{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Munich",
      "state": "",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87395",
      "recalling_firm": "Brainlab AG",
      "address_1": "Olof-palme-str. 9",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "USA: (CA, NJ, AZ, AR, NY, IN, OR, NC), Australia, Austria, Belgium, Denmark, France, Germany, Hong Kong, Italy, Slovakia, Spain, Thailand, Turkey, United Kingdom, United Arab Emirates.",
      "recall_number": "Z-1185-2021",
      "product_description": "ExacTrac Dynamic (Medical Charged-Particle Radiation Therapy System).",
      "product_quantity": "73 ExacTrac Dynamic systems (versions 1.0.0, 1.0.1, 1.0.2) have been manufactured and distributed",
      "reason_for_recall": "In case of a failed automatic marker detection, a software error causes parts of the  display to incorrectly behave as if the current patient position is within predefined tolerances  and may allow the user to proceed to treatment despite potentially exceeding shift values.",
      "recall_initiation_date": "20210301",
      "center_classification_date": "20210308",
      "termination_date": "20230320",
      "report_date": "20210317",
      "code_info": "Model/catalogue numbers: 20910-06 ETD Implanted lic.  UDI: The GTIN for ExacTrac Dynamic version 1.0 is 04056481142315.  GMDN code: 40887.  Software revisions of affected ExacTrac Dynamic software: versions 1.0.0, 1.0.1, 1.0.2.",
      "more_code_info": ""
    }
  ]
}