  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Elkton",
      "state": "MD",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76344",
      "recalling_firm": "Terumo Cardiovascular Systems Corporation",
      "address_1": "125 Blue Ball Rd",
      "address_2": "N/A",
      "postal_code": "21921-5315",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Terumo   Cardiovascular   has  identified   16  domestic   consignees   and  30   international consignees .",
      "recall_number": "Z-1201-2017",
      "product_description": "Hercules 360 Universal Stabilizing Arm, top mount, 360-degree rotation, reusable stainless steel",
      "product_quantity": "The total number of distributed devices stands at 260",
      "reason_for_recall": "Terumo Cardiovascular Systems (Terumo CVS) is recalling certain lots of Hercules TM 360 Universal Stabilizing Arms because the two pins that hold the sleeve attached to the distal end of the device may come loose, causing the pins and sleeve to separate from the device. While the arm can still function without the pins and sleeve in place, there is potential for the loose components to fall into the surgical site.",
      "recall_initiation_date": "20140820",
      "center_classification_date": "20170214",
      "termination_date": "20170215",
      "report_date": "20170222",
      "code_info": "Serial Number Range 70194 through 70504",
      "more_code_info": ""
    }
  ]
}