{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Shenyang, Liaoning",
      "state": "",
      "country": "China",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64783",
      "recalling_firm": "Philips And Neusoft Medical Systems Co., Ltd.",
      "address_1": "16 Century Road, Neusoft Park,",
      "address_2": "Hun Nan Industrial Area",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide distribution: USA including states of:  California, Connecticut, North Carolina, Nebraska, Ohio, South Carolina, Texas, and Puerto Rico.",
      "recall_number": "Z-1165-2013",
      "product_description": "NeuViz 16 Multi-Slice CT Scanner System, Part number (PN): 989605858501, Model:  NeuViz 16.  This is a computed X-ray, Tomography system.    The NeuViz 16 CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously. The acquired X-RAY transmission data is reconstructed by computer into cross-sectional images of the body from the same axial plane taken at different angles. The system is suitable for all patients",
      "product_quantity": "23 units",
      "reason_for_recall": "User error of the NeuViz 16  or NeuViz Dual may cause personnel injuries. There was a report where a finger was pinched when the table  was moved backward while the tabletop was controlled by another person from the gantry left control panel.",
      "recall_initiation_date": "20130304",
      "center_classification_date": "20130423",
      "termination_date": "20131122",
      "report_date": "20130501",
      "code_info": "Serial numbers:  N16E090002, N16E090003, N16E090004, N16E090007, N16E090008, N16E090010, N16E100015, N16E100017, N16E100023, N16E110016, N16E110023, N16E110026, N16E110027, N16E110040, N16E110053, N16E120016, N16E120020, N16E120022, N16E120023, N16E120024, N16E120052, N16E120053, and N16E130002.",
      "more_code_info": ""
    }
  ]
}