      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64918",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) including DC and Puerto Rico, excluding the state of RI and the countries of Yugoslavia YEMEN Viet Nam Venezuela Uzbekistan Uruguay United Arab Emirates Tunisia Thailand Tanzania Taiwan Turkey Syria Switzerland Sweden Spain South Korea South Africa Slovenia Slovakia Singapore Saudi Arabia Russia Romania Reunion Qatar Portugal Poland Philippines Peru Pakistan Paraguay Philippines Oman Norway Nigeria New Zealand Netherlands Morocco Mexico Martinique Malaysia Macedonia Mauritania Luxembourg Lithuania Libya Lebanon Latvia Kuwait Kenya Korea Korea (Republic Of [South] Korea) KUWAIT Kazakhstan Jordan Japan Jamaica Italy Israel Ireland Iran Iraq Indonesia India Hungary Hong Kong Honduras Guadeloupe Greece Germany Georgia, Republic of French Polynesia France Finland Estonia Ecuador Egypt Denmark Dominican Republic Czech Republic Croatia Colombia China Chile Canada Bulgaria Brazil Bosnia And Herzegovina Belgium Belarus Barbados Bahrain Austria Australia Argentina Arab Emirates Algeria and United Kingdom.",
      "recall_number": "Z-1207-2013",
      "product_description": "GE Healthcare, Discovery MR750, Discovery 450, Optima 450W (with and without GEM), Signa HDx, Signa HDxt, Signa HDi, Signa Vibrant.    Indicated for use as a diagnostic imaging device.",
      "product_quantity": "4338",
      "reason_for_recall": "GE Healthcare has recently become aware of a Reformat Image Flip Issue associated with GE MR Products.    The axial and coronal reformatted images from a sagittal 3D series (such as with applications of FSPGR-3D T1, BRAVO T1, or CUBE T1) of over 126 slices may display a left-to-right image flip .  In addition, the acquired sagittal images may be annotated with left-right location annotation revers",
      "recall_initiation_date": "20120608",
      "center_classification_date": "20130501",
      "termination_date": "20141103",
      "report_date": "20130508",
      "code_info": "Mfg Lot or Serial #  00000029854MR5  00000030086MR1  00000030154MR0  00000235678MR0  00000235678MR0  00000288408MR8  00000289961MR5  00000290137MR9  00000292634MR3  00000292964MR4  00000292964MR4  00000293015MR4  00000293338MR0  00000293646MR6  00000293874MR4  00000293918MR9  00000294018MR7  00000294057MR5  00000294098MR9  00000294280MR3  00000294472MR6  00000294494MR0  00000294556MR6  00000294576MR4  00000294643MR2  00000294665MR5  00000294718MR2  00000294913MR9  00000294953MR5  00000295024MR4  00000295092MR1  00000295112MR7  00000295134MR1  00000295215MR8  00000295228MR1  00000295241MR4  00000295506MR0  00000295904MR7  00000296014MR4  00000296046MR6  00000296070MR6  00000296096MR1  00000296097MR9  00000296171MR2  00000296226MR0  00000296228MR0  00000296307MR2  00000296369MR2  00000296418MR7  00000296429MR4  00000296461MR7  00000296462MR5  00000296469MR0  00000296512MR7  00000296517MR6  00000296573MR9  00000296595MR2  00000296602MR6  00000296626MR5  00000296640MR6  00000296821MR2  00000296829MR5  00000296840MR2  00000296860MR0  00000296888MR1  00000296894MR9  00000296908MR7  00000297006MR9  00000297134MR9  00000297147MR1  00000297355MR0  00000297374MR1  00000297387MR3  00000297469MR9  00000297493MR9  00000297624MR9  00000297649MR6  00000297833MR6  00000297880MR7  00000297891MR4  00000298050MR6  00000298083MR7  00000298181MR9  00000298247MR8  00000298248MR6  00000298249MR4  00000298283MR3  00000298310MR4  00000298334MR4  00000298386MR4  00000298387MR2  00000298399MR7  00000298474MR8  00000298494MR6  00000298512MR5  00000298535MR6  00000298537MR2  00000298552MR1  00000298573MR7  00000298574MR5  00000298580MR2  00000298591MR9  00000298655MR2  00000298656MR0  00000298675MR0  00000298677MR6  00000298722MR0  00000298727MR9  00000298738MR6  00000298834MR3  00000298861MR6  00000298869MR9  00000298870MR7  00000298889MR7  00000298901MR0  00000298907MR7  00000298908MR5  00000298932MR5  00000298943MR2  00000298959MR8  00000298981MR2  00000299141MR2  00000299160MR2  00000299161MR0  00000299171MR9  00000299172MR7  00000299173MR5  00000299181MR8  00000299199MR0  00000299200MR6  00000299213MR9  00000299230MR3  00000299250MR1  00000299291MR5  00000299294MR9  00000299320MR2  00000299321MR0  00000299329MR3  00000299331MR9  00000299406MR9  00000299411MR9  00000299491MR1  00000299503MR3  00000299520MR7  00000299522MR3  00000299544MR7  00000299559MR5  00000299560MR3  00000299627MR0  00000299634MR6  00000299644MR5  00000299645MR2  00000299647MR8  00000299683MR3  00000299823MR5  00000299851MR6  00000299852MR4  00000299908MR4  00000299952MR2  00000299966MR2  00000299967MR0  00000300038MR7  00000300039MR5  00000300122MR9  00000300296MR1  00000300473MR6  00000300493MR4  00000300494MR2  00000300562MR6  00000300564MR2  00000300571MR7  00000300572MR5  00000300588MR1  00000300590MR7  00000300671MR5  00000300687MR1  00000300688MR9  00000300704MR4  00000300713MR5  00000300722MR6  00000300723MR4  00000300724MR2  00000300765MR5  00000300793MR7  00000300794MR5  00000300795MR2  00000300847MR1  00000300867MR9  00000300898MR4  00000300900MR8  00000300929MR7  00000300967MR7  00000300968MR5  00000300987MR5  00000301046MR9  00000301094MR9  00000301126MR9  00000301156MR6  00000301189MR7  00000301190MR5  00000301228MR3  00000301235MR8  00000301260MR6  00000301262MR2  00000301292MR9  00000301295MR2  00000301309MR1  00000301379MR4  00000301380MR2  00000301401MR6  00000301469MR3  00000301522MR9  00000301599MR7  00000301820MR7  00000301822MR3  00000301843MR9  00000301866MR0  00000301869MR4  00000301926MR2  00000301962MR7  00000302193MR8  00000302194MR6  00000302433MR8  00000302434MR6  00000302492MR4  00000302494MR0  00000302545MR9  00000302570MR7  00000302631MR7  00000302633MR3  00000302712MR5  00000302748MR9  00000302997MR2  00000303041MR8  00000303310MR7  00000303322MR2  00000303356MR0  00000303375MR0  00000303377MR6  0  1  151  151545  292879  1537343  00000002143YR6  00000005215TR9  00000005216YR7  00000005232YR4  00000005233YR2  00000005526YR9  00000005536YR8  00000005538YR4  00000005542YR6  00000005545YR9  00000005551YR7  00000005555YR8  00000005564YR0  00000005568YR1  00000005573YR1  00000005965YR9  00000005976YR6  00000005984YR0  00000005992YR3  00000006556YR5  00000006558YR1  00000006568YR0  00000006570YR6  00000006571YR4  00000006572YR2  00000006583YR9  00000006586YR2  00000006592YR0  00000006656YR3  00000006660YR5  00000006667YR0  00000006669YR6  00000006679YR5  00000006681YR1  00000006684YR5  00000006696YR9  00000006700YR9  00000006707YR4  00000006708YR2  00000006713YR2  00000006724YR9  00000006726YR4  00000006727YR2  00000006729YR8  00000006731YR4  00000006734YR8  00000006736YR3  00000006737YR1  00000006739YR7  00000006740YR5  00000006741YR3  00000006757YR9  00000007091YR2  00000007097YR9  00000007100YR1  00000007102YR7  00000007109YR2  00000007113YR4  00000007122YR5  00000007133YR2  00000007498YR9  00000007501YR0  00000007507YR7  00000007514YR3  00000007516YR8  00000007524YR2  00000007532YR5  00000007533YR3  00000007534YR1  00000007537YR4  00000007543YR2  00000007544YR0  00000007549YR9  00000007553YR1  00000007556YR4  00000007564YR8  00000007565YR5  00000007567YR1  00000007572YR1  00000007579YR6  00000007582YR0  00000007587YR9  00000007591YR1  00000007850YR1  00000007856YR8  00000007858YR4  00000007862YR6  00000007863YR4  00000007864YR2  00000007865YR9  00000007866YR7  00000007867YR5  00000007871YR7  00000007879YR0  00000007880YR8  00000007882YR4  00000007884YR0  00000007886YR5  00000007887YR3  00000007892YR3  00000008059YR8  00000008061YR4  00000008063YR0  00000008198YR4  00000008332YR9  00000008335YR2  00000008342YR8  00000008345YR1  00000008349YR3  00000008352YR7  00000008359YR2  00000008367YR5  00000008369YR1  00000008372YR5  00000008374YR1  00000008378YR2  00000008380YR8  00000008387YR3  00000008391YR5  00000008392YR3  00000008396YR4  00000008397YR2  00000008398YR0  00000008403YR8  00000008515YR9  00000008516YR7  00000008518YR3  00000008520YR9  00000008521YR7  00000008524YR1  00000008530YR8  00000008536YR5  00000008539YR9  00000008541YR5  00000008543YR1  00000008545YR6  00000008546YR4  00000008548YR0  00000008551YR4  00000008553YR0  00000008556YR3  00000008558YR9  00000009003YR5  00000009005YR0  00000009007YR6  00000009008YR4  00000009012YR6  00000009014YR2  00000009020YR9  00000009021YR7  00000009026YR6  00000009028YR2  00000009034YR0  00000009039YR9  00000009045YR6  00000009046YR4  00000009049YR8  00000009056YR3  00000009057YR1  00000009062YR1  00000009063YR9  00000009065YR4  00000009066YR2  00000009076YR1  00000009078YR7  00000009084YR5  00000009419YR3  00000009420YR1  00000009421YR9  00000009423YR5  00000009424YR3  00000009425YR0  00000009426YR8  00000009432YR6  00000009437YR5  00000009438YR3  00000009439YR1  00000009444YR1  00000009452YR4  00000009462YR3  00000009525YR7  00000009526YR5  00000009529YR9  00000009536YR4  00000009546YR3  00000009550YR5  00000009551YR3  00000009554YR7  