      "country": "United States",
      "city": "Mentor",
      "address_1": "5960 Heisley Rd",
      "reason_for_recall": "The firm discovered that one of the hysteroscope models listed for processing in four existing Quick Connects has one inlet port rather than two inlet ports as indicated in the Quick Connect processing instructions.",
      "address_2": "",
      "product_quantity": "809 units",
      "code_info": "All Lots. Model numbers: QKC1736E, QKC1736INT, QKC1737E, and QKC1737INT.",
      "center_classification_date": "20160316",
      "distribution_pattern": "Worldwide distribution.  US nationwide, Australia, Bahrain, Bulgaria, Canada, Greece, Italy, Malaysia, Mexico, Qatar, Russia, Spain, and United Kingdom.",
      "state": "OH",
      "product_description": "Quick Connects. Steris Corporation.  Model numbers: QKC1736E, QKC1736INT, QKC1737E, and QKC1737INT. Each Quick Connect contains a flow unit consisting of tubing, tethers, and adapters. For sterilant and rinse water used in endoscope washers.",
      "report_date": "20160323",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Steris Corporation",
      "recall_number": "Z-1180-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "73009",
      "termination_date": "20160826",
      "more_code_info": "",
      "recall_initiation_date": "20151204",
      "postal_code": "44060-1834",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}