  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Waukesha",
      "address_1": "3000 N Grandview Blvd",
      "reason_for_recall": "Fastener bolts on the heat exchanger were not properly torqued and part of the heat exchanger can detach from the gantry and hit the inside of the covers and other components of the gantry possibly resulting in the oil conduit breaking, spilling oil within the gantry and outside the system. This issue affects GE Healthcare VCT for Discovery VCT, Discovery RX VCT, and Discovery PET/CT 690, Lightspe",
      "address_2": "",
      "product_quantity": "166",
      "code_info": "Serial Number  00000416307CN7  00000416534CN6  00000418451CN1  00000416467CN9  00000414273CN3  00000416594CN0  00000416589CN0  00000416532CN0  00000417617CN8  00000416889CN4  00000416685CN6  00000418151CN7  00000416332CN5  00000416591CN6  00000416468CN7  00000416333CN3  00000419445CN2  00000416269CN9  00000416815CN9  00000416117CN0  00000416524CN7  00000416267CN3  00000419240CN7  00000416646CN8  00000417864CN6  00000416195CN6  00000419505CN3  00000418001CN4  00000416816CN7  00000416551CN0  00000416201CN2  00000417625CN1  00000417650CN9  00000416220CN2  00000419411CN4  00000416390CN3  00000418913CN0  00000416737CN5  00000416738CN3  00000416777CN1  00000416363CN0  00000416888CN6  00000416590CN8  00000417177CN3  00000419041CN9  00000416850CN6  00000416627CN8  00000417275CN5  00000417709CN3  00000417133CN6  00000419446CN0  00000419375CN1  00000418807CN4  00000418035CN2  00000418238CN2  00000418843CN9  00000417786CN1  00000419393CN4  00000417695CN4  00000419507CN9  00000419078CN1  00000419280CN3  00000419390CN0  00000419526CN9  00000419525CN1  00000419559CN0  00000418819CN9  00000419506CN1  00000418402CN4  00000419089CN8  00000419550CN9  00000416736CN7  00000418656CN5  00000416781CN3  00000416798CN7  00000418357CN0  00000419447CN8  00000416552CN8  00000418810CN8  00000416764CN9  00000419178CN9  00000417719CN2  00000416389CN5  00000418514CN6  00000417634CN3  00000416848CN0  00000417733CN3  00000417716CN8  00000416849CN8  00000417649CN1  00000416286CN3  00000416401CN8  00000418847CN0  00000419177CN1  00000416533CN8  00000417644CN2  00000419376CN9  00000419504CN6  00000416231CN9  00000416301CN0  00000417626CN9  00000418403CN2  00000419079CN9  00000416268CN1  00000416620CN3  00000416597CN3  00000416763CN1  00000418780CN3  00000417643CN4  00000416647CN6  00000417642CN6  00000418791CN0  00000416466CN1  00000416388CN7  00000416353CN1  00000417134CN4  00000419241CN5  00000418915CN5  00000419174CN8  00000418818CN1  00000416295CN4  00000417720CN0  00000418846CN2  00000416372CN1  00000416762CN3  00000417714CN3  00000416882CN9  00000417713CN5  00000417725CN9  00000416595CN7  00000418020CN4  00000417734CN1  00000419448CN6  00000417735CN8  00000417787CN9  00000419054CN2  00000417696CN2  00000418932CN0  00000419173CN0  00000418300CN0  00000417717CN6  00000418845CN4  00000419561CN6  00000418696CN1  00000418848CN8  00000366477CN8  00000220652HM2  00000421152CN0  00000419040CN1  00000418844CN7  00000416202CN0  00000419391CN8  00000416371CN3  00000418781CN1  00000416814CN2  00000416593CN2  00000416200CN4  00000416739CN1  00000416465CN3  00000416670CN8  00000416881CN1  00000416525CN4  00000416284CN8  00000416796CN1  00000418943CN7  00000419039CN3",
      "center_classification_date": "20130429",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) including DC and the states of AZ, CA, CO, CT, FL,GA, IL, IN, KY, MA, MI, MS,MO, NH, NJ, NM, NY, NC, OK, OR, PA, SC, SD, TX, UT, VA, WA, and WI and the countries of ALBANIA ARGENTINA AUSTRALIA BELGIUM BRAZIL CANADA CHINA COLOMBIA COSTA RICA CZECH REPUBLIC DENMARK ECUADOR EGYPT FINLAND FRANCE GREECE HONG KONG HUNGARY INDIA IRELAND ITALY JAPAN KOREA, REPUBLIC OF MEXICO MOROCCO NIGERIA NORWAY POLAND RUSSIAN FEDERATION SINGAPORE SPAIN SWEDEN SWITZERLAND TURKEY UKRAINE UNITED KINGDOM.",
      "state": "WI",
      "product_description": "GE, VCT for Discovery VCT, Discovery RX VCT, and Discovery PET/CT 690, Lightspeed VCT, Discovery CT750HD.    Intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission.",
      "report_date": "20130508",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "GE Healthcare, LLC",
      "recall_number": "Z-1188-2013",
      "initial_firm_notification": "Visit",
      "product_type": "Devices",
      "event_id": "64889",
      "termination_date": "20140204",
      "more_code_info": "",
      "recall_initiation_date": "20110725",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}