  "results": [
    {
      "status": "Ongoing",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91649",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "N/A",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-1166-2023",
      "product_description": "Luminos dRF with software VD10- A diagnostic imaging system for  radiographic and fluoroscopic studies  Model Number: 10094200",
      "product_quantity": "1 unit; Expanded Recall: 25 units",
      "reason_for_recall": "Potential risk of collision with the ceiling, wall, or objects which may result in serious injury to staff or patients due to room configuration parameters of the system being set to default values",
      "recall_initiation_date": "20221221",
      "center_classification_date": "20230223",
      "report_date": "20230301",
      "code_info": "UDI: N/A S/N: 4650  Expanded Recall 2/24/23: 4043 4051 4143 4156 4199 4201 4238 4252 4291 4323 4332 4342 4354 4372 4401 4430 4508 4600 4606 4637 4641 4683 4722 4724 4734"
    }
  ]
}