  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Vantaa",
      "state": "",
      "country": "Finland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87309",
      "recalling_firm": "Thermo Fisher Scientific",
      "address_1": "Ratashe 2",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S.: Texas    O.U.S.: Belgium, France, Germany, Italy, Spain, Switzerland and United Kingdom",
      "recall_number": "Z-1202-2021",
      "product_description": "Cascadion SM Clinical Analyzer, REF 99990000 SN Type: 9910",
      "product_quantity": "19 systems (1 system in U.S. and 10 system O.U.S.)",
      "reason_for_recall": "Due to software defect, under certain assay parameters, false Vitamin D results may be reported.  The system reports a false result by not quantitating the correct analyte peak in sample chromatogram. The resulting sample would show a Vitamin D-concentration that is an unusual situation for human serum and plasma samples and would indicate a severe Vitamin D deficiency.",
      "recall_initiation_date": "20210201",
      "center_classification_date": "20210310",
      "termination_date": "20240110",
      "report_date": "20210317",
      "code_info": "Model Number 99990000  GTIN: 06438153004925  Serial Numbers:  CA100077,  CA000055,  CA100073,  CA100074,  CA000051,  CA100062,  CA100065,  CA100069,  CA100070  CA100063  CA000042,  CA100071  CA100060,  CA100059,  CA000058,  CA000053,  CA000056,  CA100061,  CA100066",
      "more_code_info": ""
    }
  ]
}