      "country": "United States",
      "city": "Utica",
      "address_1": "525 French Rd",
      "reason_for_recall": "Wire sets of certain devices were assembled with adhesive on the contact wires, which may impede conductivity and could cause failure",
      "address_2": "",
      "product_quantity": "18,720 units in total",
      "code_info": "Lot Numbers:   Y010713-02, Y03181603, Y051414-08, Y072012-12,Y09301601,  Y012312-14, Y03201511, Y05151513, Y07251703, Y100314-14,  Y013114-16, Y032513-17, Y052313-10, Y07271507, Y10061701,   Y02151602, Y03251604, Y06011502, Y07271702, Y10071603,   Y021712-02, Y03251605, Y061214-04, Y08101504, Y101413-06,  Y02171701, Y032612-15, Y061412-06, Y08171703, Y101512-06,  Y02261606, Y040314-08, Y06221508, Y082214-12, Y102714-11,  Y022712-01, Y04041701, Y062714-10, Y08241505, Y111414-09,  Y030113-13, Y041612-10, Y07061610, Y090613-11, Y11141607,  Y030514-12, Y042712-07, Y07061702, Y091214-17, Y112912-14,  Y030912-08, Y051013-06, Y071213-14, Y092012-28, Y120712-01,  Y03181602, Y05111702, Y071814-12, Y09201703, Y121714-10 & Y121813-10.",
      "center_classification_date": "20180324",
      "distribution_pattern": "Worldwide Distribution: US (Nationwide) and countries of: New Zealand, Saudi Arabia, Switzerland, United Arab Emirates Uruguay & Republic of Macedonia (The former Yugoslav Republic of Macedonia).",
      "state": "NY",
      "product_description": "MINI INFANT < 3KG, RADIOTRANSLUCENT ELECTRODE, ZOLL¿ Connector, REF/Catalog Number 2602Z     Indicated for use in external pacing, defibrillation and monitoring applications as a non-sterile, disposable device for single pediatric use, only. The PadPro 2602 electrodes provide the conductive interface between the defibrillator and/or the external transcutaneous (non-invasive) cardiac pacemaker and the pediatric patients skin. The electrode is intended for use on pediatric infants whose weight is less than 10 kg. (22 lbs.).",
      "report_date": "20180404",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "ConMed Corporation",
      "recall_number": "Z-1178-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "79337",
      "termination_date": "20200917",
      "more_code_info": "",
      "recall_initiation_date": "20180223",
      "postal_code": "13502-5945",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}