{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Oyster Bay",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79196",
      "recalling_firm": "Circulatory Technology Inc",
      "address_1": "21 Singworth St",
      "address_2": "N/A",
      "postal_code": "11771-3703",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "**Requested update** Distributed in 19 states: AL, AR, CA, FL, GA, IL, IN, KY, MD, MO, NJ, NY, OH, OR, PA, TN, TX, VA, WA, and the District of Columbia.",
      "recall_number": "Z-1171-2018",
      "product_description": "**Recall Expanded to add these models: Bigger Better-Bladder cardiopulmonary bypass blood reservoir; Models: BBB38 (individual, sterile, pouched) and BBB38NS (sold bulk to kit manufacturers)",
      "product_quantity": "410",
      "reason_for_recall": "The device is used as part of extracorporeal membrane oxygenation (ECMO) cardiopulmonary bypass circuits and may collapse during use which can increase resistance to flow in the venous line and cause a drop in circuit blood flow.",
      "recall_initiation_date": "20170922",
      "center_classification_date": "20180323",
      "termination_date": "20200520",
      "report_date": "20180307",
      "code_info": "**Recall expanded to add these models: BBB38, Lot Number: 5300-S19813; BBB38NS, Lot Number: 1707-1 and 170903"
    }
  ]
}