      "status": "Terminated",
      "city": "Franklin Lakes",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76295",
      "recalling_firm": "Becton Dickinson & Company",
      "address_1": "1 Becton Dr",
      "address_2": "N/A",
      "postal_code": "07417-1815",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide and the countries of Canada, Mexico, Indonesia",
      "recall_number": "Z-1189-2017",
      "product_description": "BD Vacutainer  Buffered Sodium Citrate: (9NC) Blood Collection Tube (4.5 ml x 13 x 75 mm)  Catalog Number: 369714    Product Usage:  The BD Vacutainer Safety Coagulation tube is an evacuated blood collection tube that provides means of collecting, transporting and processing blood in a closed tube. The buffered sodium citrate additive provides an anti-coagulated specimen that may be used for clinical laboratory coagulation assays and determinations",
      "product_quantity": "1,488,800 tubes",
      "reason_for_recall": "A portion of this lot of blood collection tubes was manufactured with approximately twice the expected amount of buffered sodium citrate additive, an anticoagulant.",
      "recall_initiation_date": "20170116",
      "center_classification_date": "20170210",
      "termination_date": "20180207",
      "report_date": "20170222",
      "code_info": "Lot 6090572",
      "more_code_info": ""
    }
  ]
}