00000009555YR4  00000009558YR8  00000009562YR0  00000009564YR6  00000009566YR1  00000009568YR7  00000009570YR3  00000009571YR1  00000009576YR0  00000009578YR6  00000009580YR2  00000009584YR4  00000009586YR9  00000009591YR9  00000009597YR6  00000009598YR4  00000009599YR2  00000009603YR2  00000009606YR5  00000009615YR6  00000009618YR0  00000009623YR0  00000009636YR2  00000009763YR4  00000009765YR9  00000009766YR7  00000009768YR3  00000009771YR7  00000009773YR3  00000009776YR6  00000009834YR3  00000009835YR0  00000009837YR6  00000009845YR9  00000009848YR3  00000009849YR1  00000009852YR5  00000009853YR3  00000010134YR5  00000010140YR2  00000010144YR4  00000010146YR9  00000010149YR3  00000010153YR5  00000010155YR0  00000010159YR2  00000010160YR0  00000010163YR4  00000010165YR9  00000010166YR7  00000010167YR5  00000010168YR3  00000010281YR4  00000010284YR8  00000010287YR1  00000010292YR1  00000010294YR7  00000010295YR4  00000010298YR8  00000010302YR8  00000010303YR6  00000010311YR9  00000010390YR3  00000010394YR5  00000010395YR2  00000010396YR0  00000010397YR8  00000010398YR6  00000010424YR0  00000010433YR1  00000010435YR6  00000010438YR0  00000010439YR8  00000010449YR7  00000010450YR5  00000010451YR3  00000010453YR9  00000010502YR3  00000010503YR1  00000010507YR2  00000010511YR4  00000010512YR2  00000010523YR9  00000010525YR4  00000010526YR2  00000010529YR6  00000010534YR6  00000010535YR3  00000010536YR1  00000010537YR9  00000010538YR7  00000010539YR5  00000010594YR0  00000010596YR5  00000010598YR1  00000010599YR9  00000010603YR9  00000010605YR4  00000010607YR0  00000010608YR8  00000010613YR8  00000010616YR1  00000010622YR9  00000010623YR7  00000010625YR2  00000010634YR4  00000010635YR1  00000010637YR7  00000010638YR5  00000010703YR7  00000010705YR2  00000010707YR8  00000010708YR6  00000010711YR0  00000010722YR7  00000010723YR5  00000010780YR5  00000010787YR0  00000010790YR4  00000010791YR2  00000010792YR0  00000010797YR9  00000010800YR1  00000010802YR7  00000010805YR0  00000010806YR8  00000010808YR4  00000010809YR2  00000010810YR0  00000010811YR8  00000010813YR4  00000010816YR7  00000010820YR9  00000010822YR5  00000010823YR3  00000010826YR6  00000010865YR4  00000010870YR4  00000010871YR2  00000010872YR0  00000010878YR7  00000010881YR1  00000010885YR2  00000010890YR2  00000010891YR0  00000010892YR8  00000010894YR4  00000010896YR9  00000010897YR7  00000010901YR7  00000010906YR6  00000010909YR0  00000010911YR6  00000010912YR4  00000010913YR2  00000010914YR0  00000010921YR5  00000010923YR1  00000010925YR6  00000010926YR4  00000010927YR2  00000010928YR0  00000010931YR4  00000010936YR3  00000010942YR1  00000010945YR4  00000010952YRO  00000010953YR8  00000010955YR3  00000010958YR7  00000010959YR5  00000010961YR1  00000010963YR7  00000010964YR5  00000010973YR6  00000010975YR1  00000010983YR5  00000010986YR8  00000010988YR4  00000010990YR0  00000010992YR6  00000010994YR2  00000010996YR7  00000011000YR7  00000011006YR4  00000011007YR2  00000011008YR0  00000011011YR4  00000011012YR2  00000011015YR5  00000011029YR6  00000011030YR4  00000011031YR2  00000011055YR1  00000011056YR9  00000011057YR7  00000011058YR5  00000011059YR3  00000011064YR3  00000011065YR0  00000011066YR8  00000011070YR0  00000011072YR6  00000011076YR7  00000011078YR3  00000011080YR9  00000011085YR8  00000011086YR6  00000011088YR2  00000011090YR8  00000011091YR6  00000011095YR7  00000011096YR5  00000011099YR9  00000011104YR7  00000011106YR2  00000011107YR0  00000011114YR6  00000011117YR9  00000011125YR2  00000011126YR0  00000011127YR8  00000011129YR4  00000011131YR0  00000011133YR6  00000011135YR1  00000011136YR9  00000011138YR5  00000011139YR3  00000011140YR1  00000011142YR7  00000011145YR0  00000011152YR6  00000011153YR4  00000011155YR9  00000011156YR7  00000011159YR1  00000011188YR0  00000011190YR6  00000011191YR4  00000011192YR2  00000011195YR5  00000011197YR1  00000011198YR9  00000011201YR1  00000011202YR9  00000011203YR7  00000011204YR5  00000011209YR4  00000011217YR7  00000011218YR5  00000011219YR3  00000011220YR1  00000011222YR7  00000011223YR5  00000011224YR3  00000011233YR4  00000011241YR7  00000011242YR5  00000011243YR3  00000011244YR1  00000011245YR8  00000011252YR4  00000011256YR5  00000011259YR9  00000011265YR6  00000011268YR0  00000011269YR8  00000011316YR7  00000011317YR5  00000011321YR7  00000011322YR5  00000011334YR0  00000011337YR3  00000011338YR1  00000011339YR9  00000011401YR7  00000011402YR5  00000011404YR1  00000011416YR5  00000011450YR4  00000011501YR4  00000011502YR2  00000011504YR8  00000011506YR3  00000011508YR9  00000011511YR3  00000011518YR8  00000011521YR2  00000011524YR6  00000011541YR0  00000011544YR4  00000011547YR7  00000011549YR3  00000011550YR1  00000011555YR0  00000011563YR4  00000011567YR3  00000011569YR1  00000011570YR9  00000011572YR5  00000011579YR0  00000011581YR6  00000011584YR0  00000011585YR7  00000011586YR5  00000011588YR1  00000011589YR9  00000011591YR5  00000011595YR6  00000011596YR4  00000011717YR6  00000011721YR8  00000011724YR2  00000011726YR7  00000011727YR5  00000011730YR9  00000011741YR6  00000011742YR4  00000011743YR2  00000011744YR0  00000011746YR5  00000011774YR7  00000011784YR6  00000011786YR1  00000011787YR9  00000011793YR7  00000011796YR0  00000011798YR6  00000011801YR8  00000011804YR2  00000011862YR0  00000011867YR9  00000011869YR5  00000011872YR9  00000011874YR5  00000011875YR2  00000011878YR6  00000011879YR4  00000011882YR8  00000011883YR6  00000011884YR4  00000011885YR1  00000011886YR9  00000011892YR7  00000011897YR6  00000011900YR8  00000011901YR6  00000011910YR7  00000011920YR6  00000011926YR3  00000011931YR3  00000011932YR1  00000011934YR7  00000011938YR8  00000011939YR6  00000011946YR1  00000011947YR9  00000011951YR1  00000011953YR7  00000011954YR5  00000011955YR2  00000011956YR0  00000011958YR6  00000011996YR6  00000012003YR0  00000012006YR3  00000012008YR9  00000012010YR5  00000012015YR4  00000012020YR4  00000012023YR8  00000012042YR8  00000012047YR7  00000012050YR1  00000012063YR4  00000012065YR9  00000012070YR9  00000012072YR5  00000012078YR2  00000012080YR8  00000012132YR7  00000012133YR5  00000012157YR4  00000012158YR2  00000012161YR6  00000012162YR4  00000012163YR2  00000012167YR3  00000012180YR6  00000012181YR4  00000012185YR5  00000012188YR9  00000012189YR7  00000012193YR9  00000012196YR2  00000012198YR8  00000012222YR6  00000012223YR4  00000012224YR2  00000012225YR9  00000012229YR1  00000012230YR9  00000012232YR5  00000012234YR1  00000012236YR6  00000012237YR4  00000012239YR0  00000012244YR0  00000012275YR4  00000012306YR7  00000012355YR4  00000012387YR7  00000012402YR4  00000022936MR8  00000024170MR6  00000025636MR4  00000028674MR1  00000028886MR2  00000028959MR7  00000029079MR2  00000029128MR0  00000029187MR8  00000029521MR6  00000029787MR9  00000052299YR0  0000005535YMR0  00000091BD5D84  000000HDMR1092  000000HDMR1151  000000HDMR2485  000000HDMR2612  000000HDMR2685  000000HDMR2686  000000HDMR2807  000000HDMR2843  000000HDMR2864  000000HDMR2904  000000HDMR2938  000000HDMR2950  000000HDMR2984  000000HDMR3464  000000HDMR3527  000000HDMR3710  000000HDMR4293  000000HDMR5532  000000HDMR5538  000001006B6464  00000105XF5011  00000108134YR4  00000006597YR9  00000218753MR2  00000223456MR5  00000228289MR3  00000229365MR2  00000229474MR2  00000229480MR9  00000230664MR5  00000230694MR2  00000231893MR9  00000231910MR1  00000231911MR9  00000231913MR5  00000232415MR0  00000232419MR2  00000233267MR4  00000233272MR4  00000233303MR7  00000234794MR6  00000234797MR9  00000234843MR1  00000234847MR2  00000234855MR5  00000235236MR7  00000235237MR5  00000235242MR5  00000235680MR6  00000235690MR5  00000236703MR5  00000236831MR4  00000236924MR7  00000237059MR1  00000237064MR1  00000237932MR9  00000237936MR0  00000237944MR4  00000237949MR3  00000237968MR3  00000238343MR8  00000238355MR2  00000238364MR4  00000238867MR6  00000239042MR5  00000239241MR3  00000239857MR6  00000239864MR2  00000241137MR9  00000241143MR7  00000241145MR2  00000241434MR3  00000241514MR9  00000241721MR0  00000006714YR0  00000241865MR5  00000241866MR3  00000241868MR9  00000242548MR6  00000242558MR5  00000242585MR8  00000243249MR0  00000243685MR5  00000243700MR2  00000243702MR8  00000244845MR4  00000244846MR2  00000244850MR4  00000245327MR2  00000245803MR2  00000245809MR9  00000245819MR8  00000246615MR9  00000246622MR5  00000247217MR3  00000247727MR1  00000248132MR3  00000248159MR6  00000248485MR5  00000249758MR4  00000007087YR0  00000250019MR7  00000251094MR9  00000251229MR1  00000251231MR7  00000251241MR6  00000008115YR8  00000252422MR1  00000252437MR9  00000252930MR3  00000253153MR1  00000253635MR7  00000008117YR4  00000254485MR6  00000255208MR1  00000255211MR5  00000255569MR6  00000255571MR2  00000255599MR3  00000255610MR8  00000255612MR4  00000256228MR8  00000256367MR4  00000256696MR6  00000256699MR0  00000256704MR8  00000256984MR6  00000256990MR3  00000257724MR5  00000257730MR2  