      "status": "Terminated",
      "city": "Austin",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76209",
      "recalling_firm": "Encore Medical, Lp",
      "address_1": "9800 Metric Blvd",
      "address_2": "N/A",
      "postal_code": "78758-5445",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide - US Nationwide in the states of MS, OH, RI, NY, OK, TX, CA, HI",
      "recall_number": "Z-1191-2017",
      "product_description": "LIMA Modular Revision Hip Stem Model 428-01-050_110    Product Usage:  The Modular Revision Femoral hip stem is made up of a modular stem coupled with a proper neck by means of a Morse taper stabilized during the implantation phase by a safety screw. This system is particularly indicated for revision surgery on both uncemented and cemented femoral implants, when there is significant bone loss and an abnormal meta-epiphyseal anatomy of the femur.",
      "product_quantity": "13 devices",
      "reason_for_recall": "Lima Proximal Bodies were inadvertently re-sterilized. The safety screw which affixes the distal and proximal bodies of the stem includes a thread-locking plug made from UHMWPE, which is not approved for repeated gamma sterilization.",
      "recall_initiation_date": "20170110",
      "center_classification_date": "20170210",
      "termination_date": "20170331",
      "report_date": "20170222",
      "code_info": "1005528A, 1006882A, 1107305A, 1101798A, 1100565A, 1005531A, 1211716A, 1007408A, 1004038B, 1005533D, 1106805A, 1203471A, 1005533C, 1100565A, 1101795A, 1007079A, 1005532A, 1100560A",