{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67257",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) including the states of  AL, AK, AZ, AR, CA, CO, CT, FL, GA, IL, IN, KS, KY, LA, ME, MD, MA, MI, MS, MO, NV, NJ, NY, NC, ND, OH, OK OR, PA, PR, RI, SC, TN, TX, UT, VA,WA,WY and  WI., and the countries of  ARGENTINA, AUSTRALIA, CANADA, CHINA, CZECH REPUBLIC, DENMARK, EGYPT, FINLAND, FRANCE, GERMANY, HUNGARY, INDIA, IRELAND, ITALY, JAPAN, JORDAN, REPUBLIC OF KOREA, MALAYSIA, MEXICO, NEW ZEALAND, NORWAY, PHILIPPINES, POLAND, RUSSIA, SAUDI ARABIA, SPAIN, SRI LANKA, TURKEY and UNITED KINGDOM.",
      "recall_number": "Z-1163-2014",
      "product_description": "GE Healthcare, Innova 2100IQ, Innova 3100, Innova 3100IQ, Innova 4100, Innova 4100IQ Innova 2121IQ, Innova 3131IQ; Cardiovascular X-ray imaging systems (X-ray generator installed with Rotor V3 version and Heater V4 version boards).    The Digital Fluoroscopic Imaging System is indicated for use in generating fluoroscopic images of human anatomy for diagnostic and intervention angiography procedures. It is intended to replace fluoroscopic images obtained through the image intensifier technology. This device is not intended for mammography applications. The Innova 3100 system is indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, and optionally, rotational and angiography procedures. It is intended to replace fluoroscopic images obtained through image intensifier technology. The Digital Fluoroscopic Imaging System is indicated for use in diagnostic and interventional angiographic procedures of human anatomy. It is intended to replace image intensifier fluoroscopic systems in all diagnostic or interventional procedures. This device is not intended for mammography applications. A new tilt table will support performing procedures like CO2 studies, Venography.  The Digital Fluoroscopic Imaging System with Bolus Chasing option is indicated for use in generating fluoroscopic images of human anatomy for diagnostic and intervention; the option Bolus enables to perform subtraction angiography of lower limbs. This device is not intended for mammography applications. The Innova 2100-IQ system is indicated for use in generating fluoroscopic images of human anatomy for vascular angiography diagnostic and interventional procedures, and optionally, rotational angiography procedures. It is also indicated for generating fluoroscopic images of human anatomy for cardiology diagnostic and interventional procedures. It is intended to replace fluoroscopic images obtained through image intensifier technology. This device is not intended for mammography applications.  For Innova 4100, Innova 4100IQ, Innova 3100, Innova 3100 IQ, Innova 2100IQ with Innova 3T Option: The Innova systems are indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and optionally, rotational imaging procedures. They are also indicated for generating fluoroscopic images of human anatomy for cardiology, diagnostic, and interventional procedures. They are intended to replace fluoroscopic images obtained through image intensifier technology. Those devices are not intended for mammography applications. Innova CT is a software option, which reconstructs 3D volumes from Rotational Fluoroscopy acquisition to provide images that assist the physician in diagnosis, surgical planning, interventional procedures and treatment follow-up. Innova CT is intended for imaging bone and soft tissues as well as other internal body structures. Innova CT is not intended for mammography applications. For Innova 4100, Innova 4100IQ, Innova 3100, Innova 3100 IQ, Innova 2100IQ with Innova 3D or InnovaSpin option: The Innova systems are indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and optionally, rotational imaging procedures. They are also indicated for generating fluoroscopic images of human anatomy for cardiology, diagnostic, and interventional procedures. They are intended to replace fluoroscopic images obtained through image intensifier technology. Those devices are not intended for mammography applications. Innova 3D is a software option which reconstructs 3D volumes from Rotational Fluoroscopy acquisition to assist the physician in diagnosis, surgical planning, interventional procedures and treatment follow-up. It is not intended for mammography applications. InnovaSpin is a software option that permits fast spin rotational angiography. It is not intended for mammography applications. :",