00000257732MR8  00000008519YR1  00000257752MR6  00000258089MR2  00000008996YR1  00000258403MR5  00000258419MR1  00000259530MR4  00000260011MR2  00000260017MR9  00000260462MR7  00000260470MR0  00000260626MR7  00000260698MR6  00000260699MR4  00000260864MR4  00000261068MR1  00000261069MR9  00000261328MR9  00000009455YR7  00000261877MR5  00000261879MR1  00000262101MR9  00000262150MR6  00000262151MR4  00000262458MR3  00000262464MR1  00000262735MR4  00000262908MR7  00000262933MR5  00000262935MR0  00000263313MR9  00000263507MR6  00000263686MR8  00000263792MR4  00000263832MR8  00000263972MR2  00000264529MR9  00000264617MR2  00000265062MR0  00000265066MR1  00000265183MR4  00000009604YR0  00000265718MR7  00000265720MR3  00000265824MR3  00000265827MR6  00000009847YR5  00000265964MR7  00000010393YR7  00000266129MR6  00000010527YR0  00000266460MR5  00000266754MR1  00000266799MR6  00000266825MR9  00000266828MR3  00000267180MR8  00000267324MR2  00000268049MR4  00000268283MR9  00000268471MR0  00000268524MR6  00000010531YR2  00000269045MR1  00000269047MR7  00000269214MR0  00000269336MR4  00000269340MR6  00000269380MR2  00000270386MR6  00000270462MR5  00000270976MR4  00000271051MR5  00000271561MR3  00000010533YR8  00000271753MR6  00000272101MR7  00000272283MR3  00000272284MR1  00000272627MR1  00000272890MR5  00000272891MR3  00000272896MR2  00000273195MR8  00000273541MR3  00000273632MR0  00000273672MR6  00000273673MR4  00000274078MR5  00000010601YR3  00000274745MR9  00000275022MR2  00000275583MR3  00000275647MR6  00000275648MR4  00000275653MR4  00000275961MR1  00000275967MR8  00000276043MR7  00000276245MR8  00000276266MR4  00000276457MR9  00000276495MR9  00000276588MR1  00000276634MR3  00000010706YR0  00000277494MR1  00000277495MR8  00000277575MR7  00000277763MR9  00000277784MR5  00000277916MR3  00000278041MR9  00000278164MR9  00000278193MR8  00000278291MR0  00000278301MR7  00000278391MR8  00000278555MR8  00000278624MR2  00000278669MR7  00000278678MR8  00000278688MR7  00000278766MR1  00000278769MR5  00000278848MR7  00000278851MR1  00000278873MR5  00000279078MR0  00000279405MR5  00000279415MR4  00000279479MR0  00000279483MR2  00000279503MR7  00000279618MR3  00000279753MR8  00000279874MR2  00000279921MR1  00000280000MR1  00000280001MR9  00000280065MR4  00000280089MR4  00000280229MR6  00000280249MR4  00000280251MR0  00000280252MR8  00000280253MR6  00000280328MR6  00000280436MR7  00000280462MR3  00000280524MR0  00000280678MR4  00000280775MR8  00000280891MR3  00000280946MR5  00000281156MR0  00000281157MR8  00000281271MR7  00000281374MR9  00000281377MR2  00000281826MR8  00000281937MR3  00000282371MR4  00000282373MR0  00000282393MR8  00000282547MR9  00000282788MR9  00000282794MR7  00000282817MR6  00000282868MR9  00000282948MR9  00000283075MR0  00000283120MR4  00000283321MR8  00000283346MR5  00000283377MR0  00000011009YR8  00000283408MR3  00000283664MR1  00000283682MR3  00000283815MR9  00000283849MR8  00000283865MR4  00000283949MR6  00000284025MR4  00000284034MR6  00000011061YR9  00000284235MR9  00000284263MR1  00000284386MR0  00000284389MR4  00000284391MR0  00000284407MR4  00000284409MR0  00000284466MR0  00000284472MR8  00000284490MR0  00000284563MR4  00000284576MR6  00000284599MR8  00000284640MR0  00000284709MR3  00000284735MR8  00000284838MR0  00000284839MR8  00000284864MR6  00000011123YR7  00000285055MR0  00000285218MR4  00000285225MR9  00000285227MR5  00000285490MR9  00000285497MR4  00000285499MR0  00000285775MR3  00000285796MR9  00000285927MR0  00000286006MR2  00000286010MR4  00000286016MR1  00000286091MR4  00000286094MR8  00000286249MR8  00000286305MR8  00000286337MR1  00000286526MR9  00000286701MR8  00000286743MR0  00000286767MR9  00000286785MR1  00000286843MR8  00000286851MR1  00000286853MR7  00000286886MR7  00000286982MR4  00000287026MR9  00000287078MR0  00000287081MR4  00000287391MR7  00000287442MR8  00000287626MR6  00000287628MR2  00000287649MR7  00000287767MR8  00000287770MR2  00000287826MR2  00000287846MR0  00000287864MR3  00000287940MR1  00000288015MR1  00000011711YR9  00000288096MR1  00000288097MR9  00000288192MR8  00000288383MR3  00000288651MR3  00000288690MR1  00000288702MR4  00000011873YR7  00000011877YR8  00000288980MR6  00000289016MR8  00000289024MR2  00000289042MR4  00000289270MR1  00000289438MR4  00000011927YR1  00000289525MR8  00000289763MR5  00000289771MR8  00000289789MR0  00000289792MR4  00000011941YR2  00000290052MR0  00000290130MR4  00000290141MR1  00000290192MR4  00000290370MR6  00000290478MR7  00000290655MR0  00000291370MR5  00000291681MR5  00000291750MR8  00000291805MR0  00000291979MR3  00000292432MR2  00000292474MR4  00000292635MR0  00000292792MR9  00000292820MR8  00000292823MR2  00000292922MR2  00000292980MR0  00000292986MR7  00000293047MR7  00000293069MR1  00000012017YR0  00000293299MR4  00000293315MR8  00000293317MR4  00000293333MR1  00000293342MR2  00000012272YR1  00000293432MR1  00000293442MR0  00000012341YR4  00000293510MR4  00000293592MR2  00000293614MR4  00000293631MR8  00000293639MR1  00000293657MR2  00000293658MR1  00000293671MR4  00000293682MR1  00000293711MR8  00000293712MR6  000000HDMR1086  00000293937MR9  00000293959MR3  00000293960MR1  00000293969MR2  00000294003MR9  00000241861MR4  00000294100MR3  00000294102MR9  00000294106MR0  00000294118MR5  00000294144MR1  00000294158MR1  00000294160MR7  00000294272MR0  00000294273MR8  00000294304MR1  00000294319MR9  00000294320MR7  00000294342MR1  00000294394MR2  00000249768MR3  00000294406MR4  00000294409MR8  00000294442MR9  00000294444MR5  00000294515MR2  00000294519MR4  00000294523MR6  00000294524MR4  00000294553MR3  00000294554MR1  00000251800MR9  00000294605MR1  00000294625MR9  00000294667MR1  00000294668MR9  00000294694MR5  00000294696MR0  00000294737MR2  00000294738MR0  00000254665MR3  00000294911MR3  00000294945MR1  00000294959MR2  00000294990MR7  00000295004MR6  00000295025MR1  00000295026MR9  00000295027MR7  00000295052MR5  00000295058MR2  00000258090MR0  00000295102MR8  00000295139MR0  00000295173MR9  00000295176MR2  00000295190MR3  00000295191MR1  00000295216MR6  00000295218MR2  00000295229MR9  00000295233MR1  00000295235MR6  00000295476MR6  00000295482MR4  00000295512MR8  00000295518MR1  00000295592MR0  00000295593MR8  00000295701MR7  00000295741MR3  00000295742MR1  00000295743MR9  00000295789MR2  00000295806MR4  00000295807MR2  00000295810MR6  00000295876MR7  00000295898MR1  00000296104MR3  00000296111MR8  00000296144MR9  00000296145MR6  00000261329MR7  00000296223MR1  00000296224MR9  00000296292MR6  00000265596MR7  00000296303MR1  00000296364MR3  00000296367MR6  00000296441MR9  00000296519MR2  00000296575MR4  00000296631MR5  00000296830MR3  00000296848MR5  00000297011MR9  00000297185MR1  00000297201MR6  00000297205MR7  00000297370MR9  00000297371MR7  00000297383MR2  00000297419MR4  00000265939MR9  00000297447MR5  00000297475MR6  00000297486MR3  00000297488MR9  00000297627MR2  00000297656MR1  00000297698MR3  00000297700MR7  00000297726MR2  00000297852MR6  00000297879MR9  00000297893MR0  00000298044MR9  00000298045MR6  00000298187MR6  00000298220MR5  00000266073MR6  00000298223MR9  00000298313MR8  00000298413MR6  00000298468MR0  00000298496MR1  00000298497MR9  00000298502MR6  00000298504MR2  00000298533MR1  00000266214MR6  00000298657MR8  00000298708MR9  00000298709MR7  00000298710MR5  00000298716MR2  00000298718MR8  00000271646MR2  00000298792MR3  00000298809MR5  00000298810MR3  00000298826MR9  00000298930MR9  00000298945MR7  00000298946MR5  00000298947MR3  00000298949MR9  00000298952MR3  00000298968MR9  00000298974MR7  00000299162MR8  00000274744MR2  00000299166MR9  00000276678MR0  00000299327MR7  00000284234MR2  00000299350MR9  00000299416MR8  00000299422MR6  00000299424MR2  00000299449MR9  00000299471MR3  00000299523MR1  00000299525MR6  00000299527MR2  00000299642MR9  00000299660MR1  00000299662MR7  00000299758MR3  00000299762MR5  00000299828MR4  00000299850MR8  00000299882MR1  00000299885MR4  00000299911MR8  00000299949MR8  00000299962MR1  00000300458MR7  00000300496MR7  00000300498MR3  00000300517MR0  00000300577MR4  00000300619MR4  00000300674MR9  00000300702MR8  00000284902MR4  00000300799MR4  00000300811MR7  00000288822MR0  00000300902MR4  00000300904MR0  00000300909MR9  00000300937MR0  00000300950MR3  00000300951MR1  00000288852MR7  00000300970MR1  00000300984MR2  00000301020MR4  00000293343MR0  00000301105MR3  00000301188MR9  00000301212MR7  00000301299MR4  00000301301MR8  00000301340MR6  00000301349MR7  00000301356MR2  00000301463MR6  00000301501MR3  00000301565MR8  00000301804MR1  00000301818MR1  00000301847MR0  00000301858MR7  00000302424MR7  00000302510MR3  00000302511MR1  00000302523MR6  00000302625MR9  00000751BD2DAE  00000806BD3416  00000831BD49F9  00000831BD4A1D  00000833BD4AD4  00000833BD4ADC  00000833BD4ADE  00000293731MR6  00000836BD4D87  00000836BD4DA8  00000836BD4DAF  00000836BD4DB4  00000836BD4DDB  00000837BD4E1A  00000837BD4E1E  00000837BD4E66  