  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Raritan",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67447",
      "recalling_firm": "Veridex, LLC",
      "address_1": "1001 US Route 202",
      "address_2": "N/A",
      "postal_code": "08869-0606",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA Nationwide",
      "recall_number": "Z-1195-2014",
      "product_description": "CELLTRACKS ANALYZER II¿ (Product Code 9555)    Product Usage:    The CELLTRACKS ANALYZER II¿ is a semi-automated fluorescence microscope, consisting of the analyzer, a dedicated computer with CELLTRACKS¿ software, monitor, keyboard, mouse and uninterruptible power supply (UPS). The system also supports an optional Remote Review Workstation (RRW), which consists of a dedicated computer with CELLTRACKS¿ software, monitor, keyboard and mouse. Use of this product requires training and should be used under the supervision of laboratory management. The CELLTRACKS ANALYZER II¿ is for analysis of rare cells that are isolated from biological fluids including whole blood. It is used in conjunction with the CELLTRACKS¿ AUTOPREP¿ System, which automates and standardizes the sample preparation with specific reagent kits. An optional Remote Review Workstation is also available to provide the capability to review images and report results remotely.",
      "product_quantity": "19 units",
      "reason_for_recall": "Janssen Diagnostics, LLC received complaints of camera failures on the CELLTRACKS ANALYZER II¿.",
      "recall_initiation_date": "20140108",
      "center_classification_date": "20140312",
      "termination_date": "20170706",
      "report_date": "20140319",
      "code_info": "First: K050145, cleared 15 March 2005 Latest: K130794, cleared 20 June 2013  9555    Serial numbers:     CT1106005  CT1104002  CT1108006  CT1111009  CT1201001  CT1203004  CT1205011  CT1205012  CT1206013  CT1210022  CT1210024"
    }
  ]
}