  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Irvine",
      "address_1": "Forty Parker",
      "reason_for_recall": "Masimo has identified a very small number (0.02%) of Rad-8 devices with 20-pin patient cable connector that, when moved, can power off without the operator pressing the power button.",
      "address_2": "",
      "product_quantity": "42,908 units",
      "code_info": "Catalog Number  9190  9193  9191  9192  9212  9213  9218  9219  9224  9282  9285  2883  9194",
      "center_classification_date": "20130429",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) including the countries of  Africa, Asia, Australia/New Zealand, Europe, Canada, Latin America, and Middle East / SW Asia.",
      "state": "CA",
      "product_description": "Rad-8 Pulse Oximeters     The Rad-8 family of pulse oximeters and accessories are indicated for the continuous, noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor).",
      "report_date": "20130508",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Masimo Corporation",
      "recall_number": "Z-1187-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "64812",
      "termination_date": "20150610",
      "more_code_info": "",
      "recall_initiation_date": "20130313",
      "postal_code": "92618",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}