  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Deerfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87147",
      "recalling_firm": "Baxter Healthcare Corporation",
      "address_1": "1 Baxter Pkwy",
      "address_2": "",
      "postal_code": "60015-4625",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution including in the states of AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NY, OH, OR, PA, SD, TN, TX, UT, VA, WA, WI and the countries of Australia, Austria, Belgium, Canada, China, Croatia, Denmark, Finland, France, Germany, Israel, Italy, Japan, Jordan, Kenya, Korea (Republic Of), Kuwait, Lithuania, Malaysia, Malta, Mexico, Morocco, Netherlands, New Zealand, Norway, Poland, Spain, Sweden, United Kingdom.",
      "recall_number": "Z-1151-2021",
      "product_description": "The PrisMax System Version II Hemodialysis Delivery System: Automatic Repositioning System Pump Assembly, specifically tubing, Model Numbers:  a) 955626  b) 955558  c)  955701  PrisMax V2 Preventive Maintenance Kit, Product Code SC616  ARPS PUMP, Product Code SC6035  ARPS Pump Assembly, Product Code SC6053",
      "product_quantity": "2314 units",
      "reason_for_recall": "Variability in the performance of the tubing in the ARPS (Automatic Repositioning System) Pump Assembly which may lead to alarm situations during system self-test and therapy. This tubing is normally replaced during annual preventative maintenance.   Alarm Situations: During System Self-Test: Alarm code B1215 (ARPS Self-Test failure); During Therapy: B1273 (ARPS Leak), B1215 (ARPS Pressure self-test), B1116 (Return Pressure self-test Failure), B1539 (Access Pod Reposition Failure), B1541 (Effluent Pod Reposition Failure), B1540 (Effluent Filter Pod Reposition Failure). If an alarm occurs, it may lead to delay or interruption of therapy. In the event that therapy is terminated without returning blood to the patient, blood loss may occur.",
      "recall_initiation_date": "20210107",
      "center_classification_date": "20210226",
      "termination_date": "20240430",
      "report_date": "20210310",
      "code_info": "Model Numbers: serial number, UDI;  a) 955626: 101227, (01) 00085412639499 (21) 101227;  101280, (01) 00085412639499 (21) 101280;  101292, (01) 00085412639499 (21) 101292;  101293, (01) 00085412639499 (21) 101293;  101327, (01) 00085412639499 (21) 101327;  101348, (01) 00085412639499 (21) 101348;  101352, (01) 00085412639499 (21) 101352;  101353, (01) 00085412639499 (21) 101353;  101389, (01) 00085412639499 (21) 101389;  101391, (01) 00085412639499 (21) 101391;  101393, (01) 00085412639499 (21) 101393;  101394, (01) 00085412639499 (21) 101394;  101398, (01) 00085412639499 (21) 101398;  101422, (01) 00085412639499 (21) 101422;  101423, (01) 00085412639499 (21) 101423;  101450, (01) 00085412639499 (21) 101450;  101452, (01) 00085412639499 (21) 101452;  101456, (01) 00085412639499 (21) 101456;  101791, (01) 00085412639499 (21) 101791;  102142, (01) 00085412639499 (21) 102142;  102151, (01) 00085412639499 (21) 102151;  102152, (01) 00085412639499 (21) 102152;  102153, (01) 00085412639499 (21) 102153;  102154, (01) 00085412639499 (21) 102154;  102155, (01) 00085412639499 (21) 102155;  102159, (01) 00085412639499 (21) 102159;  102312, (01) 00085412639499 (21) 102312;  102318, (01) 00085412639499 (21) 102318;  102326, (01) 00085412639499 (21) 102326;  102328, (01) 00085412639499 (21) 102328;  102329, (01) 00085412639499 (21) 102329;  102330, (01) 00085412639499 (21) 102330;  102353, (01) 00085412639499 (21) 102353;  102356, (01) 00085412639499 (21) 102356;  102364, (01) 00085412639499 (21) 102364;  102365, (01) 00085412639499 (21) 102365;  102397, (01) 00085412639499 (21) 102397;  102399, (01) 00085412639499 (21) 102399;  102400, (01) 00085412639499 (21) 102400;  102402, (01) 00085412639499 (21) 102402;  102403, (01) 00085412639499 (21) 102403;  102404, (01) 00085412639499 (21) 102404;  102406, (01) 00085412639499 (21) 102406;  102407, (01) 00085412639499 (21) 102407;  102408, (01) 00085412639499 (21) 102408;  102409, (01) 00085412639499 (21) 102409;  102410, (01) 00085412639499 (21) 102410;  102411, (01) 00085412639499 (21) 102411;  102413, (01) 00085412639499 (21) 102413;  102414, (01) 00085412639499 (21) 102414;  102415, (01) 00085412639499 (21) 102415;  102416, (01) 00085412639499 (21) 102416;  102417, (01) 00085412639499 (21) 102417;  102418, (01) 00085412639499 (21) 102418;  102419, (01) 00085412639499 (21) 102419;  102420, (01) 00085412639499 (21) 102420;  102422, (01) 00085412639499 (21) 102422;  102423, (01) 00085412639499 (21) 102423;  102424, (01) 00085412639499 (21) 102424;  102425, (01) 00085412639499 (21) 102425;  102426, (01) 00085412639499 (21) 102426;  102457, (01) 00085412639499 (21) 102457;  102458, (01) 00085412639499 (21) 102458;  102459, (01) 00085412639499 (21) 102459;  102460, (01) 00085412639499 (21) 102460;  102461, (01) 00085412639499 (21) 102461;  102462, (01) 00085412639499 (21) 102462;  102463, (01) 00085412639499 (21) 102463;  102464, (01) 00085412639499 (21) 102464;  102465, (01) 00085412639499 (21) 102465;  102466, (01) 00085412639499 (21) 102466;  102467, (01) 00085412639499 (21) 102467;  102468, (01) 00085412639499 (21) 102468;  102469, (01) 00085412639499 (21) 102469;  102470, (01) 00085412639499 (21) 102470;  102471, (01) 00085412639499 (21) 102471;  102487, (01) 00085412639499 (21) 102487;  