  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67473",
      "recalling_firm": "Philips Medical Systems, Inc.",
      "address_1": "3000 Minuteman Rd",
      "address_2": "N/A",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide  Foreign:  AUSTRIA  BAHRAIN  BELGIUM  BRAZIL  CANADA  CHINA  EGYPT  ESTONIA  FINLAND  FRANCE  GERMANY  GHANA  ICELAND  INDIA  INDONESIA  IRELAND  ISRAEL  ITALY  JAPAN  JORDAN  KENYA  KOREA, REPUBLIC OF  LATVIA  LEBANON  LIBYAN ARAB JAMAHIRIYA  LITHUANIA  MALAYSIA  MEXICO  MOZAMBIQUE  MYANMAR (Burma)  NETHERLANDS  NETHERLANDS ANTILLES  NEW CALEDONIA  NEW ZEALAND  NORWAY  PAKISTAN  PERU  POLAND  PORTUGAL  QATAR  RUSSIAN FEDERATION  SAUDI ARABIA  SINGAPORE  SOUTH AFRICA  SPAIN  SWEDEN  SWITZERLAND  TAIWAN, PROVINCE OF CHINA  THAILAND  TURKMENISTAN  UNITED ARAB EMIRATES  UNITED KINGDOM  VIET NAM",
      "recall_number": "Z-1155-2014",
      "product_description": "Philips HeartStart MRx  Monitor/Defibrillator  Model numbers: M3535A, M3536A, M3536M,M3536M4,M3536M6,861288,861289,861464,861483,and 861491",
      "product_quantity": "3541 units",
      "reason_for_recall": "MRx may display a Red X in the Ready For Use (RFU) indicator during automated tests informing the user that a failure has been detected that may prevent the delivery of a shock and pacing",
      "recall_initiation_date": "20140218",
      "center_classification_date": "20140307",
      "termination_date": "20200911",
      "report_date": "20140319",
      "code_info": "Serial numbers within the range:  US00100204 to  US00570921"
    }
  ]
}