      "status": "Terminated",
      "city": "Blaine",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73150",
      "recalling_firm": "QiG Group LLC",
      "address_1": "10675 Naples St NE",
      "address_2": "N/A",
      "postal_code": "55449-5802",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distributed in Germany and Luxembourg.",
      "recall_number": "Z-1193-2016",
      "product_description": "Algostim LLC, Algovita Implanted Pulse Generator stimulator, Model 2412 ( 2 x 12 channel)",
      "product_quantity": "34",
      "reason_for_recall": "The implantable pulse generator (IPG) model 2408  had suboptimal performance establishing a connection to the external devices unless they were brought into close proximity (as close as 5 cm).  The following functions may not be available:  Reprogramming the stimulator, Adjusting stimulation parameters (on/off, intensity modifications, stimulation program selection), and Recharging the stimulator.",
      "recall_initiation_date": "20160126",
      "center_classification_date": "20160318",
      "termination_date": "20170405",
      "report_date": "20160330",
      "code_info": "Serial Numbers:  03EDB5, 03EDD9, 03EDDA, 03EDDB, 03EDDD, 03EDE0, 03EDE4, 03EDF5, 03EDF6, 03EDFA, 03EDFC, 03EDFF, 03EE07, 03EE08, 03EE0A, 03EE0D, 03EE0E, 03EE12, 03EE13, 03EE1A, 03EE30, 03EE3C, 03EE3F, 03EE44, 03EE48, 03EE4C, 03EE4D, 03EE6D, 03EE74, 03EE82, 03EE88, 03EE89, 03EE8F, 03EE94."
    }
  ]
}