      "status": "Terminated",
      "city": "Osseo",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67578",
      "recalling_firm": "Gyrus Medical, Inc",
      "address_1": "6655 Wedgwood Rd N Ste 160",
      "address_2": "N/A",
      "postal_code": "55311-3613",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution.",
      "recall_number": "Z-1196-2014",
      "product_description": "Gyrus ACMI Disposable Falope-Ring Band Applicator Kits.  REF 006889-901 and REF 005280-901 (with 8 mm Disposable Trocan).  Sterile EO.  Single Use Only.    The disposable Falope-Ring Band Applicator is a sterile, two-band, dual-incision instrument.  The two-band applicator enables the physician to occlude both Fallopian tubes without removing the applicator from the abdominal cavity.  The Falope-Ring Band and Applicator Kits are indicated for use for female sterilization (permanent contraception).",
      "product_quantity": "856 units",
      "reason_for_recall": "Due to an anomaly in the packaging process for specific lots of Disposable Falope-Ring Band Kit, it is possible that a defective seal could be present that may allow a breach of the package's sterile barrier and may compromise the sterility of the product.  The breach may or may not be easily seen.",
      "recall_initiation_date": "20140219",
      "center_classification_date": "20140313",
      "termination_date": "20150414",
      "report_date": "20140319",
      "code_info": "Lots:  MK761655, MK764022, MK767655, MK526401"
    }
  ]
}