  "results": [
    {
      "country": "United States",
      "city": "Atlanta",
      "address_1": "400 Perimeter Center Ter Ne",
      "reason_for_recall": "Dose and Monitor Unit values are not computed correctly when Elekta Motorized Wedges are used with Dynamic Conformal arcs.",
      "address_2": "Ste 50",
      "product_quantity": "1776",
      "code_info": "Versions 4.1 and higher",
      "center_classification_date": "20140313",
      "distribution_pattern": "USA including AL, AK, AZ, CA, CO, CT, DE, DC, FL, GA, IL, IN, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NJ, NY, NC, ND, OH, OK, PA, RI, SC, TN, TX, VA, WA, WV, WI, WY, Puerto Rico and Internationally to  Albania, Algeria, Argentina, Australia, Austria, Bahamas, Barbados, Bermuda, British Guiana, Boliva, Brazil, British West Indies, Burundi, Canada, Chile, China, Columbia, Costa Rica, Cuba, Curacao, Cyprus, Czechoslovakia, Dominica, Equator, Estonia, Egypt, Finland, France, Gambia, Germany, Great Britain, Georgia, Greece, Guatemala, Hong Kong, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lativa, Lithunia, Madagascar, Malaysia, Malta, Mexico, Mongolia, Myanmar, Netherlands, New Zealand, Nigeria, Pakistan, Peru, Philippines, Poland, Portugal, Romania, Russia, San Salvador, Saudia Arabia, Serbia, Singapore, Sir Lanka, South Korea, Spain, Sudan Slovenia, Suriname, Tajikistan, Thailand, Turkey, Turkmenistan, Taiwan, Ukraine, Uruguay, Venezuela, Viet Nam, Zambia.",
      "state": "GA",
      "product_description": "XiO RPT System    The XiO RTP System is used to create treatment plans for any cancer patient for whom external beam radiation therapy or brachytherapy has been prescribed. The system will calculate and display, both on screen and in hard-copy, either two or three dimensional radiation dose distributions within a patient for a given treatment.",
      "report_date": "20140319",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Elekta, Inc.",
      "recall_number": "Z-1197-2014",
      "initial_firm_notification": "Telephone",
      "product_type": "Devices",
      "event_id": "67625",
      "termination_date": "20150721",
      "recall_initiation_date": "20140217",
      "postal_code": "30346-1227",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}