00000838RD4EA3  00000845BD51B0  00000846BD5247  00000848BD5325  00000903BD55CD  00000903BD55EB  00000912BD599A  00000912BD599C  00000913BD5A49  00000913BD5A62  00000294015MR3  00000915BD5AED  00000916BD5B05  00000919BD5B98  00000294402MR3  00000930BD5D36  00000930BD5D54  00000933BD5E06  00000933BD5E09  00000937BD5ED5  00000937BD5EF3  00000939BD5F41  00000939BD5F4B  00000945BD6180  00000945BD6181  00000947BD61BF  00000948BD61FD  00000950BD6297  00000950BD62BA  00000HDML10300  00000294557MR4  00000HDMR10599  00000HDMR10673  00000HDMR10743  00000HDMR10762  00000HDMR10899  00000HDMR10944  00000HDMR20043  00000HDMR20069  00000HDMR20189  00000HDMR20256  00000HDMR20258  00000HDMR20269  00001001BD6328  00001001BD6334  00001001BD633B  00000294892MR5  00001003BD6409  00000295088MR9  00001003BD6416  00001003BD642C  00001008BD6517  00001009BD6548  00001009BD654D  00001009BD654E  00001009BD6560  00001009BD65F7  00001010BD6609  00000296301MR5  00001011BD662A  00001011BD6649  00001011BD6684  00001043XF5007  00001045XF500B  00001045XF500E  00001046XF5003  00001046XF5009  00001046XF500F  00000299251MR9  00001048XF501C  00001050XF5008  00001050XF5010  00001050XF501E  00001050XF501F  00001050XF5027  00001051XF5018  00001052XF5028  00001052XF504E  00001052XF504F  00001052XF5059  00001052XF5091  00001101XF50A1  00001101XF50A4  00001101XF50AE  00001101XF50C1  00001101XF50CA  00001104XF5011  00001104XF507D  00001104XF507E  00001104XF5081  00001104XF5083  00001104XF5087  00001104XF50A2  00001105XF5015  00001105XF5018  0000267533MR8A  00002891887MR7  00010A08BD6534  000S0919BD5BC0  000S0930BD5D48  00000833BD4ADF  000S0933BD5DFD  000S0933BD5E02  000S1001BD6331  000S1002BD634A  00000250019MR7  00000294444MR5  00000919BD5BB0  00000012239YR0  00001002BD6362  2102  11797  00000001060YR5  00000001206YR5  00000005231YR6  00000005540YR0  00000005927YR5  00000006594YR6  00000006596YR1  00000006675YR3  00000006705YR8  00000006743YR9  00000006747YR0  00000007094YR6  00000007116YR7  00000007117YR5  00001011BD661E  00000007869YR1  00000008062YR2  00000008361YR8  00000008379YP0  00000008517YR5  00000008549YR8  00000008550YR6  00000008555YR4  00000008993YR*  00000008994YR6  00000009015YR9  00000009029YR0  00000009031YR6  00000009042YR3  00000009044YR9  00000009059YR7  00000009064YR7  00000009067YR0  00000009069YR6  00000009077YR7  00000009081YR1  00000009416YR9  00000009427YR6  00000009440YR9  00000009547YR1  00000009553YR9  00000009593YR5  00000009620YR6  00000009626YR3  00000009858YR2  00000009886YR3  00000009973YR9  00000009977YR0  00000009988YR7  00000010132YR9  00000010151YR9  00000010154YR3  00000010297YR0  00000010304YR4  00000010307YR7  00000010429YR9  00000010517YR1  00000010595YR7  00000010617YR9  00000HDMR10559  00000010873YR8  00000010907YR4  00000010918YR1  00000010919YR9  00000011028YR8  00000011033YR8  00000011053YR6  00000011266YR4  00000011329YR0  00000011551YR9  00000011571YR7  00000011578YR2  00000011722YR6  00000011887YR7  00000011890YR1  00000011950YR3  00000012231YR7  00000012240YR8  00000012241YR6  00000275464MR6  00000279882MR5  000S0933BD5DF3  00000283306MR9  00000288159MR7  00000289023MR4  00000833BD4AB7  00000845BD51AF  00000845BD51C0  00000849BD5391  00000903BD55F3  00000912BD59C2  00000913BD5A45  00000916BD5B08  00000930BD5D51  00000930BD5D58  00000933BD5E15  00000937BD5EE1  00000937BD5F03  00000011088YR2  00000952BD62FE  00001001BD631E  00001002BD634E  00001002BD636A  00001003BD6405  00001009BD65B3  00001009BD6563  00001012BD66A7  00001012BD66CB  00001012BD66D0  00001012BD66E8  00001012BD66F8  00001037XF5099  00001042XF5027  00001043XF5011  00001045XF500F  00001046XF5002  00001048XF500B  00001048XF500E  00001051XF5004  00001051XF5040  00001052XFS025  00001052XF5007  00001052XF5019  00001052XF5058  00001052XF5074  00001052XF509E  00001101XF50B9  00001101XF5084  00001101XF5089  00001101XF509B  00001101XF5099  00001102XF503E  00001104XF50B4  00001104XF501D  00001104XF509E  00001104XF5093  00001106XF505E  00001134FMM112  00001142FMMONW  00001142FMM0ND  00001142FMM0NJ  00001142FMM0NM  00001142FMM0P1  00001142FMM0P3  00001147FMM0C1  00001201FMMOFR  00001201FMM0FN  00001201FMM0FV  00001201FMM0GM  00001201FMM0HK  00001201FMM0H2  00001208FMM03F  00001211FMMOTO  00001211FMM0RJ  00001211FMM0RU  00001211FMM0RW  00001211FMM0R9  00001211FMM0TB  00001211FMM0T3  00001219FMM0ER  00001219FMM0EW  00001219FMM0FG  00001219FMM0FJ  00001219FMM0FT  00001219FMM0F0  00001219FMM0F5  00001219FMM0G3  00001221FMM18A  00001221FMM194  000021285E6490  00001221FMM184  00002930008MR9  00000299515MR7  00000300732MR5  00000300792MR9  00000301597MR1  00000301752MR2  00000301924MR7  00000302248MR0  00000302688MR7  00000302690MR3  00000303439MR4  00000303805MR6  00000304177MR9  00000288907MR9  00000289309MR7  00000289310MR5  00000290050MR4  00000290202MR1  00000290457MR1  00000290502MR4  00000290636MR0  00000290659MR2  00000290704MR6  00000290773MR1  00000290957MR0  00000291185MR7  00000291325MR9  00000291363MR0  00000291397MR8  00000291464MR6  00000291489MR3  00000291877MR9  00000291980MR1  00000292367MR0  00000292431MR4  00000292471MR0  00000292528MR7  00000292843MR0  00000293261MR4  00000293561MR7  00000294398MR3  00000294698MR6  00000294999MR8  00000295111MR9  00000295802MR3  00000297974MR8  00000297877MR3  00000296040MR9  00000296841MRO  00000297133MR1  00000298114MR0  00000298836MR8  00000298960MR6  000005148810-3  00001026XF5154  00001101XF5030  00001106XF503F  00001106XF5046  00001106XF5056  00001106XF5071  00001106XF509C  00001027XF500C  00001027XF5001  00001027XF5004  00001046XF5004  00001049XF500F  00001049XF5005  00001050XF5024  00001052XF5031  00001052XF5053  00001101XF50AA  00001101XF50AD  00001101XF50BC  00001101XF50B2  00001101XF50B6  00001101XF50B7  00001101XF504B  00001101XF5047  00001101XF5054  00001101XF5059  00001101XF508C  00001101XF5081  00001102XF500A  00001102XF500F  00001102XF5003  00001102XF501B  00001102XF5019  00001102XF502F  00001102XF5076  00001106XF500A  00001106XF500B  00001106XF500C  00001106XF500F  00001106XF5001  00001106XF5003  00001106XF5009  00001106XF501D  00001106XF5010  00001106XF5014  00001106XF5016  00001106XF5019  00001106XF5020  00000303646MR4  00000301841MR3  00001117FMM0C8  00001118FMM0WL  00001134FMM103  00001140FMM0G5  00001147FMM0E9  00001203FMMOUJ  00001203FMMOUV  00001203FMM0UL  00001203FMM0UU  00001203FMM0UY  00001203FMM0VJ  00001203FMM0VL  00001203FMM0VR  00001203FMM0VU  00001203FMM0V0  00001203FMM0V3  00001203FMM0V4  00001203FMM0V8  00001204FMM04F  00001204FMM04K  00001204FMM04N  00001204FMM04W  00001204FMM04X  00001204FMM046  00001204FMM047  00001204FMM05B  00001204FMM056  00001204FMM059  00001219FMM0GX  00001220FMM01D  00001221FMM06U  00001227FMM00L  00001117FMM0CW  00001140FMM0FV  00000275650MR0  00000292973MR5  00000293642MR5  00000293724MR1  00000294332MR2  00000294967MR5  00000295709MR0  00000296321MR3  00000297132MR3  00000298102MR5  00000298493MR8  00000298501MR8  00000299340MR0  00000299684MR1  00000299685MR8  00000299825MR0  00000301296MR0  00000301872MR8  00000005549YR1  00000006565YR6  00000006566YR4  00000006710YR8  00000006716YR5  00000007104YR3  00000007120YR9  00000007496YR3  00000007557YR2  00000007849YR3  00000007852YR7  00000008100YR0  00000008120YR8  00000008320YR4  00000008337YR8  00000008344YR4  00000008544YR9  00000011870YR3  00000009617YR2  00000009855YR8  00000010164YR2  00000010597YR3  00000010825YR8  00000010898YR5  00000011208YR6  00000011328YR2  00000011720YR0  00000011863YR8  00000011943YR8  00000011980YR2  00000HDMR10401  00000012024YR6  000000HDMR2484  000000HDMR3707  00000241144MR5  00000250199MR7  00000265724MR5  00000278312MR4  00000280526MR5  00000280578MR6  00000280690MR9  00000282949MR7  00000286008MR8  00000287838MR7  00000288591MR1  00000293527MR8  00000293870MR2  00000293876MR9  00000296409MR6  00000297836MR9  00000298555MR4  00000300580MR8  00000918BD5B34  00000HDMR10764  00",
