      "status": "Terminated",
      "city": "Hillsboro",
      "state": "OR",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76332",
      "recalling_firm": "Acumed LLC",
      "address_1": "5885 NW Cornelius Pass Rd",
      "address_2": "N/A",
      "postal_code": "97124-9432",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - Costa Rica, Malaysia, Mexico, South Africa, and Spain",
      "recall_number": "Z-1199-2017",
      "product_description": "6-hole Pubic Symphysis plate (PN 70-0451).  The pelvic bone plates for the pelvic ring are a series of plates with varying lengths that function as internal fixation devices for fractures, fusions and osteotomies of the pelvic ring. The plates are strategically pre-contoured where beneficial to the application, and they are secured to the bone with  3.5mm non-locking screws. The Acumed pelvic ring plate is the Pubic Symphysis Plate.      The Acumed Pelvic Plating System is intended for use by surgeons with orthopedic training and knowledge of the indications and techniques required for fixation. The device is to be implanted by the surgeon in a sterile operating room setting.",
      "product_quantity": "15 units",
      "reason_for_recall": "The 6-hole Pubic Symphysis plate (PN 70-0451) is specified to be made of Commercially Pure Titanium (ASTM F67 - CP Grade 4). However, batch 383813 was manufactured out of Titanium Alloy (Ti-6Al- 4V ELI per ASTM F135 \u0013 Ti alloy) and was distributed.",
      "recall_initiation_date": "20170119",
      "center_classification_date": "20170214",
      "termination_date": "20170508",
      "report_date": "20170222",
      "code_info": "Lot number: 383813",
      "more_code_info": ""
    }
  ]
}