  "results": [
    {
      "status": "Terminated",
      "city": "Brooklyn Park",
      "state": "MN",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87271",
      "recalling_firm": "Medtronic Perfusion Systems",
      "address_1": "7611 Northland Dr N",
      "address_2": "N/A",
      "postal_code": "55428-1088",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution: US nationwide, Afghanistan,  Argentina,  Chile,  China,  Colombia,  Egypt,  Germany,  Iraq,  Italy,  Japan,  Korea,  Kyrgyzstan,  Latvia,  Nicaragua,  Paraguay,  Poland,  Qatar,  Romania,  Tanzania,  United Arab Emirates,  United Kingdom,  Uzbekistan, and Viet Nam",
      "recall_number": "Z-1191-2021",
      "product_description": "Medtronic Affinity Pixie BP241 Hollow Fiber Oxygenator and Cardiotomy/Venous Reservoir with Balance Biosurface (OXYGENATOR BBP241 W/CVR NI BB STERILE).  The Affinity Pixie Cardiotomy/Venous Reservoir (CVR) with Balance Biosurface is a single-use device designed to collect and store blood during extracorporeal  circulation. Cardiotomy blood is collected, filtered, and defoamed before mixing with the venous blood, which is filtered. The primary blood-contacting surfaces of the CVR are coated with Balance Biosurface to reduce platelet activation and adhesion and preserve platelet function. (Generic Name: Cardiopulmonary bypass oxygenator).",
      "product_quantity": "57 units",
      "reason_for_recall": "Products being recalled due to potentially elevated level of bacterial endotoxin. Patients exposed to elevated levels of bacterial endotoxins may develop an acute inflammatory response.",
      "recall_initiation_date": "20210204",
      "center_classification_date": "20210317",
      "termination_date": "20240426",
      "report_date": "20210324",
      "code_info": "Description: OXYGENATOR BBP241 W/CVR NI BB STERILE.  Model Number: BBP241. GTIN: 20613994617303. Lot Number: 13340434."
    }
  ]
}