  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67157",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and Internationally to AUSTRIA, BELGIUM, CZECH REPUBLIC, FINLAND, FRANCE GERMANY, GREECE, ITALY, LITHUANIA, POLAND, SPAIN, SWEDEN, SWITZERLAND,  UNITED KINGDOM.",
      "recall_number": "Z-1175-2014",
      "product_description": "GE Healthcare CARESCAPE Monitor B650 and CARESCAPE Monitor B850.    K102239 CARESCAPE Monitor B650 The CARESCAPE Monitor B650 is a multi-parameter patient monitor intended for use in multiple areas and intra-hospital transport within a professional healthcare facility. The CARESCAPE Monitor B650 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time. The CARESCAPE Monitor B650 is indicated for monitoring and recording of, and to generate alarms for, hemodynamic (including ECG, ST segment, arrhythmia detection, ECG diagnostic analysis and measurement, invasive pressure, non-invasive blood pressure, pulse oximetry, cardiac output, temperature and mixed venous oxygen saturation), impedance respiration, airway gases (CO2, O2, N2O and anesthetic agents), spirometry, gas exchange, and neurophysiological (including electroencephalography, Entropy, Bispectral Index (BIS) and neuromuscular transmission(NMT) status. The CARESCAPE Monitor B650 can be a stand-alone monitor or interfaced to other devices. It can also be connected to other monitors for remote viewing and to data management software devices via a network. The CARESCAPE Monitor B650 is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility. The CARESCAPE Monitor B650 is not intended for use during MRI. K092027 CARESCAPE Monitor B850 The CARESCAPE Monitor B850 is a multi-parameter high acuity patient monitor intended for use in multiple areas within a professional healthcare facility. The CARESCAPE Monitor B850 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time. The CARESCAPE Monitor B850 system is indicated for monitoring of Hemodynamic (including ECG, ST Segment, Arrhythmia Detection, ECG Diagnostic Analysis and Measurement, Invasive Pressure, Noninvasive Blood Pressure, Pulse Oximetry, Cardiac Output, Temperature, Impedance Respiration and SvO2 (Mixed Venous Oxygen Saturation), Airway Gases (Fi/Et C02, 02, N20 and Anesthetic Agent), Spirometry, Gas Exchange (02 Consumption (VO2), C02 production (VCO2), energy expenditure (BE), and respiratory quotient (RQ)) and neurophysiological (including electroencephalography (EEG), Entropy, Bispectral Index (BIS) and Neuromuscular Transmission (NMT) Monitoring) status. The CARESCAPE Monitor B850 provides alarms, trends, snapshots and events, and calculations and can be connected to displays, printers and recording devices. The CARESCAPE Monitor B850 can be a stand-alone monitor or interfaced to other devices. It can also be connected to other monitors for bed to bed viewing and to data management software devices via a network. The CARESCAPE Monitor B850 is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility. In addition to the healthcare practitioner, the