{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Alpharetta",
      "state": "GA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91663",
      "recalling_firm": "Avanos Medical, Inc.",
      "address_1": "5405 Windward Pkwy",
      "address_2": "N/A",
      "postal_code": "30004-3894",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Visit",
      "distribution_pattern": "US Nationwide distribution in the state of Oklahoma.",
      "recall_number": "Z-1190-2023",
      "product_description": "BALLARD* ACCESS Closed Suction System for Neonates/Pediatrics, Elbow Manifold.",
      "product_quantity": "220 units",
      "reason_for_recall": "Potential for product manifold to crack during use on patients, possibly causing leaks in the respiratory circuit and can lead to the patient receiving inadequate ventilation.",
      "recall_initiation_date": "20230203",
      "center_classification_date": "20230313",
      "termination_date": "20240905",
      "report_date": "20230322",
      "code_info": "All lots are affected.  Stock Code (Lot Number): B50511 (20098277), B50611 (20098278), B50711 (20098279), B50811 (20097177, 20097177R1)."
    }
  ]
}