102488, (01) 00085412639499 (21) 102488;  102489, (01) 00085412639499 (21) 102489;  102490, (01) 00085412639499 (21) 102490;  102491, (01) 00085412639499 (21) 102491;  102492, (01) 00085412639499 (21) 102492;  102493, (01) 00085412639499 (21) 102493;  102494, (01) 00085412639499 (21) 102494;  102495, (01) 00085412639499 (21) 102495;  102496, (01) 00085412639499 (21) 102496;  102497, (01) 00085412639499 (21) 102497;  102498, (01) 00085412639499 (21) 102498;  102499, (01) 00085412639499 (21) 102499;  102500, (01) 00085412639499 (21) 102500;  102501, (01) 00085412639499 (21) 102501;  102876, (01) 00085412639499 (21) 102876;  102877, (01) 00085412639499 (21) 102877;  102878, (01) 00085412639499 (21) 102878;  102879, (01) 00085412639499 (21) 102879;  102880, (01) 00085412639499 (21) 102880;  102881, (01) 00085412639499 (21) 102881;  102882, (01) 00085412639499 (21) 102882;  102883, (01) 00085412639499 (21) 102883;  102884, (01) 00085412639499 (21) 102884;  102885, (01) 00085412639499 (21) 102885;  102886, (01) 00085412639499 (21) 102886;  102887, (01) 00085412639499 (21) 102887;  102888, (01) 00085412639499 (21) 102888;  102889, (01) 00085412639499 (21) 102889;  102890, (01) 00085412639499 (21) 102890;  102891, (01) 00085412639499 (21) 102891;  102892, (01) 00085412639499 (21) 102892;  102893, (01) 00085412639499 (21) 102893;  102894, (01) 00085412639499 (21) 102894;  102895, (01) 00085412639499 (21) 102895;  102896, (01) 00085412639499 (21) 102896;  102897, (01) 00085412639499 (21) 102897;  102898, (01) 00085412639499 (21) 102898;  102899, (01) 00085412639499 (21) 102899;  102900, (01) 00085412639499 (21) 102900;  102921, (01) 00085412639499 (21) 102921;  102922, (01) 00085412639499 (21) 102922;  102923, (01) 00085412639499 (21) 102923;  102924, (01) 00085412639499 (21) 102924;  102925, (01) 00085412639499 (21) 102925;  102926, (01) 00085412639499 (21) 102926;  102927, (01) 00085412639499 (21) 102927;  102928, (01) 00085412639499 (21) 102928;  102929, (01) 00085412639499 (21) 102929;  102930, (01) 00085412639499 (21) 102930;  103186, (01) 00085412639499 (21) 103186;  103187, (01) 00085412639499 (21) 103187;  103188, (01) 00085412639499 (21) 103188;  103189, (01) 00085412639499 (21) 103189;  103190, (01) 00085412639499 (21) 103190;  103191, (01) 00085412639499 (21) 103191;  103192, (01) 00085412639499 (21) 103192;  103193, (01) 00085412639499 (21) 103193;  103194, (01) 00085412639499 (21) 103194;  103195, (01) 00085412639499 (21) 103195;  103196, (01) 00085412639499 (21) 103196;  103197, (01) 00085412639499 (21) 103197;  103198, (01) 00085412639499 (21) 103198;  103199, (01) 00085412639499 (21) 103199;  103200, (01) 00085412639499 (21) 103200;  103201, (01) 00085412639499 (21) 103201;  103202, (01) 00085412639499 (21) 103202;  103203, (01) 00085412639499 (21) 103203;  103204, (01) 00085412639499 (21) 103204;  103205, (01) 00085412639499 (21) 103205;  103206, (01) 00085412639499 (21) 103206;  103207, (01) 00085412639499 (21) 103207;  103208, (01) 00085412639499 (21) 103208;  103209, (01) 00085412639499 (21) 103209;  103210, (01) 00085412639499 (21) 103210;  103211, (01) 00085412639499 (21) 103211;  103212, (01) 00085412639499 (21) 103212;  103213, (01) 00085412639499 (21) 103213;  103214, (01) 00085412639499 (21) 103214;  103215, (01) 00085412639499 (21) 103215;  103216, (01) 00085412639499 (21) 103216;  103217, (01) 00085412639499 (21) 103217;  103218, (01) 00085412639499 (21) 103218;  103219, (01) 00085412639499 (21) 103219;  103220, (01) 00085412639499 (21) 103220;  103221, (01) 00085412639499 (21) 103221;  103222, (01) 00085412639499 (21) 103222;  103223, (01) 00085412639499 (21) 103223;  103224, (01) 00085412639499 (21) 103224;  103225, (01) 00085412639499 (21) 103225;  103226, (01) 00085412639499 (21) 103226;  103227, (01) 00085412639499 (21) 103227;  103228, (01) 00085412639499 (21) 103228;  103229, (01) 00085412639499 (21) 103229;  103230, (01) 00085412639499 (21) 103230;  103231, (01) 00085412639499 (21) 103231;  103232, (01) 00085412639499 (21) 103232;  103233, (01) 00085412639499 (21) 103233;  103234, (01) 00085412639499 (21) 103234;  103235, (01) 00085412639499 (21) 103235;  103326, (01) 00085412639499 (21) 103326;  103327, (01) 00085412639499 (21) 103327;  103328, (01) 00085412639499 (21) 103328;  103329, (01) 00085412639499 (21) 103329;  103330, (01) 00085412639499 (21) 103330;  103331, (01) 00085412639499 (21) 103331;  103332, (01) 00085412639499 (21) 103332;  103333, (01) 00085412639499 (21) 103333;  103334, (01) 00085412639499 (21) 103334;  103335, (01) 00085412639499 (21) 103335;  103336, (01) 00085412639499 (21) 103336;  103337, (01) 00085412639499 (21) 103337;  103338, (01) 00085412639499 (21) 103338;  103339, (01) 00085412639499 (21) 103339;  103340, (01) 00085412639499 (21) 103340;  103341, (01) 00085412639499 (21) 103341;  103342, (01) 00085412639499 (21) 103342;  103343, (01) 00085412639499 (21) 103343;  103344, (01) 00085412639499 (21) 103344;  103345, (01) 00085412639499 (21) 103345;  103366, (01) 00085412639499 (21) 103366;  103367, (01) 00085412639499 (21) 103367;  103368, (01) 00085412639499 (21) 103368;  103369, (01) 00085412639499 (21) 103369;  103370, (01) 00085412639499 (21) 103370;  103371, (01) 00085412639499 (21) 103371;  103372, (01) 00085412639499 (21) 103372;  103373, (01) 00085412639499 (21) 103373;  103374, (01) 00085412639499 (21) 103374;  103375, (01) 00085412639499 (21) 103375;  103426, (01) 00085412639499 (21) 103426;  103427, (01) 00085412639499 (21) 103427;  103428, (01) 00085412639499 (21) 103428;  103429, (01) 00085412639499 (21) 103429;  103430, (01) 00085412639499 (21) 103430;  103431, (01) 00085412639499 (21) 103431;  103432, (01) 00085412639499 (21) 103432;  103433, (01) 00085412639499 (21) 103433;  103434, (01) 00085412639499 (21) 103434;  103435, (01) 00085412639499 (21) 103435;  103456, (01) 00085412639499 (21) 103456;  103457, (01) 00085412639499 (21) 103457;  103458, (01) 00085412639499 (21) 103458;  103459, (01) 00085412639499 (21) 103459;  103460, (01) 00085412639499 (21) 103460;  103461, (01) 00085412639499 (21) 103461;  103462, (01) 00085412639499 (21) 103462;  103463, (01) 00085412639499 (21) 103463;  103464, (01) 00085412639499 (21) 103464;  103465, (01) 00085412639499 (21) 103465;  103536, (01) 00085412639499 (21) 103536;  103537, (01) 00085412639499 (21) 103537;  103538, (01) 00085412639499 (21) 103538;  103539, (01) 00085412639499 (21) 103539;  103540, (01) 00085412639499 (21) 103540;  103541, (01) 00085412639499 (21) 103541;  103542, (01) 00085412639499 (21) 103542;  103543, (01) 00085412639499 (21) 103543;  103544, (01) 00085412639499 (21) 103544;  103545, (01) 00085412639499 (21) 103545;  103546, (01) 00085412639499 (21) 103546;  103547, (01) 00085412639499 (21) 103547;  103548, (01) 00085412639499 (21) 103548;  103549, (01) 00085412639499 (21) 103549;  103550, (01) 00085412639499 (21) 103550;  103551, (01) 00085412639499 (21) 103551;  103552, (01) 00085412639499 (21) 103552;  103553, (01) 00085412639499 (21) 103553;  103554, (01) 00085412639499 (21) 103554;  103555, (01) 00085412639499 (21) 103555;  103696, (01) 00085412639499 (21) 103696;  103697, (01) 00085412639499 (21) 103697;  103698, (01) 00085412639499 (21) 103698;  103699, (01) 00085412639499 (21) 103699;  103700, (01) 00085412639499 (21) 103700;  103701, (01) 00085412639499 (21) 103701;  103702, (01) 00085412639499 (21) 103702;  103703, (01) 00085412639499 (21) 103703;  103704, (01) 00085412639499 (21) 103704;  103705, (01) 00085412639499 (21) 103705;  103716, (01) 00085412639499 (21) 103716;  103717, (01) 00085412639499 (21) 103717;  103718, (01) 00085412639499 (21) 103718;  103719, (01) 00085412639499 (21) 103719;  103720, (01) 00085412639499 (21) 103720;  103721, (01) 00085412639499 (21) 103721;  103722, (01) 00085412639499 (21) 103722;  103723, (01) 00085412639499 (21) 103723;  103724, (01) 00085412639499 (21) 103724;  103725, (01) 00085412639499 (21) 103725;  103726, (01) 00085412639499 (21) 103726;  103727, (01) 00085412639499 (21) 103727;  103728, (01) 00085412639499 (21) 103728;  103729, (01) 00085412639499 (21) 103729;  103730, (01) 00085412639499 (21) 103730;  103731, (01) 00085412639499 (21) 103731;  103732, (01) 00085412639499 (21) 103732;  103733, (01) 00085412639499 (21) 103733;  103734, (01) 00085412639499 (21) 103734;  103735, (01) 00085412639499 (21) 103735;  103746, (01) 00085412639499 (21) 103746;  103747, (01) 00085412639499 (21) 103747;  103748, (01) 00085412639499 (21) 103748;  103749, (01) 00085412639499 (21) 103749;  103750, (01) 00085412639499 (21) 103750;  103751, (01) 00085412639499 (21) 103751;  103752, (01) 00085412639499 (21) 103752;  103753, (01) 00085412639499 (21) 103753;  103754, (01) 00085412639499 (21) 103754;  103755, (01) 00085412639499 (21) 103755;  103866, (01) 00085412639499 (21) 103866;  103867, (01) 00085412639499 (21) 103867;  103868, (01) 00085412639499 (21) 103868;  103869, (01) 00085412639499 (21) 103869;  103870, (01) 00085412639499 (21) 103870;  103871, (01) 00085412639499 (21) 103871;  103872, (01) 00085412639499 (21) 103872;  103873, (01) 00085412639499 (21) 103873;  103874, (01) 00085412639499 (21) 103874;  103875, (01) 00085412639499 (21) 103875;  103876, (01) 00085412639499 (21) 103876;  103877, (01) 00085412639499 (21) 103877;  103878, (01) 00085412639499 (21) 103878;  103879, (01) 00085412639499 (21) 103879;  103880, (01) 00085412639499 (21) 103880;  103881, (01) 00085412639499 (21) 103881;  103882, (01) 00085412639499 (21) 103882;  103883, (01) 00085412639499 (21) 103883;  103884, (01) 00085412639499 (21) 103884;  103885, (01) 00085412639499 (21) 103885;  103886, (01) 00085412639499 (21) 103886;  103887, (01) 00085412639499 (21) 103887;  103888, (01) 00085412639499 (21) 103888;  103889, (01) 00085412639499 (21) 103889;  103890, (01) 00085412639499 (21) 103890;  103891, (01) 00085412639499 (21) 103891;  103892, (01) 00085412639499 (21) 103892;  103893, (01) 00085412639499 (21) 103893;  103894, (01) 00085412639499 (21) 103894;  103895, (01) 00085412639499 (21) 103895;  103896, (01) 00085412639499 (21) 103896;  103897, (01) 00085412639499 (21) 103897;  103898, (01) 00085412639499 (21) 103898;  103899, (01) 00085412639499 (21) 103899;  103900, (01) 00085412639499 (21) 103900;  103901, (01) 00085412639499 (21) 103901;  103902, (01) 00085412639499 (21) 103902;  103903, (01) 00085412639499 (21) 103903;  103904, (01) 00085412639499 (21) 103904;  103905, (01) 00085412639499 (21) 103905;  103916, (01) 00085412639499 (21) 103916;  103917, (01) 00085412639499 (21) 103917;  103918, (01) 00085412639499 (21) 103918;  103919, (01) 00085412639499 (21) 103919;  103920, (01) 00085412639499 (21) 103920;  103921, (01) 00085412639499 (21) 103921;  103922, (01) 00085412639499 (21) 103922;  103923, (01) 00085412639499 (21) 103923;  103924, (01) 