CARESCAPE Monitor B850 is designed to provide configuration and troubleshooting capabilities to qualified service personnel. The CARESCAPE Monitor B850 is not intended for use during MRI. K131414 CARESCAPE MONITOR B850 The CARESCAPE Monitor B850 is a multi-parameter high acuity patient monitor intended for use in multiple areas within a professional healthcare facility. The CARESCAPE Monitor B850 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time. The CARESCAPE Monitor B850 system is indicated for monitoring of Hemodynamic (including ECG, ST Segment, Arrhythmia Detection, ECG Diagnostic Analysis and Measurement, Invasive Pressure, Noninvasive Blood Pressure, Pulse Oximetry, Cardiac Output, Thermodilution and pulse contour), Temperature, Mixed Venous Oxygen Saturation, and Central Venous Oxygen Saturation). Respiratory [Impedance Respiration, Airway Gases (C02, 02, N20 and Anesthetic Agents), Spirometry, Gas Exchange] and N",
      "product_quantity": "676 units OUS",
      "reason_for_recall": "GE Healthcare has recently become aware of potential safety issues associated with the CARESCAPE Monitor B850 and CARESCAPE Monitor B650.  These issues include both NIBP (Non-Invasive Blood Pressure)  and Bed to Bed Issues.",
      "recall_initiation_date": "20140228",
      "center_classification_date": "20140311",
      "termination_date": "20150429",
      "report_date": "20140319",
      "code_info": "Mfg Lot or Serial #  SEW13380452HA  SEW13380454HA  SEW13380455HA  SEW13380500HA  SEW13380501HA  SEW13380502HA  SEW13380503HA  SEW13380504HA  SEW13380505HA  SEW13380506HA  SEW13380507HA  SEW13380508HA  SEW13380509HA  SEW13380512HA  SEW13380520HA  SEW13380524HA  SEW13380525HA  SEW13380526HA  SEW13380534HA  SEW13380535HA  SEW13380536HA  SKA13130068GA  SKA13130071GA  SKA13130075GA  SKA13130077GA  SKA13130078GA  SKA13130082GA  SKA13130085GA  SKA13140118GA  SKA13370167GA  SKA13370211GA  SKA13370214GA  SKA13370215GA  SKA13370216GA  SKA13370217GA  SKA13370218GA  SKA13370219GA  SKA13370220GA  SKA13370221GA  SKA13370222GA  SKA13370223GA  SKA13370224GA  SKA13370225GA  SKA13370226GA  SKA13370227GA  SKA13370228GA  SKA13370229GA  SKA13370230GA  SKA13370231GA  SKA13370232GA  SKA13370233GA  SKA13370234GA  SKA13370235GA  SKA13370236GA  SKA13370237GA  SKA13370238GA  SKA13370239GA  SKA13370240GA  SKA13370241GA  SKA13370242GA  SKA13370263GA  SKA13370265GA  SKA13370267GA  SKA13370268GA  SKA13370269GA  SKA13370270GA  SKA13370271GA  SKA13370272GA  SEW13380133HA  SEW13380155HA  SEW13380256HA  SEW13380258HA  SEW13380259HA  SEW13380260HA  SEW13380264HA  SEW13380265HA  SEW13380266HA  SEW13380267HA  SEW13380268HA  SEW13380269HA  SEW13380271HA  SEW13380305HA  SEW13380319HA  SEW13380371HA  SEW13380375HA  SEW13380377HA  SEW13380400HA  SEW13380413HA  SEW13380415HA  SEW13380422HA  SEW13380423HA  SEW13380424HA  SEW13380425HA  SEW13380426HA  SEW13380427HA  SEW13380429HA  SEW13380432HA  SEW13380433HA  SEW13380434HA  SEW13380435HA  SEW13380436HA  SEW13380437HA  SEW13380438HA  SEW13380439HA  SEW13380440HA  SEW13380441HA  SEW13380442HA  