{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Utica",
      "address_1": "525 French Rd",
      "reason_for_recall": "Wire sets of certain devices were assembled with adhesive on the contact wires, which may impede conductivity and could cause failure",
      "address_2": "",
      "product_quantity": "18,720 units in total",
      "code_info": "Lot Numbers: Y01221601, Y03241702, Y10121501, Y020713-17, Y04071501, & Y12021501.",
      "center_classification_date": "20180324",
      "distribution_pattern": "Worldwide Distribution: US (Nationwide) and countries of: New Zealand, Saudi Arabia, Switzerland, United Arab Emirates Uruguay & Republic of Macedonia (The former Yugoslav Republic of Macedonia).",
      "state": "NY",
      "product_description": "MINI INFANT < 3KG, RADIOTRANSLUCENT ELECTRODE, Philips Plug Style Connector, REF/Catalog Number 2602H     Indicated for use in external pacing, defibrillation and monitoring applications as a non-sterile, disposable device for single pediatric use, only. The PadPro 2602 electrodes provide the conductive interface between the defibrillator and/or the external transcutaneous (non-invasive) cardiac pacemaker and the pediatric patients skin. The electrode is intended for use on pediatric infants whose weight is less than 10 kg. (22 lbs.).",
      "report_date": "20180404",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "ConMed Corporation",
      "recall_number": "Z-1175-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "79337",
      "termination_date": "20200917",
      "more_code_info": "",
      "recall_initiation_date": "20180223",
      "postal_code": "13502-5945",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}