00085412639499 (21) 103924;  103925, (01) 00085412639499 (21) 103925;  103936, (01) 00085412639499 (21) 103936;  103937, (01) 00085412639499 (21) 103937;  103938, (01) 00085412639499 (21) 103938;  103939, (01) 00085412639499 (21) 103939;  103940, (01) 00085412639499 (21) 103940;  103941, (01) 00085412639499 (21) 103941;  103942, (01) 00085412639499 (21) 103942;  103943, (01) 00085412639499 (21) 103943;  103944, (01) 00085412639499 (21) 103944;  103945, (01) 00085412639499 (21) 103945;  104076, (01) 00085412639499 (21) 104076;  104077, (01) 00085412639499 (21) 104077;  104078, (01) 00085412639499 (21) 104078;  104079, (01) 00085412639499 (21) 104079;  104080, (01) 00085412639499 (21) 104080;  104081, (01) 00085412639499 (21) 104081;  104082, (01) 00085412639499 (21) 104082;  104083, (01) 00085412639499 (21) 104083;  104084, (01) 00085412639499 (21) 104084;  104085, (01) 00085412639499 (21) 104085;  104106, (01) 00085412639499 (21) 104106;  104107, (01) 00085412639499 (21) 104107;  104108, (01) 00085412639499 (21) 104108;  104109, (01) 00085412639499 (21) 104109;  104110, (01) 00085412639499 (21) 104110;  104111, (01) 00085412639499 (21) 104111;  104112, (01) 00085412639499 (21) 104112;  104113, (01) 00085412639499 (21) 104113;  104114, (01) 00085412639499 (21) 104114;  104115, (01) 00085412639499 (21) 104115;  104126, (01) 00085412639499 (21) 104126;  104127, (01) 00085412639499 (21) 104127;  104128, (01) 00085412639499 (21) 104128;  104129, (01) 00085412639499 (21) 104129;  104130, (01) 00085412639499 (21) 104130;  104131, (01) 00085412639499 (21) 104131;  104132, (01) 00085412639499 (21) 104132;  104133, (01) 00085412639499 (21) 104133;  104134, (01) 00085412639499 (21) 104134;  104135, (01) 00085412639499 (21) 104135;  104146, (01) 00085412639499 (21) 104146;  104147, (01) 00085412639499 (21) 104147;  104148, (01) 00085412639499 (21) 104148;  104149, (01) 00085412639499 (21) 104149;  104150, (01) 00085412639499 (21) 104150;  104151, (01) 00085412639499 (21) 104151;  104152, (01) 00085412639499 (21) 104152;  104153, (01) 00085412639499 (21) 104153;  104154, (01) 00085412639499 (21) 104154;  104155, (01) 00085412639499 (21) 104155;  104206, (01) 00085412639499 (21) 104206;  104207, (01) 00085412639499 (21) 104207;  104208, (01) 00085412639499 (21) 104208;  104209, (01) 00085412639499 (21) 104209;  104210, (01) 00085412639499 (21) 104210;  104211, (01) 00085412639499 (21) 104211;  104212, (01) 00085412639499 (21) 104212;  104213, (01) 00085412639499 (21) 104213;  104214, (01) 00085412639499 (21) 104214;  104215, (01) 00085412639499 (21) 104215;  104226, (01) 00085412639499 (21) 104226;  104227, (01) 00085412639499 (21) 104227;  104228, (01) 00085412639499 (21) 104228;  104229, (01) 00085412639499 (21) 104229;  104230, (01) 00085412639499 (21) 104230;  104231, (01) 00085412639499 (21) 104231;  104232, (01) 00085412639499 (21) 104232;  104233, (01) 00085412639499 (21) 104233;  104234, (01) 00085412639499 (21) 104234;  104235, (01) 00085412639499 (21) 104235;  104326, (01) 00085412639499 (21) 104326;  104327, (01) 00085412639499 (21) 104327;  104328, (01) 00085412639499 (21) 104328;  104329, (01) 00085412639499 (21) 104329;  104330, (01) 00085412639499 (21) 104330;  104331, (01) 00085412639499 (21) 104331;  104332, (01) 00085412639499 (21) 104332;  104333, (01) 00085412639499 (21) 104333;  104334, (01) 00085412639499 (21) 104334;  104335, (01) 00085412639499 (21) 104335;  104346, (01) 00085412639499 (21) 104346;  104347, (01) 00085412639499 (21) 104347;  104348, (01) 00085412639499 (21) 104348;  104349, (01) 00085412639499 (21) 104349;  104350, (01) 00085412639499 (21) 104350;  104351, (01) 00085412639499 (21) 104351;  104352, (01) 00085412639499 (21) 104352;  104353, (01) 00085412639499 (21) 104353;  104354, (01) 00085412639499 (21) 104354;  104355, (01) 00085412639499 (21) 104355;  104376, (01) 00085412639499 (21) 104376;  104377, (01) 00085412639499 (21) 104377;  104378, (01) 00085412639499 (21) 104378;  104379, (01) 00085412639499 (21) 104379;  104380, (01) 00085412639499 (21) 104380;  104381, (01) 00085412639499 (21) 104381;  104382, (01) 00085412639499 (21) 104382;  104383, (01) 00085412639499 (21) 104383;  104384, (01) 00085412639499 (21) 104384;  104385, (01) 00085412639499 (21) 104385;  104656, (01) 00085412639499 (21) 104656;  104657, (01) 00085412639499 (21) 104657;  104658, (01) 00085412639499 (21) 104658;  104659, (01) 00085412639499 (21) 104659;  104660, (01) 00085412639499 (21) 104660;  104661, (01) 00085412639499 (21) 104661;  104662, (01) 00085412639499 (21) 104662;  104663, (01) 00085412639499 (21) 104663;  104664, (01) 00085412639499 (21) 104664;  104665, (01) 00085412639499 (21) 104665;  104696, (01) 00085412639499 (21) 104696;  104697, (01) 00085412639499 (21) 104697;  104698, (01) 00085412639499 (21) 104698;  104699, (01) 00085412639499 (21) 104699;  104700, (01) 00085412639499 (21) 104700;  104701, (01) 00085412639499 (21) 104701;  104702, (01) 00085412639499 (21) 104702;  104703, (01) 00085412639499 (21) 104703;  104704, (01) 00085412639499 (21) 104704;  104705, (01) 00085412639499 (21) 104705;  104716, (01) 00085412639499 (21) 104716;  104717, (01) 00085412639499 (21) 104717;  104718, (01) 00085412639499 (21) 104718;  104719, (01) 00085412639499 (21) 104719;  104720, (01) 00085412639499 (21) 104720;  104721, (01) 00085412639499 (21) 104721;  104722, (01) 00085412639499 (21) 