SEW13380443HA  SEW13380444HA  SEW13380445HA  SEW13380446HA  SEW13380447HA  SEW13380453HA  SEW13380459HA  SEW13380461HA  SEW13380463HA  SEW13380464HA  SEW13380465HA  SEW13380497HA  SEW13380498HA  SEW13380499HA  SKA13130089GA  SKA13130101GA                                      SED13017065GA  SED13017067GA  SED13027233GA  SED13027251GA  SED13027253GA  SED13027259GA  SED13027261GA  SED13027263GA  SED13027265GA  SED13027266GA  SED13027267GA  SED13027268GA  SED13027269GA  SED13027270GA  SED13027271GA  SED13027272GA  SED13027273GA  SED13027274GA  SED13027275GA  SED13027276GA  SED13027277GA  SED13027278GA  SED13027412GA  SED13027413GA  SED13027415GA  SEW13379826HA  SEW13379860HA  SEW13379863HA  SEW13380254HA  SEW13380306HA  SEW13380313HA  SEW13380399HA  SEW13380402HA  SKA13370170GA        SED10353754GA  SED10353755GA  SED10353756GA  SED10353757GA  SED10353758GA  SED10353759GA  SED10353760GA  SED10353761GA  SED10373970GA  SED10373971GA  SED10373972GA  SED10373973GA  SED10373974GA  SED10373975GA  SED10373977GA  SED10373980GA  SED10373981GA  SED10373983GA  SED10373988GA  SED10374003GA  SED10384049GA  SED10384064GA  SED10384085GA  SED10384087GA  SED10384092GA  SED10384100GA  SED10384102GA  SED10384107GA  SED10384109GA  SED10384112GA  SED10384117GA  SED10384119GA  SED10384120GA  SED10384121GA  SED10384122GA  SED10384123GA  SED10384124GA  SED10384125GA  SED10384126GA  SED10384127GA  SED10384128GA  SED10384129GA  SED10384131GA  SED10384132GA  SED10384143GA  SED10384149GA  SED10394174GA  SED10404188GA  SED10485073GA  SED10536439GA  SED10536442GA  SED10536443GA  SED10536444GA  SED10536447GA  SED10536457GA  SED10536459GA  SED10536473GA  SED11026570GA  SED11036588GA  SED11036609GA  SED11036610GA  SED11036618GA  SED11036619GA  SED11036620GA  SED11036622GA  SED11036658GA  SED11036660GA  SED11036662GA  SED11036668GA  SED11036675GA  SED11036686GA  SED11057067GA  SEW10334205HA  SEW10334221HA  SEW10334222HA  SEW10334244HA  SEW10334255HA  SEW10385224HA  SEW10385242HA  SEW10385273HA  SEW10385281HA  SEW10466987HA  SEW10466989HA  SEW10466990HA  SEW10466991HA  SEW10466992HA  SEW10466993HA  SEW10466995HA  SEW10466997HA  SEW10466999HA  SEW10467002HA  SEW10467006HA  SEW10467007HA  SEW10467087HA  SEW10467088HA  SEW10467089HA  SEW10467092HA  SEW10467093HA  SEW10467094HA  SEW10467096HA  SEW10467358HA  SEW10467370HA  SEW11141062HA  SEW11151201HA  SEW11151278HA  SEW11171443HA  SEW11171494HA  SEW11171513HA  SEW11181832HA  SEW11202312HA  SEW11202689HA  SEW11202691HA  SEW13380257HA  SEW13380263HA  SEW13380275HA  SEW13380276HA  SEW13380277HA  SEW13380286HA  SEW13380322HA  SEW13380367HA  SEW13380368HA  SEW13380373HA  SEW13380386HA  SEW13380387HA  SEW13380391HA  SEW13380511HA  SEW13380514HA  SEW13380517HA  SEW13380527HA  SEW13380528HA  SEW13380529HA  SEW13380531HA  SEW13380533HA  SEW13380537HA  SEW13380541HA  SEW13380542HA  SEW13390694HA  SEW13390695HA  SEW13390696HA  SEW13390697HA  SEW13390704HA  SEW13390710HA  SKA13360147GA  SKA13360148GA  SKA13360149GA  SKA13360150GA  SKA13360152GA  SKA13360156GA  SKA13360157GA  SKA13360158GA  SKA13360159GA  SKA13360160GA  