  "results": [
    {
      "status": "Terminated",
      "city": "Glens Falls",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70322",
      "recalling_firm": "Navilyst Medical, Inc",
      "address_1": "10 Glens Falls Technical Park",
      "address_2": "N/A",
      "postal_code": "12801",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.",
      "recall_number": "Z-1192-2015",
      "product_description": "NAMIC Convenience Kit, LEFT HEART KIT, UPN H965600125911, REF/Catalog No. 60012591, STERILE, Rx ONLY ---For single use only. --- Device Listing Number D136588.     NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.",
      "product_quantity": "2 units",
      "reason_for_recall": "The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.",
      "recall_initiation_date": "20150109",
      "center_classification_date": "20150225",
      "termination_date": "20170310",
      "report_date": "20150304",
      "code_info": "Lot 4440346",
      "more_code_info": ""
    }
  ]
}