  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Selzach",
      "state": "N/A",
      "country": "Switzerland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91666",
      "recalling_firm": "Stryker GmbH",
      "address_1": "Bohnackerweg 1",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-1191-2023",
      "product_description": "stryker Hoffmann LRF, Transport Strut, Model Number 4933-0-400; Orthopedic external fixation device intended to keep bone fragments/segments in a correct anatomical position until reliable bone consolidation is achieved",
      "product_quantity": "488 units",
      "reason_for_recall": "Stryker has identified a nonconformance in specific lots of Hoffmann LRF Bone Transport Struts. Specifically, the thread pitch is oversized at the distal end of the threaded rod, and as a result the distal ring fixation cannot be locked between 0- 50mm of the threaded rod.",
      "recall_initiation_date": "20230207",
      "center_classification_date": "20230301",
      "report_date": "20230308",
      "code_info": "Serial numbers:  D32158, D32160, D32161, D32162, D32163, D32164, D34343, D34344, D40690, D41629, D41630, D41631, D44713, G16331, G16332, G16333, G26278, G29759, G29760, G32730, G32731, G32735, G45119, H30398, H30399, H66392",
      "more_code_info": ""
    }
  ]
}