{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Raynham",
      "address_1": "325 Paramount Drive",
      "reason_for_recall": "The IFU provided with the free strand ORTHOCORD sutures currently indicates that approximately 100% suture strength remains at six weeks after implantation. While this statement is true for Violet ORTHOCORD sutures, it is an inaccurate statement for Blue ORTHOCORD free strand sutures. Based on in-vivo strength retention study, a Blue ORTHOCORD sutures maintain approximately 80% of strength at six",
      "address_2": "",
      "product_quantity": "65062",
      "code_info": "ALL from Jan 1 2010 - Corrected IFU implementation date",
      "center_classification_date": "20130425",
      "distribution_pattern": "Nationwide Distribution.",
      "state": "MA",
      "product_description": "ORTHOCORD BLUE BRAIDED COMPOSITE SUTURE 2 (5 METRIC), Catalog Number: 223111.    For use in general soft tissue approximation, and/or ligation, including orthopedic procedures.",
      "report_date": "20130501",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "DePuy Mitek, Inc., a Johnson & Johnson Co.",
      "recall_number": "Z-1172-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "64706",
      "termination_date": "20140224",
      "more_code_info": "",
      "recall_initiation_date": "20130322",
      "postal_code": "02767-5199",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}