  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Mahwah",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67495",
      "recalling_firm": "Mindray DS USA, Inc. dba Mindray North America",
      "address_1": "800 MacArthur Blvd",
      "address_2": "N/A",
      "postal_code": "07430-2001",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA Nationwide Distribution",
      "recall_number": "Z-1156-2014",
      "product_description": "Panaroma Central Station      Product Usage:  This device can view real time, store, print, graph, and trend patient clinical and demographic data. The device can also set independent alarm limits for data send by the bedside monitor.",
      "product_quantity": "732 units in the US",
      "reason_for_recall": "Mindray has identified two issues with the Panorama Central Station. Full Disclosure data may lag behind the Panorama system time clock when the Panorama Central  Station is in use with Mindray¿s V-Series Monitor.  Additionally, event indicators lag behind the Panorama system time clock when the Panorama Central Station is in use with a Panorama Telepack.",
      "recall_initiation_date": "20140108",
      "center_classification_date": "20140307",
      "termination_date": "20150121",
      "report_date": "20140319",
      "code_info": "p/n 0998-00-0708-01; 0998-UC-0708-01 and 0020-00-0205-XXX"
    }
  ]
}