104722;  104723, (01) 00085412639499 (21) 104723;  104724, (01) 00085412639499 (21) 104724;  104725, (01) 00085412639499 (21) 104725;  104746, (01) 00085412639499 (21) 104746;  104747, (01) 00085412639499 (21) 104747;  104748, (01) 00085412639499 (21) 104748;  104749, (01) 00085412639499 (21) 104749;  104750, (01) 00085412639499 (21) 104750;  104751, (01) 00085412639499 (21) 104751;  104752, (01) 00085412639499 (21) 104752;  104753, (01) 00085412639499 (21) 104753;  104754, (01) 00085412639499 (21) 104754;  104755, (01) 00085412639499 (21) 104755;  104856, (01) 00085412639499 (21) 104856;  104857, (01) 00085412639499 (21) 104857;  104858, (01) 00085412639499 (21) 104858;  104859, (01) 00085412639499 (21) 104859;  104860, (01) 00085412639499 (21) 104860;  104861, (01) 00085412639499 (21) 104861;  104862, (01) 00085412639499 (21) 104862;  104863, (01) 00085412639499 (21) 104863;  104864, (01) 00085412639499 (21) 104864;  104865, (01) 00085412639499 (21) 104865;  104886, (01) 00085412639499 (21) 104886;  104887, (01) 00085412639499 (21) 104887;  104888, (01) 00085412639499 (21) 104888;  104889, (01) 00085412639499 (21) 104889;  104890, (01) 00085412639499 (21) 104890;  104891, (01) 00085412639499 (21) 104891;  104892, (01) 00085412639499 (21) 104892;  104893, (01) 00085412639499 (21) 104893;  104894, (01) 00085412639499 (21) 104894;  104895, (01) 00085412639499 (21) 104895;  105006, (01) 00085412639499 (21) 105006;  105007, (01) 00085412639499 (21) 105007;  105008, (01) 00085412639499 (21) 105008;  105009, (01) 00085412639499 (21) 105009;  105010, (01) 00085412639499 (21) 105010;  105011, (01) 00085412639499 (21) 105011;  105012, (01) 00085412639499 (21) 105012;  105013, (01) 00085412639499 (21) 105013;  105014, (01) 00085412639499 (21) 105014;  105015, (01) 00085412639499 (21) 105015;  105016, (01) 00085412639499 (21) 105016;  105017, (01) 00085412639499 (21) 105017;  105018, (01) 00085412639499 (21) 105018;  105019, (01) 00085412639499 (21) 105019;  105020, (01) 00085412639499 (21) 105020;  105021, (01) 00085412639499 (21) 105021;  105022, (01) 00085412639499 (21) 105022;  105023, (01) 00085412639499 (21) 105023;  105024, (01) 00085412639499 (21) 105024;  105025, (01) 00085412639499 (21) 105025;  105036, (01) 00085412639499 (21) 105036;  105037, (01) 00085412639499 (21) 105037;  105038, (01) 00085412639499 (21) 105038;  105039, (01) 00085412639499 (21) 105039;  105040, (01) 00085412639499 (21) 105040;  105041, (01) 00085412639499 (21) 105041;  105042, (01) 00085412639499 (21) 105042;  105043, (01) 00085412639499 (21) 105043;  105044, (01) 00085412639499 (21) 105044;  105045, (01) 00085412639499 (21) 105045;  105246, (01) 00085412639499 (21) 105246;  105247, (01) 00085412639499 (21) 105247;  105248, (01) 00085412639499 (21) 105248;  105249, (01) 00085412639499 (21) 105249;  105250, (01) 00085412639499 (21) 105250;  105251, (01) 00085412639499 (21) 105251;  105252, (01) 00085412639499 (21) 105252;  105253, (01) 00085412639499 (21) 105253;  105254, (01) 00085412639499 (21) 105254;  105255, (01) 00085412639499 (21) 105255;  105256, (01) 00085412639499 (21) 105256;  105257, (01) 00085412639499 (21) 105257;  105258, (01) 00085412639499 (21) 105258;  105259, (01) 00085412639499 (21) 105259;  105260, (01) 00085412639499 (21) 105260;  105261, (01) 00085412639499 (21) 105261;  105262, (01) 00085412639499 (21) 105262;  105263, (01) 00085412639499 (21) 105263;  105264, (01) 00085412639499 (21) 105264;  105265, (01) 00085412639499 (21) 105265;  105276, (01) 00085412639499 (21) 105276;  105277, (01) 00085412639499 (21) 105277;  105278, (01) 00085412639499 (21) 105278;  105279, (01) 00085412639499 (21) 105279;  105280, (01) 00085412639499 (21) 105280;  105281, (01) 00085412639499 (21) 105281;  105282, (01) 00085412639499 (21) 105282;  105283, (01) 00085412639499 (21) 105283;  105284, (01) 00085412639499 (21) 105284;  105285, (01) 00085412639499 (21) 105285;  105496, (01) 00085412639499 (21) 105496;  105497, (01) 00085412639499 (21) 105497;  105498, (01) 00085412639499 (21) 105498;  105499, (01) 00085412639499 (21) 105499;  105500, (01) 00085412639499 (21) 105500;  105501, (01) 00085412639499 (21) 105501;  105502, (01) 00085412639499 (21) 105502;  105503, (01) 00085412639499 (21) 105503;  105504, (01) 00085412639499 (21) 105504;  105505, (01) 00085412639499 (21) 105505;  105526, (01) 00085412639499 (21) 105526;  105527, (01) 00085412639499 (21) 105527;  105528, (01) 00085412639499 (21) 105528;  105529, (01) 00085412639499 (21) 105529;  105530, (01) 00085412639499 (21) 105530;  105531, (01) 00085412639499 (21) 105531;  105532, (01) 00085412639499 (21) 105532;  105533, (01) 00085412639499 (21) 105533;  105534, (01) 00085412639499 (21) 105534;  105535, (01) 00085412639499 (21) 105535;  105626, (01) 00085412639499 (21) 105626;  105627, (01) 00085412639499 (21) 105627;  105628, (01) 00085412639499 (21) 105628;  105629, (01) 00085412639499 (21) 105629;  105630, (01) 00085412639499 (21) 105630;  105631, (01) 00085412639499 (21) 105631;  105632, (01) 00085412639499 (21) 105632;  105633, (01) 00085412639499 (21) 105633;  105634, (01) 00085412639499 (21) 105634;  105635, (01) 00085412639499 (21) 105635;  105822, (01) 00085412639499 (21) 105822;  105823, (01) 00085412639499 (21) 105823;  105824, (01) 00085412639499 (21) 105824;  105825, (01) 00085412639499 (21) 105825;  105826, (01) 00085412639499 (21) 105826;  105827, (01) 00085412639499 (21) 105827;  105828, (01) 00085412639499 (21) 105828;  105829, (01) 00085412639499 (21) 105829;  105830, (01) 00085412639499 (21) 105830;  105831, (01) 00085412639499 (21) 105831;  106062, (01) 00085412639499 (21) 106062;  106063, (01) 00085412639499 (21) 106063;  106064, (01) 00085412639499 (21) 106064;  106067, (01) 00085412639499 (21) 106067;  106068, (01) 00085412639499 (21) 106068;  106069, (01) 00085412639499 (21) 106069;  106070, (01) 00085412639499 (21) 106070 (603 units).    