      "status": "Completed",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79006",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "N/A",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Other",
      "distribution_pattern": "Australia, Austria, Belarus,  Belgium, Bosnia and Herzegovina, Brazil,  Canada, China, Croatia,  Cyprus, Czech Republic, Denmark,  Finland, France, Germany,  Gibraltar, Greece, Hong Kong,  Hungary, India, Ireland,  Israel, Italy, Korea,  Kuwait, Latvia, Lithuania,  Macao, Mayotte, Mexico,  Netherlands, New Zealand, Norway,  Panama, Poland, Portugal,  Qatar, Russia, Saudi Arabia,  Singapore, Slovakia, Slovenia,  South Africa, Spain, Sweden,  Switzerland, Taiwan, Turkey,  United Kingdom, United States",
      "recall_number": "Z-1169-2018",
      "product_description": "AutoMate 1200 Catalog Number ODL25120;  AutoMate 1250 Catalog Number ODL25125;  AutoMate 2500 Catalog Number ODL25250;  AutoMate 2550 Catalog Number ODL25255",
      "product_quantity": "98 units in the US",
      "reason_for_recall": "Belts on the x-axis (and on the y-axis, to a lesser extent) manufactured from 2015 are showing premature failure, posing the remote risk of cross-contamination.",
      "recall_initiation_date": "20180119",
      "center_classification_date": "20180322",
      "report_date": "20180328",
      "code_info": "All systems up to the following Serial Numbers:  ODL25120 Revision T; SN3101,  ODL25125 Revision W; SN1508,  ODL25250 Revision T; SN2416,  ODL25255 Revision W; SN4617"
    }
  ]
}