      "more_code_info": "001002BD6340  000000HDMR2932  00000006748YR8  00000007523YR4  00000007594YR5  00000008098YR6  00000008329YR5  00000008399YR8  00000009992YR9  00000010614YR6  00000010922YR3  00000010933YR0  00000277444MR6  00000279870MR0  00000281465MR5  00000912BD59C6  00001002BD633F  00001011BD668C  00001201FMM0HB  00000300712MR7  00000303280MR2  00000290456MR3  00000290545MR3  00000291274MR9  00000029804MR5  00000296514MR3  00001106XF501E  00001203FMM0VT  00001147FMM0EF  00001147FMM0E7  00000299646MR0  00000300796MR0  00000007103YR5  00000010869YR6  00000011084YR1  00000010139YR4  00000010308YR5  00000010391YR1  00000010442YR2  00000010785YR4  00000010798YR7  00000011026YR2  00000011255YR7  00000011552YR7  000001106XF022  00000241723MR6  00000273525MR6  00000282690MR7  00000287024MR4  00000288979MR8  00000289785MR8  00000293936MR1  00000293951MR0  00000294271MR2  00000294441MR1  00000294514MR5  00000294607MR7  00000294669MR7  00000294907MR1  00000295014MR5  00000295056MR6  00000295477MR4  00000295481MR6  00000296360MR1  00000296366MR8  00000296404MR7  00000297973MR0  00000298093MR6  00000298097MR7  00000298488MR8  00000298843MR4  00000298975MR4  00000299348MR3  00000299417MR6  00000299451MR5  00000299494MR5  00000299675MR9  00000300186MR4  00000301039MR4  00000010796YR1  00000010866YR2  00000011719YR2  00000012013YR9  00000012014YR7  00000288343MR7  00000294510MR3  00001004BD6448  00000007584YR6  00000010704YR5  00000293217MR6  00000298534MR9  00000300778MR8  00001052XF50AE  00000295174MR7  00000296558MR0  00000296846MR9  00000296863MR4  00000297381MR6  00001003BD642F  00000294308MR2  00000294986MR5  00000298063MR9  00000300501MR4  00000300849MR7  00000301739MR9  00000006723YR1  00000007114YR2  00000007847YR7  00000008325YR3  00000009054YR8  00000268685MR5  00000011097YR3  00000298557MR0  00000011189YR8  00000011199YR7  00000011326YR6  00000011561YR8  00000012059YR2  00000012121YR0  00000275533MR8  00000285677MR1  00000286649MR9  00000289162MR0  00000290054MR6  00000294344MR7  00000296238MR9  00000296865MR9  00000298215MR5  00000912BD59A1  00001012BD66A2  00000009574YR5  00000010254YR3  00000011553YR5  00001046XF502A  00001047XF5005  00001052XF509C  00001102XF502E  00001219FMM0FY  00000303326MR3  00001101XF5025  00000298265MR0  00000298790MR7  00000298804MR6  00000298806MR1  00000301038MR6  00001011BD667B  00000294516MR0  00000297422MR8  00000298944MR0  00000299292MR3  00000301293MR7  00000301963MR5  00000005567YR3  00000007520YR0  00000007539YR0  00000010425YR7  00000290007MR4  00000011734YR1  00000241874MR7  00000262064MR9  00000275175MR8  00000283302MR8  00000284233MR4  00001001BD631D  UNK51078490008  00000006721YR5  00000009418YR5  00000009531YR5  00000297175MR2  00000297386MR5  00000293521MR1  00000293920MR5  00000295538MR3  00000295710MR8  00000298674MR3  00000300298MR7  00000006665YR4  00000008532YR4  00000009456YR5  00000012127YR7  00000012187YR1  00000027810MR1  000000HDMR5592  00000295902MR1  00000296047MR4  00000297404MR6  00000950BD628D  00000950BD62AA  00000009429YR2  00000291675MR7  00000293000MR6  00000296547MR3  00001106XF50B1  00001204FMM04T  00001140FMM0G1  00000297154MR7  00000299452MR3  00000300452MR0  00000301496MR6  00000302432MR0  00000008311YR3  00000010141YR0  00000010309YR3  000000105258M0  00000010618YR3  00000010972YR8  00000010991YR8  00000293935MR3  00000011206YR0  00000011221YR9  00000011232YR6  00000011264YR9  00000011324YR1  00000011332YR4  00000011788YR7  00000012011YR3  00000012046YR9  00000012134YR3  00000243187MR2  00000261729MR8  00000265005MR9  00000275466MR1  00000280781MR6  00000281056MR2  00000283516MR3  00000286157MR3  00000287425MR3  00000287847MR8  00000288193MR6  00000289319MR6  00000289339MR4  00000293672MR2  00000294285MR2  00000294329MR8  00000294396MR7  00000297421MR0  00000297905MR2  00000298658MR6  00000299648MR6  00000299650MR2  00000300700MR2  00000914BD5A83  00000VALUE1159  00000VALUE1177  00001003BD6426  00000006555YR7  00000010068YR5  00000010917YR3  00000280327MR8  00000950BD6286  00001043XF5013  00001043XF5014  00001045XF500D  00001046XF502B  00001048XF5021  00001102XF5002  00000297007MR7  00000298272MR6  00000298331MR0  00000299267MR5  00000299419MR2  00000300733MR3  00000301050MR1  00000301193MR9  00000006662YR1  00000007497YR1  00000007542YR4  00000007568YR9  00000008116YR6  00000008322YR0  00000008365YR9  00000008384YR0  00000008562YR1  00000010300YR2  00000010448YR9  00000010712YR8  00000010874YR6  00000010934YR8  00000011150YR0  00000011154YR2  00000011309YR2  00000012311YR7  00000256706MR3  00000262807MR1  00000273674MR2  00000284863MR8  00000285435MR4  00000289198MR4  00000290689MR9  00000293585MR6  00000294322MR3  00000294506MR1  00000294606MR9  00000294961MR8  00000295051MR7  00000295089MR7  00000296363MR5  00000296642MR2  00000297008MR5  00000297697MR5  00000297821MR1  00000297878MR1  00000298487MR0  00000300504MR8  00000301764MR7  00000302644MR0  00000903BD55DE  000S0934BD5E4A  00000009538YR0  00000011158YR3  00000845BD51D2  00000918BD5B36  00000933BD5E0C  00000300476MR9  00001105XF501F  00001105XF5021  00001142FMM0NH  00001203FMMOMV  00001208FMM03E  00001219FMM0ET  00001219FMM0G4  00001101XF506C  00001102XF509B  00001106XF5004  00000299140MR4  00000007550YR7  00000010804YR3  00000011503YR0  00000238341MR2  00000284387MR8  00000285018MR8  00000298647MR9  00000298972MR1  00001043XF5019  00000288040MR9  00000012323YR2  00001142FMMONN  00000011560YR0  00000301494MR1  00000005539YR2  00000007495YR5  00000009532YR3  00000011063YR5  00000011713YR5  00000012049YR3  00000269046MR9  00000278844MR6  00000285352MR1  00000288888MR1  00000294392MR6  00000293984MR1  00000296613MR3  00000007528YR3  00000010520YR5  00000010789YR6  00000011112YR0  00000011330YR8  00000011515YR4  00000282382MR1  00000293644MR1  00000294130MR0  00000295700MR9  00000295762MR9  00000295877MR5  00000298707MR1  00000300745MR7  00000010057YR8  00000011307YR6  00000011593YR1  00000286965MR9  00001003BD641A  00001052XF50A8  00000299965MR4  00000298266MR8  00001203FMM0UN  00000300865MR3  00000005571YR5  00000007589YR5  00000009582YR8  00000009613YR1  00000010293YR9  00000010306YR9  00000010868YR8  00000011005YR6  00000011032YR0  00000011419YR9  00000011447YR0  00000011554YR3  00000011771YR3  00000011781YR2  00000011794YR5  00000012271YR3  00000229366MR0  00000265607MR2  00000280923MR4  00000293051MR9  00000293591MR4  00000294693MR7  00000294895MR8  00000294946MR9  00000295090MR5  00000297012MR7  00000298042MR3  00000298684MR2  00000299253MR5  00000299910MR0  00000302358MR7  00001008BD652C  00000009460YR7  00000011262YR3  00000011323YR3  00000011559YR2  00000011795YR2  00000011802YR6  00000110XF5079  00000288387MR4  00001101XF507E  00001102XF5049  00001211FMM0TE  00000301051MR9  00000296594MR5  00001147FMM0EG  00000288041MR7  00000011542YR8  00000011785YR3  00000251799MR3  00000286263MR9  00000294692MR9  00000012348YR9  00000298309MR6  00000011187YR2  00000298531MR5  00000300810MR9  00000300982MR6  00000301848MR8  00001003BD642E  00000012191YR3  00000010631YR0  00000010784YR7  00000011194YR8  00000278463MR5  00000296237MR1  00000298758MR4  00001003BD640D  00000296493MR0  00000296622MR4  00000011868YR7  00000009548YR9  00000009839YR2  00000010070YR1  00000010899YR3  00000010985YR0  00000011226YR8  00000011231YR8  00000300798MR6  00000012052YR7  00000278605MR1  00000283127MR9  00000287275MR2  00000290000MR9  00000291166MR7  00000293023MR8  00000293523MR7  00000293615MR1  00000293875MR1  00000296362MR7  00000298276MR7  00000298679MR2  00000298955MR6  00000298970MR5  00000299168MR5  00000301385MR1  00000VALUE1166  00000010027YR1  00000912BD59A4  00001043XF5003  00001043XF5010  00001101XF5003  00001102XF5008  00000008315YR4  00000008987YR0  00000009539YR8  00000294525MR1  00000011733YR3  00001009BD6579  00000009612YR3  00000011412YR4  00000295829MR6  00000296864MR2  00000297013MR5  00000295830MR4  00000297987MR0  00000007105YR0  00000012253YR1  000000HDMR3417  00000280068MR8  00000282923MR2  00000279398MR2  00001204FMM05H  00000005561YR6  00000008095YR2  00000008101YR8  00000009009YR2  00000011279YR7  00000238868MR4  00000259581MR7  00000268511MR3  00000271125MR7  00000276137MR7  00000276265MR6  00000277531MR0  00000286384MR2  00000010143YR6  00001003BD640F  00000288409MR6  00000293603MR7  00000005982YR4  00000009560YR4  00000239536MR6  00000241131MR2  00000245669MR7  00000259379MR6  00000285167MR3  00000294101MR1  00000299830MR0  00000009871YR5  00000011010YR6  00001219FMM0G6  00000303279MR4  00000292216MR9  00000011102YR1  00000295015MR2  00000299757MR5  00000010824YR1  00000011782YR0  00000283770MR6  00000294156MR5  1112  00000006589YR6  00000006702YR5  00000010156YR8  00000010290YR5  00000010291YR3  00000010440YR6  00000010509YR8  00000011318YR3  00000011779YR6  00000253167MR1  00000262791MR7  00000284833MR1  00000010509YR8  00000294116MR9  00000294722MR4  00000295474MR1  00000296365MR0  00000298213MR0  00000298845MR9  00000008522YR5  00000010169YR1  00000010524YR7  00001049XF5026  00000299521MR5  00000011406YR6  00000009569YR5  00000009583YR6  00000011513YR9  00000278389MR2  00000284710MR1  00000290194MR0  00000298584MR4  00000939BD5FA4  00000239237MR1  00000012077YR4  00000010137YR8  00000011942YR0  00000011881YR0  00000288823MR8  00000296521MR8  00000301503MR9  00000005553YR3  00000295540MR9  00001102XF50A4  00000299854MR0  00000301844MR7  00000011228YR4  00000279563MR1  00000831BD4A04  00000848BD5316  00000298211MR4  00000294520MR2  00000298754MR3  00001102XF500D  00000009757YR6  00000297489MR7  00000298069MR6  00000298189MR2  00000298566MR1  00000299147MR9  00000299293MR1  00000299330MR1  00000299407MR7  00000299455MR6  00000299545MR4  00000299546MR2  00000300791MR1  00000301194MR7  00000301970MR0  00000302516MR0  00000006672YR0  00000007092YR0  00000007128YR2  00000007130YR8  00000007503YR6  00000007504YR4  00000007526YR7  00000007574YR7  00000007855YR0  00000008122YR4  00000008195YR0  00000008324YR6  00000008371YR7  00000008527YR4  00000008986YR2  