  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82605",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide",
      "recall_number": "Z-1204-2019",
      "product_description": "Siemens Digital Linear Accelerators of type:  ARTISTE MV System  MEVATRON M2/Primus Mid-Energy PRIMUS  PRIMUS HI  ONCOR Impression  ONCOR Impression Plus  ONCOR Avant Garde  ONCOR Expression    ARTISTE, ONCOR, and PRIMUS running the following: Control Console from software version 13.0.302 and higher; Control Console from software version 9.2.400 and higher; and Control Console from software version 11.0.400 and higher.",
      "product_quantity": "120 US units",
      "reason_for_recall": "Advisory issued not to disable or bypass the AMP function. Disabling or bypassing the AMP function for reasons of comfort is not advisable may lead to patient or user injury due to collision with LINAC equipment disabling the  Auto Field Sequence (AFS) Automatic Motion Protection (AMP)  function implemented at the control of the Digital Linear Accelerator.",
      "recall_initiation_date": "20190320",
      "center_classification_date": "20190419",
      "termination_date": "20200618",
      "report_date": "20190501",
      "code_info": "SIEMENS Digital Linear Accelerators of type ARTISTE\", ONCOR\" and PRIMUS\" running:  - Control Console from software version  13.0.302 and higher  - Control Console from software version  9.2.400 and higher  - Controle Console from software version  11.0.400 and higher"
    }
  ]
}