      "product_quantity": "257(159 US, 98 OUS)",
      "reason_for_recall": "GE Healthcare has become aware of a potential safety issue involving the innova systems. The Innova system can unexpectedly stop delivering X-rays after a power-on or a reset cycle which may result in the loss of real-time interventional imaging. No injuries have been reported due to this issue.",
      "recall_initiation_date": "20131211",
      "center_classification_date": "20140310",
      "termination_date": "20140729",
      "report_date": "20140319",
      "code_info": "Mfg Lot or Serial # System ID  00000550134BU1 2035763100A  00000581573BU3 203932VA4100  00000523132BU9 205783BCL5  00000560743BU7 206598UWBP  00000446090BU3 207351YCL1  00000504213BU0 212305CATH1  00000437215BU7 214324INNOVA2  00000562417BU6 2198864100  00000093723VE5 229228INNOV21C  00000583484BU1 240566CATH2  00000478109BU2 251435ICV3  00000477896BU5 3016094100  00000441137BU7 3018774100  00000466659BU0 303788CL1  00000466984BU2 304424INNOVA  00000445301BU5 309691P3100  00000424103BU0 310423INNOVA1  00000579612BU3 313745IN1  00451726BU4BU8 316268INNOVAA  00000466810BU9 318449INNOVA2  00000482326BU6 3184734100  00000499386BU1 330375LAB2  00000445906BU1 334286BSP  00000551081BU3 336832MCCL6  00000464558BU6 352333CATH2  00000463833BU4 361980INNOVA2  00000427056BU7 401274IN41  00000585912BU9 404712SP11  00000585915BU2 405271P4100  00000593051BU6 405610MWC3100EP  00000449428BU2 405840OIC3100  00000541272BU1 407518EP  00000431502BU4 407518SP3  00000452126BU6 407SPECIALS  00000490443BU9 408559INV1  00000421698BU2 4103374100  00000461599BU3 410337INNOVA6  00000571721BU0 410368INNOVA1  00000442436BU2 4105434100  00000533867BU8 410554INNOVA1  00000548841BU6 410554INNOVA4  00000560325BU3 410614CV1  00000443912BU1 412623LAB3  00000454264BU3 419251LAB1  00000576989BU8 423493CL1  00000450075BU7 480821IN2  00000468193BU8 501257VALAB1  00000434782BU9 502587LAB4  FMI12204DUMM03 508383IN4100  00000581740BU8 509473RM1  00000459476BU8 5135844100  00000415698BU0 5136864100  00000479625BU6 516562INNOVA5A  00000460641BU4 516562INNOVA6  00000460969BU9 516562INNOVA7  00000441966BU9 516663INNOVA2  00000467462BU8 518525LAB1  00000467645BU8 540953INNOVA  00000513959BU7 570476INNOVA21  00000466830BU7 573256MHC  00000479758BU5 573815BCR2  00000433021BU3 573882UMC1  00000446439BU2 6012884100SP1  00000558894BU2 601376SP4100  00000437985BU5 6082634100  00000410733BU0 608324DAL  00000543076BU4 608775NOVA1  00000561529BU9 609267IN2100  00000546435BU9 609757IN3100  00000491706BU8 610250INNOVA3  00000442603BU7 614293LIBERTY2  00000563635BU2 614722INNOVA1  00000458814BU1 615338CV2  00000493727BU2 615446CV1  00000458501BU4 619229INNOVA1  00000438341BU0 626359INNOVA  00000463834BU2 65049331  00000546155BU3 7018573100  00000469683BU7 7028533100  00000433013BU0 706278INNOVA1  00000421893BU9 706802FINN  00000454140BU5 7083333100  00000427535BU0 713776M3100  00000555540BU4 717782INNOVA31B  00000459460BU2 718283INNOVA3  00000490237BU5 718470INNOVA2  00000436326BU3 718518CVL1  00000398531BU4 727725SPEC1  00000571258BU3 727942CATH  00000559817BU2 732776CATH2  00000533869BU4 732776CATH6  00000460652BU1 765453INNOVA1  00000467397BU6 770535N4100  00000478491BU4 770991CL3  00000449237BU7 801268INNOVA  00000432806BU8 80140841  00000479709BU8 804594CATH  00000438367BU5 804764CATH3  00000442737BU3 812283IN31  00000592458BU4 812450CL2  00000479578BU7 812858GWLAB2  00000492209BU2 813972INTV  00000439429BU2 815226INN2  00000434732BU4 815599FHN3100  00000538053BU0 8157412100  00000467323BU2 816478MC1  00000477463BU4 817329BG2100  00000557168BU2 8189813100  00000447651BU1 828298VASC  