SKA13360161GA  SKA13360162GA  SKA13370194GA  SKA13370205GA  SKA13370206GA  SKA13370207GA  SKA13370208GA  SKA13370210GA  SKA13370212GA  SKA13370213GA  SED11199036GA  SED11199040GA  SED11209172GA  SED11209176GA  SED11209177GA  SED11209178GA  SED11209188GA  SED11209213GA  SED11209215GA  SED11209219GA  SED11209221GA  SED11209222GA  SED11209223GA  SED11209225GA  SED11209226GA  SED11209227GA  SED11209228GA  SED11209229GA  SED11209230GA  SED11209231GA  SED11219232GA  SED11219233GA  SED11219234GA  SED11219235GA  SED11219238GA  SED11219239GA  SED11219240GA  SED11219241GA  SED11219242GA  SED11219243GA  SED11219245GA  SED11219246GA  SED11219247GA  SED11219248GA  SED11219249GA  SED11219252GA  SED11219253GA  SED11219254GA  SED11219255GA  SED11219256GA  SED11219258GA  SED11219259GA  SED11219262GA  SED11219263GA  SED11219265GA  SED11219266GA  SED11219267GA  SED11219270GA  SED11219273GA  SED11219279GA  SED11219280GA  SED11219281GA  SED11219282GA  SED11219284GA  SED11219299GA  SED11219300GA  SED11219302GA  SED11219303GA  SED11219304GA  SED11219305GA  SED11219306GA  SED11219307GA  SED11219308GA  SED11219309GA  SED11219310GA  SED11219311GA  SED11219312GA  SED11219313GA  SED11219314GA  SED11219315GA  SED11219317GA  SED11219318GA  SED11219319GA  SED11219320GA  SED11219321GA  SED11219322GA  SED11219325GA  SED11219326GA  SED11219327GA  SED11219328GA  SED11219330GA  SED11219339GA  SED11219343GA  SED11219387GA  SED11219388GA  SED11219403GA  SEW11140794HA  SEW11140799HA  SEW11140801HA  SEW11140805HA  SEW11140808HA  SEW11140878HA  SEW11140990HA  SEW11141000HA  SEW11141013HA  SEW11141016HA  SEW11202251HA  SEW11202256HA  SEW11202257HA  SEW11202259HA  SEW11202265HA  SEW11202269HA  SEW11202270HA  SEW11202272HA  SEW11202273HA  SEW11223135HA  SEW11405474HA  SEW11405475HA  SEW11405515HA  SEW11405517HA  SEW11405518HA  SEW11405519HA  SEW11405520HA  SEW11405522HA  SEW11405526HA  SEW11405527HA  SEW11405528HA  SEW11405529HA  SEW11405536HA  SEW11507853HA  SEW11507854HA  SEW11507856HA  SEW13380392HA  SEW13380393HA  SEW13380394HA  SEW13380395HA  SEW13380396HA  SEW13380397HA  SEW13380416HA  SEW13380417HA  SEW13380418HA  SEW13380419HA  SEW13380420HA  SEW13380421HA  SEW13380532HA  SKA13370163GA  SKA13370165GA  SKA13370169GA  SKA13370172GA  SKA13370173GA  SKA13370175GA  SKA13370176GA  SKA13370177GA  SKA13370178GA  SKA13370179GA  SKA13370180GA  SKA13370182GA  SKA13370183GA    SEW13380157HA  SEW13380158HA  SEW13380162HA  SEW13380206HA  SEW13380207HA  SEW13380208HA  SEW13380270HA  SEW13380272HA  SEW13380274HA  SEW13380280HA  SEW13380281HA  SEW13380282HA  SEW13380283HA  SEW13380284HA  SEW13380285HA  SEW13380288HA  SEW13380289HA  SEW13380290HA  SEW13380291HA  SEW13380292HA  SEW13380293HA  SEW13380295HA  SEW13380296HA  SEW13380297HA  SEW13380298HA  SEW13380299HA  SEW13380301HA  SEW13380312HA  SEW13380314HA  SEW13380315HA  SEW13380316HA  SEW13380317HA  SKA13360151GA  SKA13360154GA    SEW13380518HA  SEW13401073HA  SEW13380309HA  SEW13380389HA  SEW13380401HA  SEW13380403HA  SEW13380404HA  SEW13380405HA  SEW13380411HA    SEW13379847HA  SEW13379861HA  