  "results": [
    {
      "status": "Ongoing",
      "city": "Louvain-la-neuve",
      "state": "N/A",
      "country": "Belgium",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90213",
      "recalling_firm": "Ion Beam Applications S.A.",
      "address_1": "3 Chemin Du Cyclotron",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of IL, VA, WA and the country of Germany.",
      "recall_number": "Z-1205-2022",
      "product_description": "Proteus235- Proton Therapy System:  to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation",
      "product_quantity": "4 units",
      "reason_for_recall": "First layer of Uniform Scanning treatment fields is sometimes irradiated with incorrect Scanning Magnet setpoints resulting in the field size may be smaller than expected and the irradiated area may receive more dose than prescribed",
      "recall_initiation_date": "20220420",
      "center_classification_date": "20220527",
      "report_date": "20220608",
      "code_info": "UDI-DI: (01)05404013801138  Serial Numbers: PAT108 (US), PAT112 (US), SAT116 (US). PAT107 (EU)."
    }
  ]
}