00000460469BU0 845365V19  00000523086BU7 845368LAB1  00000544815BU4 850469BCV1  00000544202BU5 850747IN4100  00000485433BU7 856596IN3100  00000491032BU9 858554INNOVA2  00000422738BU5 8595724100  00000467881BU9 864255SF4100  00000548598BU2 864255SFCL2  00000537821BU1 865693CCL1  00000596870BU6 903CSM2121  00000458184BU9 903GS4100  00000454594BU3 904202CATH3  00000496915BU0 907562CATH1  00000436762BU9 918744JR4100  00000549261BU6 9284532100  00000442333BU1 9408984100  00000439384BU9 951788INNOVA4  00000545588BU6 954987CL4  00000482458BU7 954MW4100  00000430321BU0 956632INN4100  00000469847BU8 956698INN31  00000430322BU8 985867INNOVA2  00000514612BU1 COR359867  00000403215BU7 FWANG3217  00000580430BU7 2036884100B  00000590065BU9 303399EP  00000571259BU1 318449INNOVA3  00000057472BU0 352333CV3  00000590868BU6 361788CATH1  00000543267BU9 410337INNOVA2  00000590608BU6 412623H  00000480416BU7 423493I4100  00000594389BU9 541743OR15  00000436759BU5 614293LIBERTY1  00000590864BU5 615396EP  00000538571BU1 631968INNOVA2  00000425774BU7 650369SV5  00000589833BU3 770382CINN  00000465705BU2 904276CTH1  00000435786BU9 904276SP  00000476725BU7 910343EP1  00000612550BU4 973972LAB3  00000411137BU3 312792R4100  00000466829BU9 352333CATH1  00000444377BU6 757398INNOVA  00000444937BU7 77573841  00000437218BU1 4103323100  00000593466BU6 918OSUCATH2  00000453934BU2 417269SCLD  00000572422BU4 00052VAS04  00000467850BU4 378INV31  00000444914BU6 0910162006  00000464419BU1 0910163011  00000438629BU8 0910064007  00000417372BU0 250370INNOVA  00000429328BU8 416756INNOVA  00000491353BU9 514842AINNOVA  00000432687BU2 519685INNOVA  00000518369BU4 604682INNOVA1  00000537026BU7 604682INNOVA2  00000476531BU9 604875AINNOVA1  00000482420BU7 604875AINNOVA2  00000007C20329 905848INNOVA1  00000413716BU2 082406040022  00000600076BU4 082406070013  00000473100BU6 082416010005  00000501586BU2 082416040018  00000462912BU7 082416050002  00000585908BU7 082416090008  00000430320BU2 082416100006  00000461622BU3 082416100009  00000626767BU8 082416100065  00000446837BU7 082416120006  00000456458BU9 082416120007  00000459879BU3 082416140002  00000446285BU9 082416160005  00000480352BU4 082416160007  00000592459BU2 082416170004  00000461649BU6 082416200004  00000461008BU5 082416210004  00000559815BU6 082416210012  00000439906BU9 082416310003  00000544638BU0 082416310007  00000590598BU9 082416110007  00000446438BU4 082416130015  00000498661BU8 082416100012  00000424554BU4 082406120016  00000463449BU9 CS1006VA03  00000445594BU5 CS1054VA01  00000590862BU9 DK1066VA05  00000590866BU0 EG1843VA01  00000456800BU2 FI1015VA01  00000440111BU3 M2002624  00000438631BU4 M40480151  00000477885BU8 M4163030  00000000LS8-C1 R4211476  00000590317BU4 M4171816  00000412256BU0 407953XR01  00000465383BU8 HC4355XR06  00000576419BU6 HC4355XR07  00000514614BU7 HU1094VA01  00000582487BU5 083016102002169  00000467880BU1 083016102100795  00000590867BU8 083016860019010  00000501587BU0 IE1100VA01  00000476726BU5 A5605743  00000425390BU2 A5630803  00000594768BU4 A5195416  00000518355BU3 A5333037  00000608125BU1 A5365426  0000010C2M0154 A5401383  00000591390BU0 A5719853  0000010C2M0164 A5810824  00000463078BU6 A5697709  00000493966BU6 XF0153  00000564505BU6 YV0013  00000438669BU4 YV0022  00000439612BU3 YV0023  00000478097BU9 YV0035  00000547029BU9 YV0076  00000514380BU5 YX1483  00000459461BU0 YV0030  00000560738BU7 YV0363  00000612546BU2 YV0388  00000454589BU3 JO1000VA01  00000600863BU5 0850060744  00000560740BU3 0834160005  FMI12204DUMM01 28626INNOVA  00000445752BU9 477XVIN2553211  00000522945BU5 644INNOVA21  00000459368BU7 0920169006  00000435051BU8 NO1005VA05  0000009C2M0125 0843160016  00000476720BU8 600042VA02  00000441173BU2 787653INNOVA  FMI12204DUMM02 RU1138VA01  00000551948BU3 RU3053VA01  00000418618BU5 SA1188VA03  00000589400BU1 MEX68922  0000010C2M0176 MEX68928  00000593472BU4 083016808780010  00000580432BU3 34078VAS01  00000537448BU3 60001VAS01  00000428165BU5 00171VAS01  00000462909BU3 00234VAS01  00000424824BU1 77137VAS01  00000483163BU2 00282VAS02"
    }
  ]
}