{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Blaine",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73150",
      "recalling_firm": "QiG Group LLC",
      "address_1": "10675 Naples St NE",
      "address_2": "N/A",
      "postal_code": "55449-5802",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distributed in Germany and Luxembourg.",
      "recall_number": "Z-1192-2016",
      "product_description": "Algostim LLC, Algovita Implanted Pulse Generator stimulator, Model 2408 ( 3 x 8 channel)",
      "product_quantity": "40",
      "reason_for_recall": "The implantable pulse generator (IPG) model 2408  had suboptimal performance establishing a connection to the external devices unless they were brought into close proximity (as close as 5 cm).  The following functions may not be available:  Reprogramming the stimulator, Adjusting stimulation parameters (on/off, intensity modifications, stimulation program selection), and Recharging the stimulator.",
      "recall_initiation_date": "20160126",
      "center_classification_date": "20160318",
      "termination_date": "20170405",
      "report_date": "20160330",
      "code_info": "Serial Numbers:  03EDBF, 03EDC1, 03EDC5, 03EDC7, 03EDC9, 03EDCA, 03EDCE, 03EDD0, 03EDE5, 03EDE8, 03EDE9, 03EDEA, 03EDEF, 03EDF0, 03EE01, 03EE03, 03EE05, 03EE1C, 03EE1F, 03EE24, 03EE25, 03EE26, 03EE28, 03EE34, 03EE35, 03EE36, 03EE37, 03EE3A, 03EE3B, 03EE5A, 03EE5B, 03EE60, 03EE96, 03EE97, 03EE98, 03EE9D, 03EE9F, 03EEA2, 03EEA3, 03EEA4, 03EEAC."
    }
  ]
}