00000008992YR0  00000009027YR4  00000009032YR4  00000009051YR4  00000009058YR9  00000009422YR7  00000009556YR2  00000009619YR8  00000009844YR2  00000010130YR3  00000010157YR6  00000010158YR4  00000010296YR2  00000010399YR4  00000010441YR4  00000010487YR7  00000010505YR6  00000010514YR8  00000010519YR7  00000010627YR8  00000010629YR4  00000010713YR6  00000010782YR1  00000010793YR8  00000010814YR2  00000010875YR3  00000010876YR1  00000010888YR6  00000010902YR5  00000010904YR1  00000010947YR0  00000010950YR4  00000010956YR1  00000010960YR3  00000010965YR2  00000010969YR4  00000010989YR2  00000010998YR3  00000011068YR4  00000011100YR5  00000011157YR5  00000011200YR3  00000011253YR2  00000011261YR5  00000011263YR1  00000011267YR2  00000011271YR4  00000011505YR5  00000011545YR1  00000011557YR6  00000011597YR2  00000011723YR4  00000011729YR1  00000011735YR8  00000011737YR4  00000011773YR9  00000011777YR0  00000011799YR4  00000011891YR9  00000011912YR3  00000011922YR2  00000011959YR4  00000011995YR8  00000011999YR0  00000012002YR2  00000012004YR8  00000012016YR2  00000012022YR0  00000012041YR0  00000012045YR1  00000012048YR5  00000012053YR5  00000012055YR0  00000012058YR4  00000012061YR8  00000012079YR0  00000012123YR6  00000012124YR4  00000012130YR1  00000012156YR6  00000012182YR2  00000012238YR2  00000012309YR1  00000012318YR2  00000012320YR8  00000012347YR1  000000HDMR4345  00000150XF5002  00000232401MR0  00000232405MR1  00000236701MR9  00000237697MR8  00000243250MR8  00000245674MR7  00000245820MR6  00000247457MR5  00000248172MR9  00000248175MR2  00000249230MR4  00000249655MR2  00000249757MR6  00000251700MR1  00000252424MR7  00000254196MR9  00000254489MR8  00000254504MR4  00000257427MR5  00000257442MR4  00000258402MR7  00000259647MR6  00000261680MR3  00000267728MR4  00000268047MR8  00000268301MR9  00000268417MR3  00000268470MR2  00000268690MR5  00000268692MR1  00000269436MR2  00000269584MR9  00000270097MR9  00000270528MR3  00000272536MR4  00000272968MR9  00000273266MR7  00000274169MR2  00000275964MR5  00000275968MR6  00000276244MR1  00000276434MR8  00000276679MR8  00000277533MR6  00000277643MR3  00000277691MR2  00000277920MR5  00000277960MR1  00000277961MR9  00000278127MR6  00000278163MR1  00000278221MR7  00000278556MR6  00000278801MR6  00000279407MR1  00000280780MR8  00000280782MR4  00000281078MR6  00000281364MR0  00000281375MR6  00000281934MR0  00000282003MR3  00000282509MR9  00000282550MR3  00000283114MR7  00000283116MR2  00000283406MR7  00000283717MR7  00000283773MR0  00000284261MR5  00000284392MR8  00000284759MR8  00000284869MR5  00000285051MR9  00000285053MR5  00000285165MR7  00000285437MR0  00000285627MR6  00000285628MR4  00000285674MR8  00000286007MR0  00000286272MR0  00000286335MR5  00000286469MR2  00000286721MR6  00000287070MR7  00000287269MR5  00000287314MR9  00000287399MR0  00000287866MR8  00000288073MR0  00000288160MR5  00000288359MR3  00000288939MR2  00000289320MR4  00000289384MR0  00000289655MR3  00000289657MR9  00000290324MR3  00000291306MR9  00000292078MR3  00000292288MR8  00000292642MR6  00000292688MR9  00000292942MR0  00000292983MR4  00000293323MR2  00000293426MR3  00000293562MR5  00000293626MR8  00000293680MR5  00000293879MR3  00000293881MR9  00000293883MR5  00000293884MR3  00000293953MR6  00000293974MR2  00000294002MR1  00000294099MR7  00000294105MR2  00000294115MR1  00000294321MR5  00000294324MR9  00000294555MR8  00000294709MR1  00000294736MR4  00000294894MR1  00000294905MR5  00000294926MR1  00000294927MR9  00000294960MR0  00000295012MR9  00000295095MR4  00000295189MR5  00000295192MR9  00000295234MR9  00000295237MR2  00000295240MR6  00000295475MR8  00000295542MR5  00000295571MR4  00000295697MR7  00000295702MR5  00000295745MR4  00000295809MR8  00000295895MR7  00000295896MR5  00000295897MR3  00000295901MR3  00000296036MR7  00000296037MR5  00000296049MR0  00000296072MR2  00000296092MR0  00000296108MR4  00000296213MR2  00000296304MR9  00000296306MR4  00000296358MR5  00000296361MR9  00000296402MR1  00000296405MR4  00000296411MR2  00000296426MR0  00000296457MR5  00000296643MR0  00000297204MR0  00000297463MR2  00000297626MR4  00000297810MR4  00000297903MR7  00000297904MR5  00000298076MR1  00000298186MR8  00000298197MR5  00000298212MR2  00000298282MR5  00000298411MR0  00000298549MR7  00000298558MR8  00000298582MR8  00000298671MR9  00000298715MR4  00000298756MR8  00000298825MR1  00000298857MR4  00000298948MR1  00000298954MR9  00000299183MR4  00000299186MR7  00000299194MR1  00000299203MR0  00000299268MR3  00000299270MR9  00000299274MR1  00000299308MR7  00000299315MR2  00000299316MR0  00000299470MR5  00000299473MR9  00000299493MR7  00000299514MR0  00000299524MR9  00000299641MR1  00000299666MR8  00000299827MR6  00000299829MR2  00000299912MR6  00000300184MR9  00000300185MR6  00000300495MR9  00000300497MR5  00000300578MR2  00000300771MR3  00000301065MR9  00000301338MR0  00000301339MR8  00000301351MR3  00000301405MR7  00000301736MR5  00000301737MR3  00000301815MR7  00000301816MR5  00000301817MR3  00000301833MR0  00000301859MR5  00000301910MR6  00000301916MR3  00000302504MR6  00000302623MR4  00000302677MR0  00000806BD3403  00000834BD4B8B  00000903BD55E8  00000913BD5A5B  00000950BD6299  00000950BD62B7  00000VALUE1281  00000VALUE1294  00001003BD6407  00001009BD654C  00001009BD6585  00001009BD65FB  00001011BD665D  00001011BD668D  00001012BD66B0  00001043XF5006  00001050XF5029  00001052XF5065  00001052XF506D  00001052XF50AC  00001101XF50A3  00001101XF50A6  00001101XF50BA  00001104XF5082  00001104XF5084  00001105XF5024  000S0934BD5E3A  00000006706YR6  00000006725YR6  00000006754YR6  00000007111YR8  00000007525YR9  00000009563YR8  00000009762YR6  00000010031YR3  00000012001YR4  00000012074YR1  00001002BD63A5  00001002BD6344  00001002BD638F  00001009BD6568  00001037XF5078  00001042XF5038  00001049XF5024  00001052XF5024  00001101XF50B3  00001102XF5020  00001102XF506C  00001104XF5097  00001105XF5003  00001105XF5010  00001105XF5027  00001105XF5028  00001134FMM115  00001201FMM0FK  00001201FMM0FM  00001201FMM0FX  00001201FMM0FY  00001203FMMONT  00001203FMM0MY  00001203FMM0NJ  00001203FMM0N8  00001208FMM03W  00001208FMM03X  00001211FMM0RG  00001211FMM0RM  00000299853MR2  00000300734MR1  00000301458MR6  00000301577MR3  00000301726MR6  00000301925MR4  00000302689MR5  00000302920MR4  00000303054MR1  00000294943MR6  00000295180MR4  00000295486MR5  00000296560MR6  00000301586MR4  00000302493MR2    2  00000291184MR0  00000293613MR6  00000296820MR4  00000297807MR0  00000299454MR9  00000300591MR5  00000301191MR3  00000005525YR1  00000005566YR5  00000006577YR1  00000006587YR0  00000006661YR3  00000007095YR3  00000007115YR9  00000008118YR2  00000008364YR2  00000008401YR2  00000008533YR2  00000008564YR7  00000009053YR0  00000009464YR9  00000009535YR6  00000009573YR7  00000009846YR7  00000010710YR2  00000010715YR1  00000010786YR2  00000010807YR6  00000011052YR8  00000011121YR1  00000011146YR8  00000011193YR0  00000011225YR0  00000011247YR4  00000011319YR1  00000011509YR7  00000011522YR0  00000011576YR6  00000011925YR5  00000012068YR3  00000012197YR0  00000012261YR4  00000012302YR6  00000028626MR7  00000239031MR8  00000241136MR1  00000251245MR7  00000255594MR4  00000263872MR4  00000270389MR0  00000271659MR5  00000273622MR1  00000276660MR8  00000278103MR7  00000278799MR2  00000283398MR6  00000283574MR2  00000289774MR2  00000289821MR1  00000293563MR3  00000293641MR7  00000293877MR7  00000293878MR5  00000293952MR8  00000294407MR2  00000294443MR7  00000294723MR2  00000295442MR8  00000295484MR0  00000296110MR0  00000297971MR4  00000298010MR0  00000298532MR3  00000298583MR6  00000298673MR5  00000298757MR6  00000299252MR7  00000299326MR9  00000299425MR9  00000299849MR0  00000300555MR0  00000301106MR1  00000838BD4EC9  00001011BD6643  00001101XF5001  00000010710YR2  00000009053YR0  00000007580YR4  00000009048YR0  00000009527YR3  00000009881YR4  00000012069YR1  00001008BD6541  00001052XF509B  00001208FMM03P  00000299418MR4  00000291345MR7  00000294545MR9  00000295757MR9  00001204FMM04H  00001118FMM0WP  00000294771MR1  00000296834MR5  00000008534YR0  00000008537YR3  00000010721YR9  00000010984YR3  00000012060YR0  00000238347MR9  00000283023MR0  00000291778MR9  00001117FMM0CT  00000008103YR4  00000273538MR9  00000276197MR1  00000283038MR8  00000289266MR9  00000297657MR9  00000298495MR3  00000831BD4A21  00000HDMR20150  00000010488YR5  00000010718YR5  00000011745YR7  00000296518MR4  00000297487MR1  00000297629MR8  00000297724MR7  00000297819MR5  00000298548MR9  00000299195MR8  00000299196MR6  00000299197MR4  00000299204MR8  00000010889YR4  00000011949YR5  00000028870MR1  00000232427MR5  00000260690MR3  00000261204MR2  00000281158MR6  00000282813MR5  00000287313MR1  00000295900MR5  00000297380MR8  00001002BD639E  00001208FMM03L  00000288909MR5  00000300561MR8  00000006697YR7  00000007527YR5  00000007558YR0  00000007894YR9  00000008102YR6  00000009417YR7  00000010437YR2  00000010513YR0  00000010877YR9  