  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Miami Lakes",
      "address_1": "14201 Nw 60th Ave",
      "reason_for_recall": "Incorrect cannula of the sheath introducer (smaller than intended).",
      "address_2": "",
      "product_quantity": "1,985 units",
      "code_info": "Catalog No. 504656X (GTIN: 20705032010389)  Lot No. 17305580, Expiration date: 2018-07-31.",
      "center_classification_date": "20160314",
      "distribution_pattern": "US Nationwide Distribution including states of:AL, AZ, CA, CO, CT, FL, GA, IN, KY, LA, MA, MD, MI, MN, MO, MT, NC, NH, NJ,NY, OH, OR, PA, RI, TN, TX, VA, WA, WV, and Hawaii.",
      "state": "FL",
      "product_description": "CORDIS AVANTI + CATHETER SHEATH INTRODUCER SYSTEM    The product includes a Catheter Sheath Introducer (CSI) and a vessel dilator. A Catheter Sheath Introducer facilitates percutaneous entry of an intravascular device. A vessel dilator facilitates the percutaneous entry of the CSI by forming an atraumatic transition from the skin through the subcutaneous tissue to the vessel.",
      "report_date": "20160323",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Cordis Corporation",
      "recall_number": "Z-1174-2016",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "product_type": "Devices",
      "event_id": "73183",
      "termination_date": "20160825",
      "more_code_info": "",
      "recall_initiation_date": "20151231",
      "postal_code": "33014-2802",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}