  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76378",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution- US (nationwide) including states of: CA, DC, IL, MN, MO, NY, SC, TX, WI, WV; and countries of: Canada, Czech Republic, Denmark, Finland, France, Germany, Italy, Netherlands, Norway, Poland, Saudi Arabia, Spain, Sweden, Switzerland, Taiwan, United Kingdom, and Uruguay.",
      "recall_number": "Z-1188-2017",
      "product_description": "Discovery MR450    The systems are whole body magnetic resonance scanners designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps. and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate. blood vessels, and musculoskeletal regions of the body.",
      "product_quantity": "60",
      "reason_for_recall": "Potential safety issue with the patient bore heating on the Discovery MR450 narrow bore 1.5T MRI. In the event of a RF component failure, typically happening after extended system use, a small area on the bore surface may become warmer than normal. If this happens in an area where there is patient contact with the bore, and proper padding is not used, it may cause a serious burn.",
      "recall_initiation_date": "20170120",
      "center_classification_date": "20170210",
      "termination_date": "20190924",
      "report_date": "20170222",
      "code_info": "Mfg. Lot or Serial # System ID  00001106XF509E 209225VC450  00001106XF50A6 858939MR4  000000000R7216 858939MR3  000000000R2645 202715MR2  00000842BD50C2 202476MR1  000000000R7025 217383CMR5  00000944BD6155 MR129036  101BD66B6  573884MRI  00000945BD6182 417COXMR1  000000000R2768 HSSFMR450  00001009BD65C1 HSS75MR450  00000944BD6166 803434MRI  00000298722MR0 803434MR2  00000300039MR5 832522MR2  Not Available 713566SMITH1  1106xf508e                      713566SMITH3  00000296429MR4 281364MR2  00000833BD4AE7 920830MR450  00000933BD5DF8 304256MR2  00000294451MR0 030020MR03  000000000R7320 403541MR1  00000944BD616B 403955MR1  000000000R7149 250727MR2  Not Available 905522MR2  00001101XF5028 514281MR450  00000297813MR8 CS1006MR04  00000294548MR3 DK1083MR06  00000294397MR5 FI1128MR01  00000295485MR7 FI1052MR01  00000296032MR6 A2032112  00000295847MR8 M4182438  00000293919MR7 M206565701  00000294575MR6 M40330108  00000298332MR8 RMUE01MR04  0002001-0917-3 A5127236  00000304354MR4 A5808528  00000296820MR4 A5624132  00000297807MR0 D5354206  00000293603MR7 RTD0127  DUMFMI60898001 NO1017MR03  00000297987MR0 NO1009MR02  00000296493MR0 PL1539MR01  00000300882MR8 SA1058MR02  00000296613MR3 SA1011MR03  00000299452MR3 CPZ43620  00000303993MR0 M001MR01  00000293000MR6 A010MR06  00000293920MR5 N002MR02  00000303992MR2 541010MR03  00000300849MR7 0853270113  00000292973MR5 00242MRS02  00000293206MR9 70001MRS09  00000298041MR5 05217MRS02  00000293642MR5 10448MRS02  00000293724MR1 10050MRS03  00000299340MR0 00010MRS01  00000295709MR0 00200MRS04  00000294967MR5 00100MRS01  00000296514MR3 00105MRS02  00000300689MR7 2948529MR",
      "more_code_info": ""
    }
  ]
}