00000010916YR5  00000011051YR0  00000011119YR5  00000011893YR5  00000011924YR8  00000011930YR5  00000011933YR9  00000011935YR4  00000012120YR2  00000012128YR5  00000012259YR8  00000012319YR0  00000012342YR2  00000012386YR9  00001052XF5013  00000248176MR0  00000250436MR3  00000272433MR4  00000273369MR9  00000276459MR5  00000276585MR7  00000278621MR8  00000278748MR9  00000279448MR5  00000279915MR3  00000280004MR3  00000280858MR2  00000282008MR2  00000282511MR5  00000284023MR9  00000284279MR7  00000284642MR6  00000284739MR0  00000286601MR0  00000286613MR5  00000287550MR8  00000288821MR2  00000292467MR8  00000293048MR5  00000293292MR9  00000294004MR7  00000294145MR8  00000294146MR6  00000294161MR5  00000294341MR3  00000294434MR6  00000294659MR8  00000295028MR5  00000295217MR4  00000295808MR0  00000296147MR2  00000296215MR7  00000296216MR5  00000296232MR2  00000296235MR5  00000297417MR8  00000297699MR1  00000298224MR7  00000298311MR2  00000298466MR4  00000298844MR2  00000298931MR7  00000298971MR3  00000299663MR5  00000300885MR1  00000301350MR5  0000000B523YR3  00000005565YR7  00000011017YR1  00000012051YR9  00000012166YR5  00000290408MR4  00000930BD5D56  00000931BD5D8F  00001001BD6338  00001003BD641F  00001009BD6551  00001012BD66E0  00001050XF5012  00001052XF50A9  00001106XF5103  00000293730MR8  00000836BD4DB0  00000008362YR6  00000006584YR7  00000006658YR9  00000278668MR9  00000299272MR5  00000VALUE1176  00000283486MR9  00000284390MR2  00000000HM0427  00000011207YR8  00000297906MR0  00000300703MR6  00000950BD62AB  00000293710MR0  00000297155MR4  00000007093YR8  00000007509YR3  00000009607YR3  00000010624YR5  00000012076YR6  00000281467MR1  00000284686MR3  00000290193MR2  00000290255MR9  00000280779MR0  00000295131MR7  00000296406MR2  00000298332MR8  00000298665MR1  00000298676MR8  00000299261MR8  00000301850MR4  00000302543MR4  00000005222YR5  00000007515YR0  00000008366YR7  00000007134YR0  00000007518YR4  00000007566YR3  00000009442YR5  00000009461YR5  00000009533YR1  00000009581YR0  00000012066YR7  00000254189MR4  00000010788YR8  00000010818YR3  00000010920YR7  00000010948YR8  00000010962YR9  00000011071YR8  00000011445YR4  00000011520YR4  00000011546YR9  00000011888YR5  00000011911YR5  00000012044YR4  00000012164YR0  00000012251YR5  000000HDMR2838  000000HDMR2862  00000243246MR6  00000256205MR6  00000256211MR4  00000258409MR2  00000267466MR1  00000268303MR5  00000279400MR6  00000281683MR3  00000282582MR6  00000283698MR9  00000284592MR3  00000285351MR3  00000285589MR8  00000286265MR4  00000286894MR1  00000294419MR7  00000294624MR2  00000294636MR6  00000295507MR8  00000295508MR6  00000295522MR7  00000296146MR4  00000296403MR9  00000297372MR5  00000297628MR0  00000297659MR5  00000298071MR2  00000299664MR3  00000300939MR6  00000833BD4AEE  00000950BD62A0  00001101XF50CE  00000007118YR3  00000009450YR8  00000010636YR9  00000276243MR3  00000298489MR6  00001048XF501E  00001050XF501D  00001102XF502D  00001221FMM190  00000288156MR3  00000292292MR0  00000293440MR4  00001118FMM0WT  00001140FMM0FE  00000301824MR9  00000278105MR2  00000007570YR5  00000292844MR8  00000293042MR8  00000293327MR3  00000293919MR7  00000294575MR6  00000295847MR8  00000296032MR6  00000296141MR5  00000296227MR2  00000296332MR0  00000296499MR7  00000296814MR7  00000296879MR0  00000297164MR6  00000297468MR1  00000297494MR7  00000297636MR3  00000297701MR5  00000297715MR5  00000297861MR7  00000298073MR8  00000298242MR9  00000298385MR6  00000298430MR0  00000298567MR9  00000298734MR5  00000298850MR9  00000299341MR8  00000299464MR8  00000299633MR8  00000300589MR9  00000301125MR1  00000301291MR1  00000301868MR6  00000302711MR7  00000005228YR2  00000005234YR0  00000005527YR7  00000005534YR3  00000006718YR1  00000009446YR6  00000007089YR6  00000007101YR9  00000007107YR6  00000007505YR1  00000008323YR8  00000008400YR4  00000009030YR8  00000009047YR2  00000009083YR7  00000009454YR0  00000009549YR7  00000009565YR3  00000009572YR9  00000009756YR8  00000010066YR9  00000010142YR8  00000010282YR2  00000010288YR9  00000010632YR8  00000010971YR0  00000011027YR0  00000011101YR3  00000297446MR7  00000011120YR3  00000011148YR4  00000011162YR5  00000011235YR9  00000011237YR5  00000011281YR3  00000011327YR4  00000011336YR5  00000011403YR3  00000011414YR0  00000011514YR7  00000011543YR6  00000011582YR4  00000011725YR9  00000011732YR5  00000011736YR6  00000011800YR0  00000300886MR9  00000011929YR7  00000011945YR3  00000011998YR2  00000012000YR6  00000012007YR1  00000012018YR8  00000012062YR6  00000012064YR2  00000012122YR8  00000012235YR8  00000012243YR2  00000012305YR9  00000012321YR6  00000012322YR4  00000012345YR5  00000012379YR4  000000295001-3  000000HDMRXXXX  00000241872MR1  00000250442MR1  00000253158MR0  00000258091MR8  00000263339MR4  00000264618MR0  00000265869MR8  00000266467MR0  00000267327MR5  00000268121MR1  00000269325MR7  00000272922MR6  00000273833MR4  00000274073MR6  00000274836MR6  00000275538MR7  00000279165MR5  00000279616MR7  00000279727MR2  00000280922MR6  00000281867MR2  00000282338MR3  00000283122MR0  00000283350MR7  00000283508MR0  00000283772MR2  00000283933MR0  00000284238MR3  00000284648MR3  00000284757MR2  00000284901MR6  00000285496MR6  00000286286MR0  00000287964MR1  00000288043MR3  00000289773MR4  00000289820MR3  00000290544MR6  00000293318MR2  00000293344MR8  00000293524MR5  00000293643MR3  00000293656MR5  00000293681MR3  00000293683MR9  00000293970MR0  00000293999MR9  00000294114MR4  00000294117MR7  00000294343MR9  00000294529MR3  00000294552MR5  00000294661MR4  00000294735MR6  00000294919MR6  00000295182MR0  00000295193MR7  00000295804MR9  00000296305MR6  00000296440MR1  00000296833MR7  00000297444MR2  00000297465MR7  00000297660MR3  00000297809MR6  00000297834MR4  00000297851MR8  00000297853MR4  00000297892MR2  00000297970MR6  00000298070MR4  00000298222MR1  00000298281MR7  00000298808MR7  00000298929MR1  00000299526MR4  00000299661MR9  00000299763MR3  00000300125MR2  00000300673MR1  00000300726MR7  00000300750MR7  00000301138MR4  00000301139MR2  00001002BD6349  00001008BD651A  00001009BD655A  00001101XF50A8  000S0918BD5B3E  00000006567YR2  00000006686YR0  00000009850YR9  00000010620YR3  00000289782MR5  00000845BD51C9  00000918BD5B43  00000939BD5F65  00001012BD66F9  00001106XF50F6  00001105XF5017  00000293584MR9  00001211FMM0RR  00001219FMM0FL  00001223FMM09C  00000290985MR1  00000292928MR9  00001111FMM0B4  00001139FMM0D5  00001140FMM0FD  00001140FMM0G3  00001203FMM0UH  00001203FMM0VC  00001203FMM0V1  00001204FMM04M  00001204FMM049  00001219FMM0GK  00001219FMM0GP  00001111FMM0B3  00001111FMM0MX  00001118FMM0WB  00001134FMM10T  00001147FMM0EC  00000294397MR5  00000295485MR7  00000298431MR8  00000299453MR1  00000301912MR2  00000005966YR7  00000009010YR0  00000009552YR1  00000010714YR4  00000010970YR2  00000012199YR6  000000HDMR4304  00000266744MR2  00000269339MR8  00000282855MR6  00000283822MR5  00000286896MR6  00000293732MR4  00000298094MR4  00000300615MR2  00000008330YR3  00000294713MR3  00000006575YR5  00000295827MR0  00000282691MR5  00000289474MR9  00000294306MR6  00000298686MR7  00000299450MR7  00000296291MR8  00000294548MR3  00000006573YR0  00000006689YR4  00000008347YR7  00000008354YR3  00000011054YR4  00000011103YR9  00000011306YR8  00000011865YR3  00000228373MR7  00000229478MR3  00000238871MR8  00000265589MR2  00000265916MR7  00000279461MR8  00000280067MR0  00000294628MR3  00000297813MR8  00000298847MR5  00000007577YR0  00000011899YR2  00000011936YR2  00000259585MR8  00000286696MR0  00000294626MR7  00000296267MR8  00000297835MR1  00001201FMM0GL  00000300966MR9  00000298412MR8  00001211FMM0RK  00000263264MR4  00000282959MR6  00000298214MR8  00000298553MR9  00000299271MR7  00000299883MR9  00001002BD6381  00000297467MR3  00000298116MR5  00000299182MR6  00000299909MR2  00000300121MR1  00000300686MR3  00000300766MR3  00000300767MR1  00000300963MR6  00000301192MR1  00000301297MR8  00000301927MR0  00000010929YR8  00000011109YR6  00000011160YR9  00000008560YR5  00000011876YR0  00000011952YR9  00000012190YR5  00000010285YR5  00000257734MR4  00000289523MR3  00000010716YR9  00000293050MR1  00000010915YR7  00000293163MR2  00000293499MR0  00000011019YR7  00000293673MR0  00000011074YR2  00000011082YR5  00000011108YR8  00000011110YR4  00000298221MR3  00000011558YR4  00000011574YR1  00000298774MR1  00000011738YR2  00000011772YR1  00000011864YR6  00000299163MR6  00000299167MR7  00000011871YR1  00000299265MR9  00000299349MR1  00000011894YR3  00000011923YR0  00000011928YR9  00000011937YR0  00000300952MR9  00000301104MR6  00000914BD5A77  00001003BD6415  00000012125YR1  00000012165YR7  00000012183YR0  00000012184YR8  00000012186YR3  00001046XF501D  00000242552MR8  00000012270YR5  00000012317YR4  00000939BD5F52  00000012352YR1  00000012360YR4  00001001BD6333  000000HDMR5534  00000242552MR8  00000262664MR6  00000267973MR6  00000277190MR5  00000277214MR3  00000288874MR1  00000289039MR0  