SKA13370174GA  SKA13370181GA  SKA13370184GA  SKA13370187GA  SKA13370188GA  SKA13370190GA  SKA13370191GA  SKA13370192GA  SEW13380278HA  SEW13380279HA  SEW13380287HA  SEW13380294HA  SEW13380300HA  SEW13380307HA  SEW13380311HA  SEW13380398HA  SEW13380519HA  SEW13380521HA  SEW13380522HA  SEW13390703HA  SKA13360153GA  SEW13380510HA  SEW13380530HA  SEW13380539HA  SEW13380543HA  SEW13380544HA  SEW13380555HA  SEW13390701HA  SEW13390702HA  SEW13380414HA  SEW13380448HA  SEW13380449HA  SEW13380450HA  SEW13380451HA  SEW13380458HA  SEW13380460HA          SED10384090GA  SED10394175GA  SED10394176GA  SED10404203GA  SED10404204GA  SED10404209GA  SED10404210GA  SED10404257GA  SED11310355GA  SED11461107GA  SED11461113GA  SED11461116GA  SED11461119GA  SED11461120GA  SED11461121GA  SED11461123GA  SED12051712GA  SED12051713GA  SED12051714GA  SED12051715GA  SED12051717GA  SED12051718GA  SED12051726GA  SED12051727GA  SED12051728GA  SED12051729GA  SED12051730GA  SED12051731GA  SED12051732GA  SED12051734GA  SED12051738GA  SED12061859GA  SED12061861GA  SED12061863GA  SED12061864GA  SED12061865GA  SED12061866GA  SED12061870GA  SED12061871GA  SED12061874GA  SED12061875GA  SED12071892GA  SED12071896GA  SED12071899GA  SED12071900GA  SED12082106GA  SED12082108GA  SED12082144GA  SED12082156GA  SED12082158GA  SED12082165GA  SED12082169GA  SED12082170GA  SED12082199GA  SED12394520GA  SED12394580GA  SED12394585GA  SED12394589GA  SED12394598GA  SED12394694GA  SED12394704GA  SED12475648GA  SED12485742GA  SED12485782GA  SEW10293722HA  SEW10467037HA  SEW10467141HA  SEW10467161HA  SEW10467163HA  SEW10467174HA  SEW10467179HA  SEW10467181HA  SEW10467182HA  SEW10467183HA  SEW10467185HA  SEW12193055HA  SEW13380273HA  SEW13380302HA  SEW13380303HA  SEW13380304HA  SEW13380308HA  SEW13380310HA  SEW13380318HA  SEW13380320HA  SEW13380321HA  SEW13380366HA  SEW13380369HA  SEW13380370HA  SEW13380372HA  SEW13380374HA  SEW13380376HA  SEW13380378HA  SEW13380379HA  SEW13380380HA  SEW13380381HA  SEW13380382HA  SEW13380383HA  SEW13380384HA  SEW13380385HA  SEW13380388HA  SEW13380390HA  SEW13380406HA  SEW13380407HA  SEW13380408HA  SEW13380409HA  SEW13380410HA  SEW13380412HA  SEW13380428HA  SEW13380430HA  SEW13380431HA  SEW13380457HA  SEW13380462HA  SEW13380513HA  SEW13380515HA  SEW13380523HA  SEW13380540HA  SEW13380546HA  SEW13380547HA  SEW13380548HA  SEW13380549HA  SEW13380550HA  SEW13380551HA  SEW13380553HA  SEW13380554HA  SEW13390698HA  SEW13390699HA  SEW13390712HA  SEW13390719HA  SEW13390720HA  SEW13390775HA  SKA13130069GA  SKA13130070GA  SKA13130072GA  SKA13130073GA  SKA13130079GA  SKA13130080GA  SKA13130081GA  SKA13130083GA  SKA13130084GA  SKA13130086GA  SKA13130087GA  SKA13130097GA  SKA13130098GA  SKA13130099GA  SKA13360155GA  SKA13370164GA  SKA13370166GA  SKA13370168GA  SKA13370171GA  SKA13370185GA  SKA13370186GA  SKA13370189GA  SKA13370193GA  SKA13370195GA  SKA13370196GA  SKA13370197GA  SKA13370198GA  SKA13370199GA  SKA13370200GA  SKA13370201GA  SKA13370202GA  SKA13370203GA  SKA13370204GA  SKA13370209GA  SKA13370257GA  SKA13370259GA  SKA13370264GA"
    }
  ]
}