{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Palm Beach Gardens",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67183",
      "recalling_firm": "Biomet 3i, LLC",
      "address_1": "4555 Riverside Dr",
      "address_2": "N/A",
      "postal_code": "33410-4200",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA Nationwide in the states of AL, AZ, CA, CO, CT, DC, DE, FL, GA, IL, KS, KY, LA, MA, MD, MI, MN, MS, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WI, WV, including Hawaii and Puerto Rico and the countries of : Argentina, Austria, Canada, China, Costa Rica, Germany, Dominican Republic, Spain, France, Great Britain, Hong Kong, Hungary, Israel, Japan, Korea, Mexico, Poland, Portugal, Paraguay, Sweden, Singapore, Taiwan, and Uruguay.",
      "recall_number": "Z-1167-2014",
      "product_description": "Navigator Surgical Kit    Product Usage:  Molded plastic tray used in holding various dental instruments for autoclave processing.",
      "product_quantity": "21",
      "reason_for_recall": "During recent revised sterilization validation testing, the included surgical trays did not meet the Sterility Assurance Level required.",
      "recall_initiation_date": "20130722",
      "center_classification_date": "20140311",
      "termination_date": "20150421",
      "report_date": "20140319",
      "code_info": "SGKIT (SGTRAY)"
    }
  ]
}