00000289315MR4  00000289591MR0  00000290654MR3  00000291970MR2  00000292591MR5  00000292982MR6  00000293297MR8  00000293298MR6  00000293586MR4  00000293593MR0  00000293650MR8  00000294051MR8  00000294284MR5  00000294547MR5  00000294920MR4  00000294928MR7  00000294929MR5  00000294936MR0  00000294944MR4  00000295020MR2  00000295059MR0  00000295219MR0  00000295528MR4  00000295529MR2  00000295561MR5  00000295828MR8  00000295878MR3  00000296094MR6  00000296439MR3  00000296442MR7  00000296458MR3  00000296468MR2  00000297014MR3  00000297202MR4  00000297385MR7  00000297420MR2  00000297474MR9  00000297658MR7  00000297725MR4  00000297728MR8  00000297806MR2  00000298155MR3  00000298314MR6  00000298328MR6  00000298329MR4  00000298330MR2  00000298467MR2  00000298659MR4  00000298672MR7  00000298680MR0  00000298759MR2  00000298793MR1  00000298858MR2  00000298911MR9  00000298928MR3  00000298934MR1  00000298935MR8  00000298969MR7  00000299164MR4  00000299165MR1  00000299191MR7  00000299201MR4  00000299269MR1  00000299275MR8  00000299311MR1  00000299328MR5  00000299448MR1  00000299506MR6  00000299507MR4  00000299643MR7  00000299677MR5  00000299946MR4  00000300123MR7  00000300183MR1  00000300617MR8  00000300729MR1  00000300747MR3  00000300859MR6  00000300887MR7  00000300922MR2  00000300949MR5  00000301010MR5  00000301168MR1  00000301187MR1  00000301264MR8  00000301383MR6  00000301392MR7  00000301397MR6  00000301498MR2  00000301762MR1  00000302188MR8  00000302426MR2  00000302474MR2  00000302532MR7  00000302643MR2  00000302679MR6  00000848BD5311  00000919BD5BB3  00000944BD6168  00000950BD629C  00000VALUE1162  00000VALUE1179  00000VALUE1280  00001002BD6360  00001003BD640C  00001008BD6539  00001009BD6558  00001009BD65DF  00001011BD6645  00001046XF501A  00001046XF501B  00001051XF5016  00001052XF506B  00001102XF5001  00001102XF503D  00000298250MR2  00000011314YR2  00000011590YR7  00000011866YR1  00000011997YR4  00000836BD4DB3  00000837BD4E13  00000851BD54EB  00000903BD55FC  00000905BD567A  00000293321MR6  00000918BD5B2B  00000919BD5BA8  00000919BD5BC2  00000938BD5F17  00000939BD5F69  00000950BD62BD  00001002BD6346  00001003BD6404  00001007BD6470  00001008BD652A  00001009BD65B0  00001009BD654A  00001010BD6606  00001011BD6685  00001012BD66B3  00001043XF5012  00001048XF500A  00001049XF5020  00001050XF5006  00001050XF5009  00001050XF501A  00001051XF5014  00001051XF5042  00001052XF50A3  00001052XF501C  00001052XF507A  00001052XF5098  00001101XF50A2  00001101XF5065  00001101XF509A  00001101XF509D  00001101XF5094  00001101XF5098  00001102XF503B  00001102XF5069  00001104XF50AB  00001104XF507C  00001104XF508E  00001104XF5080  00001104XF509D  00001105XF5019  00001113FMM1D3  00001134FMM114  00001134FMM116  00001142FMM0NE  00001142FMM0NY  00001142FMM0N5  00001201FMM0GE  00001201FMM0GT  00001201FMM0HP  00001203FMM0NF  00001203FMM0N6  00001203FMM0N9  00001208FMM03M  00001208FMM040  00001211FMM0RD  00001211FMM0T9  00001213FMM0LL  00001213FMM0L9  00001219FMM0FM  00001219FMM0FN  00000299492MR9  00000300451MR2  00000300768MR9  00000301459MR4  00000301753MR0  00000302245MR6  00000302246MR4  00000303964MR1  00000297408MR7  00000284356MR3  00000295483MR2  00000299184MR2  00000010924YR9  00001203FMM0N7  00000294119MR3  00000294601MR0  00000296200MR9  00000296265MR2  00000297850MR0  00000298717MR0  00000298951MR5  00000299547MR0  00000299628MR8  00000299686MR6  00000300037MR9  00000300297MR9  00000300986MR7  111  00000005548YR3  00000006574YR8  00000006691YR0  00000006698YR5  00000006715YR7  00000007090YR4  00000007099YR5  00000007522YR6  00000007530YR9  00000007535YR8  00000007547YR3  00000007555YR6  00000008096YR0  00000009415YR1  00000009545YR5  00000009595YR0  00000010162YR6  00000010289YR7  00000010720YR1  00000010903YR3  00000010954YR6  00000011081YR7  00000011889YR3  00000011944YR6  00000030BD4960  00000233440MR7  00000233623MR8  00000238361MR3  00000239043MR3  00000239230MR6  00000243412MR4  00000247728MR9  00000263292MR5  00000270630MR7  00000274168MR4  00000274932MR3  00000276461MR1  00000010900YR9  00000277779MR5  00000282692MR3  00000284491MR8  00000285119MR4  00000288849MR3  00000291415MR8  00000291846MR4  000002008AUG20  00000293954MR4  00000294918MR8  00000295763MR7  00000298581MR0  00000299202MR2  00000300553MR5  00000301499MR0  00000831BD49FA  00000837BD4E16  00000950BD629A  00001006BD6461  00001043XF5008  00001104XF5090  00000009588YR5  00000010392YR9  00000011776YR2  00000837BD4E18  00000006738YR9  00000007548YR1  00001007BD6498  00000300563MR4  00000300711MR9  00000303327MR1  00000291008MR1  00000291154MR3  00000295023MR6  00001147FMM0EA  00001203FMM0VA  00001204FMM051  00001219FMM0GW  00000030090MR1  00000010887YR8  00000011411YR6  00000295060MR8  00000297156MR2  00000297898MR9  00000298251MR0  00000298473MR0  00000299824MR3  0000030047 4MR4  00000300586MR5  00000300848MR9  00000301197MR0  000000008393Y1  00000006569YR8  00000006677YR9  00000006742YR1  00000007508YR5  00000007511YR9  00000007546YR5  00000007551YR5  00000007861YR8  00000007890YR7  00000008104YR2  00000008357YR6  00000008382YR4  00000008547YR2  00000009082YR9  00000009575YR2  00000009621YR4  00000009764YR2  00000010286YR3  00000010443YR0  00000010615YR3  00000010812YR6  00000010968YR6  00000010974YR4  00000011034YR6  00000011144YR3  00000011151YR8  00000011230YR0  00000011257YR3  00000011277YR1  00000011325YR8  00000011417YR3  00000011444YR7  00000011507YR1  00000011562YR6  00000011598YR0  00000011599YR8  00000011790YR3  00000011792YR9  00000011895YR0  00000011898YR4  00000011909YR9  00000012009YR7  00000012266YR3  00000012268YR9  00000012354YR7  00000028686MR4  00000252377MR7  00000252665MR5  00000008327YR9  00000261576MR3  00000261590MR4  00000266158MR5  00000278554MR1  00000280673MR5  00000280860MR8  00000281315MR2  00000282512MR3  00000285591MR4  00000286160MR7  00000286930MR3  00000011948YR7  00000293433MR9  00000029401MR1  00000294604MR4  00000294627MR5  00000294910MR5  00000295230MR7  00000295514MR4  00000296112MR6  00000296633MR1  00000296645MR5  00000296832MR9  00000297625MR6  00000298219MR7  00000298262MR7  00000298824MR4  00000298977MR0  00000277642MR5  00000283404MR2  00000299335MR0  00000299950MR6  00000300124MR5  00000300554MR3  00000300699MR6  00000300812MR5  00000302253MR0  00000910BD57E9  00000937BD5EF2  00000938BD5F16  00001002BD6343  00000296203MR3  00001043XF5005  00001007BD6496  00000007870YR9  00000008097YR8  00000280921MR8  00000285223MR4  00000836BD4DE2  00000903BD5606  00001003BD63FD  00001048XF5017  00001051XF502C  00001052XF50B0  00000948BD61E2  00001052XF5069  00001052XF5081  00001101XF5096  00001102XF5075  00001104XF50B1  00001104XF5089  00001142FMM0N9  00001048XF5027  00001102XF501A  00001102XF5010  00000303048MR3  00000302203MR5  00001007BD6492  00001012BD66C3  00001052XF5067  00001011BD6629  00000011196YR3  00000011331YR6  00000011580YR8  00000011714YR3  00000011783YR8  00000012057YR6  00000012071YR7  00000283517MR1  00000287773MR6  000S1001BD632F  00000011001YR1  00000950BD629D  00000294451MR0  00000294988MR1  00000012159YR0  00000300587MR3  00000299229MR5  00000008394YR9  00000009080YR3  00000298068MR8  00000919BD5B8E  00000293438MR8  00000295521MR9  00000297157MR0  00000298246MR0  00000302713MR3  00000302749MR7  00000007126YR6  00000007541YR6  00000007563YR0  00000007581YR2  00000008099YR4  00000008376YR6  00000009016YR7  00000009434YR2  00000009441YR7  00000009443YR3  00000009534YR9  00000009854YR1  00000010283YR0  00000010815YR9  00000010867YR0  00000010957YR9  00000011105YR4  00000011564YR2  00000011896YR8  00000012005YR5  00000233294MR8  00000246611MR8  00000256686MR7  00000264748MR5  00000265594MR2  00000268761MR4  00000277762MR1  00000283407MR5  00000283774MR8  00000284060MR1  00000284594MR9  00000286370MR2  00000286892MR5  00000293334MR9  00000293869MR4  00000295055MR8  00000295541MR7  00000912BD59B4  00000006557YR3  00000008317YR0  00000008351YR9  00000011134YR4  00000011803YR4  00000914BD5A88  00000930BD5D38  00003052527MR7  00000291429MR9  00001101XF50BB  00001117FMM0C7  00001134FMM10B  00000006717YR3  00000007521YR8  00000010879YR5  00000011093YR2  00000012019YR6  00000270465MR8  00000288222MR3  00000290369MR8  00000295003MR8  00000300776MR2  00000301223MR4  00001011BD6688  00001201FMMOHM  00001203FMM0NR  00000301047MR7  00000010148YR5  00000010709YR4  00000280857MR4  00000008340YR2  00000010799YR5  00000010987YR6  00000011283YR9  00000011728YR3  00000011739YR0  00000294047MR6  00000298098MR5  00000298755MR0  00001104XF5026  00000006574YR8  00000009026YR6  00000009591YR9  00000009599YR2  00000012041YR0  000000000R1740  00000010310YR1  00000011137YR7  00000011731YR7  00000011775YR4  00000010521YR3  00000010726YR8  00000300616MR0  00000300618MR6  00000931BD5D7B  00000948BD61EE  00001101XF5086  00001211FMM0T1  00000276632MR7  00000295513MR6  00001009BD65FE  00000290191MR6  00000914BD5A7E  00000937BD5EE2  00001002BD639A  00000299674MR2  00000300689MR7  00000297445MR9"
    }
  ]
}