  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Irvine",
      "state": "CA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82456",
      "recalling_firm": "Edwards Lifesciences, LLC",
      "address_1": "1 Edwards Way",
      "address_2": "N/A",
      "postal_code": "92614-5688",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide :  FL, MD, NY, TX, VA, CO, OK, MO, CA, DE, TN, NC, DC, PA, MA, OR, SC, OH, WI, KS, NV, MI, IL, KY, GA, NJ, WA, CT, IN, MN, WV, LA, MS, AL, SD, NE, IA, AZ, ID, HI, UT, ME,MT, NM, AK, NH, RI, AR, ND, WY, PR.     Foreign (OUS):  IT, FI, CO, KR, JP, PL, BE, GR, NL, IL, AT, DK,CA, GB, BR, SA, SE, DE, AU, FR, MX, IN, IC, ZA, ES, TR, NZ, MY, IE, SG, CZ, SI, PT, CH, RE, LK, CR, AR, HK, NO, PY, PA, MT, EG, NP, TN, RU, RO, NA,CL, BH, IR, PH, TH, KZ, DZ, BG, EE, KW, VN, OM, PE, EC, AE, MA, VE, BO, QA, DO, GT, BY, JO, LT, SK, HU, CY, MN, TW, CN",
      "recall_number": "Z-1193-2019",
      "product_description": "EV1000 Clinical Platform, Models: EV1000A, EV1000NI, EV1000CS    Product Usage:  EV1000A Monitor is used to optimize fluid management through the use of minimally invasive monitoring with the FloTrac sensor (or the Volume View sensor).",
      "product_quantity": "22443",
      "reason_for_recall": "Potential for an electrical short circuit leading to possible fire hazard due to liquid ingress into AC power outlets.",
      "recall_initiation_date": "20190321",
      "center_classification_date": "20190521",
      "termination_date": "20231128",
      "report_date": "20190529",
      "code_info": "All lots"
    }
  ]
}