  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Bartlett",
      "state": "TN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82451",
      "recalling_firm": "Gyrus ACMI, Inc.",
      "address_1": "2925 Appling Rd",
      "address_2": "N/A",
      "postal_code": "38133-3901",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide  Foreign: Germany,Japan, Singapore",
      "recall_number": "Z-1178-2019",
      "product_description": "Olympus Diego Elite Console  MDCONS 100",
      "product_quantity": "US: 107 consoles, OUS: 157 consoles",
      "reason_for_recall": "Olympus Diego Elite Consoles may inadvertently permit activation of the RF energy feature when used with Diego Elite blades",
      "recall_initiation_date": "20190308",
      "center_classification_date": "20190417",
      "termination_date": "20200508",
      "report_date": "20190424",
      "code_info": "Consoles containing software UIC version 2.3.9.0 and RTC version 00.01.23  US Serial Numbers:  GB00197  GB02851  GB04383  GB00235  GB02878  GB04384  GB00238  GB02995  GB04386  GB00270  GB03000  GB04400  GB00278  GB03054  GB04401  GB00280  GB03171  GB04412  GB00870  GB03173  GB04414  GB00886  GB03201  GB04416  GB00917  GB03232  GB04417  GB00925  GB03431  GB04419  GB00937O  GB03519  GB04420  GB00949  GB03675  GB04481  GB02173  GB03714  GB04482  GB02250  GB03717  GB04483  GB02263  GB03733  GB04484  GB02268  GB03746  GB04485  GB02289  GB03782  GB04486  GB02366  GB03790  GB04488  GB02368  GB03791  GB04489  GB02372  GB03793  GB04490  GB02381  GB03923  GB04491  GB02390  GB03941  GB04496  GB02406  GB03961  GB04497  GB02422  GB03966  GB04498  GB02442  GB04008  GB04503  GB02446  GB04022  GB04506  GB02453  GB04035  GB04507  GB02457  GB04041  GB04515  GB02507  GB04195  GB04516  GB02538  GB04212  GB04517  GB02539  GB04214  GB04534  GB02602  GB04241  GB04540  GB02613  GB04243  GB04592  GB02716  GB04246  GB02807  GB04315  GB02808  GB04338  GB04367    OUS:  GB03672  GB02338  GB02414  GB02863  GB00205  GB02324  GB02719  GB02933  GB02955  GB03799  GB02172  GB02460  GB02204  GB00857  GB00907  GB02928  GB00876  GB02413  GB00850  GB00879  GB00852  GB02466  GB02584  GB00698  GB02580  GB00904  GB02924  GB02930  GB00844  GB03946  GB03747  GB03226  GB03550  GB02940  GB02582  GB03432  GB04013  GB00959  GB03983  GB02655  GB02694  GB02576  GB02704  GB02925  GB02193  GB02216  GB00180  GB04244  GB00206  GB00266  GB00837  GB02463  GB00267  GB00848  GB00873  GB00821  GB03007  GB02558  GB00822  GB03233  GB02480  GB02569  GB02833  GB02559  GB02563  GB02557  GB00833  GB00702  GB02560  GB02656  GB02658  GB03005  GB02565  GB02500  GB02573  GB03677  GB02561  GB04204  GB03819  GB03921  GB02805  GB02282  GB02594  GB03189  GB02593  GB03716  GB04023  GB02659  GB02861  GB00700  GB03043  GB02283  GB03045  GB02568  GB00830  GB02889  GB03162  GB02567  GB02511  GB03023  GB03530  GB02304  GB02461  GB02519  GB02465  GB02415  GB02308  GB02328  GB02518  GB02325  GB02313  GB02494  GB04347  GB00897  GB02336  GB02579  GB00919  GB02196  GB00964  GB02194  GB00269  GB00899  GB00969  GB02280  GB02443  GB02864  GB03522  GB02217  GB03984  GB03207  GB00854  GB02497  GB02445  GB02517  GB02326  GB02299  GB02316  GB02501  GB02299  GB02316  GB02501  GB03800  GB03956  GB04543  GB04531  GB04536  GB04538  GB04518  GB04519  GB04520  GB04492  GB04493  GB04494  GB04495  GB04411  GB04413  GB04487"
    }
  ]
}