b) 955558: 100266, (01) 07332414124359 (21) 100266;  100267, (01) 07332414124359 (21) 100267;  100269, (01) 07332414124359 (21) 100269;  100271, (01) 07332414124359 (21) 100271;  100272, (01) 07332414124359 (21) 100272;  100273, (01) 07332414124359 (21) 100273;  100274, (01) 07332414124359 (21) 100274;  100275, (01) 07332414124359 (21) 100275;  100279, (01) 07332414124359 (21) 100279;  100300, (01) 07332414124359 (21) 100300;  100301, (01) 07332414124359 (21) 100301;  100303, (01) 07332414124359 (21) 100303;  100304, (01) 07332414124359 (21) 100304;  100316, (01) 07332414124359 (21) 100316;  100317, (01) 07332414124359 (21) 100317;  100318, (01) 07332414124359 (21) 100318;  100320, (01) 07332414124359 (21) 100320;  100331, (01) 07332414124359 (21) 100331;  100332, (01) 07332414124359 (21) 100332;  100333, (01) 07332414124359 (21) 100333;  100335, (01) 07332414124359 (21) 100335;  100336, (01) 07332414124359 (21) 100336;  100337, (01) 07332414124359 (21) 100337;  100340, (01) 07332414124359 (21) 100340;  100351, (01) 07332414124359 (21) 100351;  100352, (01) 07332414124359 (21) 100352;  100353, (01) 07332414124359 (21) 100353;  100371, (01) 07332414124359 (21) 100371;  100373, (01) 07332414124359 (21) 100373;  100375, (01) 07332414124359 (21) 100375;  100376, (01) 07332414124359 (21) 100376;  100377, (01) 07332414124359 (21) 100377;  100381, (01) 07332414124359 (21) 100381;  100384, (01) 07332414124359 (21) 100384;  100385, (01) 07332414124359 (21) 100385;  100387, (01) 07332414124359 (21) 100387;  100388, (01) 07332414124359 (21) 100388;  100389, (01) 07332414124359 (21) 100389;  100390, (01) 07332414124359 (21) 100390;  100406, (01) 07332414124359 (21) 100406;  100408, (01) 07332414124359 (21) 100408;  100410, (01) 07332414124359 (21) 100410;  100411, (01) 07332414124359 (21) 100411;  100412, (01) 07332414124359 (21) 100412;  100413, (01) 07332414124359 (21) 100413;  100414, (01) 07332414124359 (21) 100414;  100432, (01) 07332414124359 (21) 100432;  100433, (01) 07332414124359 (21) 100433;  100434, (01) 07332414124359 (21) 100434;  100435, (01) 07332414124359 (21) 100435;  100461, (01) 07332414124359 (21) 100461;  100465, (01) 07332414124359 (21) 100465;  100467, (01) 07332414124359 (21) 100467;  100469, (01) 07332414124359 (21) 100469;  100471, (01) 07332414124359 (21) 100471;  100480, (01) 07332414124359 (21) 100480;  100481, (01) 07332414124359 (21) 100481;  100483, (01) 07332414124359 (21) 100483;  100484, (01) 07332414124359 (21) 100484;  100485, (01) 07332414124359 (21) 100485;  100508, (01) 07332414124359 (21) 100508;  100510, (01) 07332414124359 (21) 100510;  100511, (01) 07332414124359 (21) 100511;  100512, (01) 07332414124359 (21) 100512;  100513, (01) 07332414124359 (21) 100513;  100514, (01) 07332414124359 (21) 100514;  100515, (01) 07332414124359 (21) 100515;  100517, (01) 07332414124359 (21) 100517;  100518, (01) 07332414124359 (21) 100518;  100519, (01) 07332414124359 (21) 100519;  100520, (01) 07332414124359 (21) 100520;  100521, (01) 07332414124359 (21) 100521;  100523, (01) 07332414124359 (21) 100523;  100524, (01) 07332414124359 (21) 100524;  100525, (01) 07332414124359 (21) 100525;  100526, (01) 07332414124359 (21) 100526;  100528, (01) 07332414124359 (21) 100528;  100556, (01) 07332414124359 (21) 100556;  100557, (01) 07332414124359 (21) 100557;  100558, (01) 07332414124359 (21) 100558;  100559, (01) 07332414124359 (21) 100559;  100560, (01) 07332414124359 (21) 100560;  100561, (01) 07332414124359 (21) 100561;  100563, (01) 07332414124359 (21) 100563;  100565, (01) 07332414124359 (21) 100565;  100568, (01) 07332414124359 (21) 100568;  100570, (01) 07332414124359 (21) 100570;  100571, (01) 07332414124359 (21) 100571;  100573, (01) 07332414124359 (21) 100573;  100574, (01) 07332414124359 (21) 100574;  100576, (01) 07332414124359 (21) 100576;  100577, (01) 07332414124359 (21) 100577;  100578, (01) 07332414124359 (21) 100578;  100609, (01) 07332414124359 (21) 100609;  100611, (01) 07332414124359 (21) 100611;  100613, (01) 07332414124359 (21) 100613;  100615, (01) 07332414124359 (21) 100615;  100616, (01) 07332414124359 (21) 100616;  100617, (01) 07332414124359 (21) 100617;  100618, (01) 07332414124359 (21) 100618;  100620, (01) 07332414124359 (21) 100620;  100621, (01) 07332414124359 (21) 100621;  100622, (01) 07332414124359 (21) 100622;  100623, (01) 07332414124359 (21) 100623;  100624, (01) 07332414124359 (21) 100624;  100625, (01) 07332414124359 (21) 100625;  100629, (01) 07332414124359 (21) 100629;  100656, (01) 07332414124359 (21) 100656;  100657, (01) 07332414124359 (21) 100657;  100659, (01) 07332414124359 (21) 100659;  100660, (01) 07332414124359 (21) 100660;  100661, (01) 07332414124359 (21) 100661;  100662, (01) 07332414124359 (21) 100662;  100664, (01) 07332414124359 (21) 100664;  100667, (01) 07332414124359 (21) 100667;  100668, (01) 07332414124359 (21) 100668;  100670, (01) 07332414124359 (21) 100670;  100671, (01) 07332414124359 (21) 100671;  100673, (01) 07332414124359 (21) 100673;  100674, (01) 07332414124359 (21) 100674;  100676, (01) 07332414124359 (21) 100676;  100678, (01) 07332414124359 (21) 100678;  100679, (01) 07332414124359 (21) 100679;  100681, (01) 07332414124359 (21) 100681;  100682, (01) 07332414124359 (21) 100682;  100683, (01) 07332414124359 (21) 100683;  100685, (01) 07332414124359 (21) 100685;  100687, (01) 07332414124359 (21) 100687;  100688, (01) 07332414124359 (21) 100688;  100689, (01) 07332414124359 (21) 100689;  100692, (01) 07332414124359 (21) 100692;  100693, (01) 07332414124359 (21) 100693;  100696, (01) 07332414124359 (21) 100696;  100697, (01) 07332414124359 (21) 100697;  100698, (01) 07332414124359 (21) 100698;  100699, (01) 07332414124359 (21) 100699;  100700, (01) 07332414124359 (21) 100700;  100701, (01) 07332414124359 (21) 100701;  100702, (01) 07332414124359 (21) 100702;  100703, (01) 07332414124359 (21) 100703;  100705, (01) 07332414124359 (21) 100705;  100707, (01) 07332414124359 (21) 100707;  100712, (01) 07332414124359 (21) 100712;  100713, (01) 07332414124359 (21) 100713;  100715, (01) 07332414124359 (21) 100715;  100716, (01) 07332414124359 (21) 100716;  100721, (01) 07332414124359 (21) 100721;  100722, (01) 07332414124359 (21) 100722;  100728, (01) 07332414124359 (21) 100728;  100730, (01) 07332414124359 (21) 100730;  100731, (01) 07332414124359 (21) 100731;  100733, (01) 07332414124359 (21) 100733;  100736, (01) 07332414124359 (21) 100736;  100745, (01) 07332414124359 (21) 100745;  100746, (01) 07332414124359 (21) 100746;  100750, (01) 07332414124359 (21) 100750;  100751, (01) 07332414124359 (21) 100751;  100754, (01) 07332414124359 (21) 100754;  100834, (01) 07332414124359 (21) 100834;  100835, (01) 07332414124359 (21) 100835;  100836, (01) 07332414124359 (21) 100836;  100838, (01) 07332414124359 (21) 100838;  100842, (01) 07332414124359 (21) 100842;  100843, (01) 07332414124359 (21) 100843;  100846, (01) 07332414124359 (21) 100846;  100847, (01) 07332414124359 (21) 100847;  100907, (01) 07332414124359 (21) 100907;  100908, (01) 07332414124359 (21) 100908;  100909, (01) 07332414124359 (21) 100909;  100912, (01) 07332414124359 (21) 100912;  100913, (01) 07332414124359 (21) 100913;  100929, (01) 0733241",
      "more_code_info": "4124359 (21) 100929;  100930, (01) 07332414124359 (21) 100930;  100956, (01) 07332414124359 (21) 100956;  100958, (01) 07332414124359 (21) 100958;  100959, (01) 07332414124359 (21) 100959;  100962, (01) 07332414124359 (21) 100962;  100963, (01) 07332414124359 (21) 100963;  100964, (01) 07332414124359 (21) 100964;  100966, (01) 07332414124359 (21) 100966;  100967, (01) 07332414124359 (21) 100967;  100968, (01) 07332414124359 (21) 100968;  100970, (01) 07332414124359 (21) 100970;  100971, (01) 07332414124359 (21) 100971;  100972, (01) 07332414124359 (21) 100972;  100974, (01) 07332414124359 (21) 100974;  100975, (01) 07332414124359 (21) 100975;  101036, (01) 07332414124359 (21) 101036;  101037, (01) 07332414124359 (21) 101037;  101038, (01) 07332414124359 (21) 101038;  101039, (01) 07332414124359 (21) 101039;  101040, (01) 07332414124359 (21) 101040;  101041, (01) 07332414124359 (21) 101041;  101042, (01) 07332414124359 (21) 101042;  101045, (01) 07332414124359 (21) 101045;  101046, (01) 07332414124359 (21) 101046;  101047, (01) 07332414124359 (21) 101047;  101048, (01) 07332414124359 (21) 101048;  101049, (01) 07332414124359 (21) 101049;  101050, (01) 07332414124359 (21) 101050;  101051, (01) 07332414124359 (21) 101051;  101052, (01) 07332414124359 (21) 101052;  101053, (01) 07332414124359 (21) 101053;  101054, (01) 07332414124359 (21) 101054;  101057, (01) 07332414124359 (21) 101057;  101058, (01) 07332414124359 (21) 101058;  101059, (01) 07332414124359 (21) 101059;  101063, (01) 07332414124359 (21) 101063;  101064, (01) 07332414124359 (21) 101064;  101066, (01) 07332414124359 (21) 101066;  101067, (01) 07332414124359 (21) 101067;  101071, (01) 07332414124359 (21) 101071;  101072, (01) 07332414124359 (21) 101072;  101073, (01) 07332414124359 (21) 101073;  101074, (01) 07332414124359 (21) 101074;  101075, (01) 07332414124359 (21) 101075;  101076, (01) 07332414124359 (21) 101076;  101078, (01) 07332414124359 (21) 101078;  101080, (01) 07332414124359 (21) 101080;  101163, (01) 07332414124359 (21) 101163;  101164, (01) 07332414124359 (21) 101164;  101165, (01) 07332414124359 (21) 101165;  101166, (01) 07332414124359 (21) 101166;  101167, (01) 07332414124359 (21) 101167;  101193, (01) 07332414124359 (21) 101193;  101194, (01) 07332414124359 (21) 101194;  101195, (01) 07332414124359 (21) 101195;  101196, (01) 07332414124359 (21) 101196;  101197, (01) 07332414124359 (21) 101197;  101201, (01) 07332414124359 (21) 101201;  101204, (01) 07332414124359 (21) 101204;  101205, (01) 07332414124359 (21) 101205;  101207, (01) 07332414124359 (21) 101207;  101210, (01) 07332414124359 (21) 101210;  101213, (01) 07332414124359 (21) 101213;  101215, (01) 07332414124359 (21) 101215;  101216, (01) 07332414124359 (21) 101216;  101217, (01) 07332414124359 (21) 101217;  101297, (01) 07332414124359 (21) 101297;  101298, (01) 07332414124359 (21) 101298;  101299, (01) 07332414124359 (21) 101299;  101300, (01) 07332414124359 (21) 101300;  101301, (01) 07332414124359 (21) 101301;  101302, (01) 07332414124359 (21) 101302;  101305, (01) 07332414124359 (21) 101305;  101306, (01) 07332414124359 (21) 101306;  101307, (01) 07332414124359 (21) 101307;  101308, (01) 07332414124359 (21) 101308;  101309, (01) 07332414124359 (21) 101309;  101310, (01) 07332414124359 (21) 101310;  101312, (01) 07332414124359 (21) 101312;  101315, (01) 07332414124359 (21) 101315;  101318, (01) 07332414124359 (21) 101318;  101493, (01) 07332414124359 (21) 101493;  101497, (01) 07332414124359 (21) 101497;  101498, (01) 07332414124359 (21) 101498;  101500, (01) 07332414124359 (21) 101500;  101501, (01) 07332414124359 (21) 101501;  101502, (01) 07332414124359 (21) 101502;  101503, (01) 07332414124359 (21) 101503;  101505, (01) 07332414124359 (21) 101505;  101506, (01) 07332414124359 (21) 101506;  101507, (01) 07332414124359 (21) 101507;  101525, (01) 07332414124359 (21) 101525;  101526, (01) 07332414124359 (21) 101526;  101527, (01) 07332414124359 (21) 101527;  101528, (01) 07332414124359 (21) 101528;  101530, (01) 07332414124359 (21) 101530;  101535, (01) 07332414124359 (21) 101535;  101536, (01) 07332414124359 (21) 101536;  101563, (01) 07332414124359 (21) 101563;  101565, (01) 07332414124359 (21) 101565;  101566, (01) 07332414124359 (21) 101566;  101569, (01) 07332414124359 (21) 101569;  101572, (01) 07332414124359 (21) 101572;  101576, (01) 07332414124359 (21) 101576;  101578, (01) 07332414124359 (21) 101578;  101579, (01) 07332414124359 (21) 101579;  101587, (01) 07332414124359 (21) 101587;  101591, (01) 07332414124359 (21) 101591;  101592, (01) 07332414124359 (21) 101592;  101618, (01) 07332414124359 (21) 101618;  101619, (01) 07332414124359 (21) 101619;  101621, (01) 07332414124359 (21) 101621;  101624, (01) 07332414124359 (21) 101624;  101625, (01) 07332414124359 (21) 101625;  101626, (01) 07332414124359 (21) 101626;  101627, (01) 07332414124359 (21) 101627;  101628, (01) 07332414124359 (21) 101628;  101629, (01) 07332414124359 (21) 101629;  101631, (01) 07332414124359 (21) 101631;  101650, (01) 07332414124359 (21) 101650;  101654, (01) 07332414124359 (21) 101654;  101655, (01) 07332414124359 (21) 101655;  101661, (01) 07332414124359 (21) 101661;  101662, (01) 07332414124359 (21) 101662;  101663, (01) 07332414124359 (21) 101663;  101665, (01) 07332414124359 (21) 101665;  101670, (01) 07332414124359 (21) 101670;  101674, (01) 07332414124359 (21) 101674;  101677, (01) 07332414124359 (21) 101677;  101687, (01) 07332414124359 (21) 101687;  101703, (01) 07332414124359 (21) 101703;  101712, (01) 07332414124359 (21) 101712;  101713, (01) 07332414124359 (21) 101713;  101717, (01) 07332414124359 (21) 101717;  101718, (01) 07332414124359 (21) 101718;  101719, (01) 07332414124359 (21) 101719;  101722, (01) 07332414124359 (21) 101722;  101725, (01) 07332414124359 (21) 101725;  101726, (01) 07332414124359 (21) 101726;  101727, (01) 07332414124359 (21) 101727;  101764, (01) 07332414124359 (21) 101764;  101772, (01) 07332414124359 (21) 101772;  101773, (01) 07332414124359 (21) 101773;  101775, (01) 07332414124359 (21) 101775;  101776, (01) 07332414124359 (21) 101776;  101778, (01) 07332414124359 (21) 101778;  101787, (01) 07332414124359 (21) 101787;  101788, (01) 07332414124359 (21) 101788;  101799, (01) 07332414124359 (21) 101799;  101802, (01) 07332414124359 (21) 101802;  101804, (01) 07332414124359 (21) 101804;  101805, (01) 07332414124359 (21) 101805;  101809, (01) 07332414124359 (21) 101809;  101811, (01) 07332414124359 (21) 101811;  101812, (01) 07332414124359 (21) 101812;  101815, (01) 07332414124359 (21) 101815;  101822, (01) 07332414124359 (21) 101822;  101823, (01) 07332414124359 (21) 101823;  101824, (01) 07332414124359 (21) 101824;  101866, (01) 07332414124359 (21) 101866;  101872, (01) 07332414124359 (21) 101872;  101892, (01) 07332414124359 (21) 101892;  101903, (01) 07332414124359 (21) 101903;  101926, (01) 07332414124359 (21) 101926;  101928, (01) 07332414124359 (21) 101928;  101931, (01) 07332414124359 (21) 101931;  101937, (01) 07332414124359 (21) 101937;  101939, (01) 07332414124359 (21) 101939;  101941, (01) 07332414124359 (21) 101941;  101956, (01) 07332414124359 (21) 101956;  101958, (01) 07332414124359 (21) 101958;  101959, (01) 07332414124359 (21) 101959;  101965, (01) 07332414124359 (21) 101965;  101966, (01) 07332414124359 (21) 101966;  101970, (01) 07332414124359 (21) 101970;  101997, (01) 07332414124359 (21) 101997;  102004, (01) 07332414124359 (21) 102004;  102006, (01) 07332414124359 (21) 102006;  102011, (01) 07332414124359 (21) 102011;  102012, (01) 07332414124359 (21) 102012;  102014, (01) 07332414124359 (21) 102014;  102042, (01) 07332414124359 (21) 102042;  102044, (01) 07332414124359 (21) 102044;  102045, (01) 07332414124359 (21) 102045;  102170, (01) 07332414124359 (21) 102170;  102172, (01) 07332414124359 (21) 102172;  102197, (01) 07332414124359 (21) 102197;  102198, (01) 07332414124359 (21) 102198;  102200, (01) 07332414124359 (21) 102200;  102201, (01) 07332414124359 (21) 102201;  102204, (01) 07332414124359 (21) 102204;  102206, (01) 07332414124359 (21) 102206;  102210, (01) 07332414124359 (21) 102210;  102211, (01) 07332414124359 (21) 102211;  102212, (01) 07332414124359 (21) 102212;  102213, (01) 07332414124359 (21) 102213;  102214, (01) 07332414124359 (21) 102214;  102255, (01) 07332414124359 (21) 102255;  102256, (01) 07332414124359 (21) 102256;  102257, (01) 07332414124359 (21) 102257;  102258, (01) 07332414124359 (21) 102258;  102259, (01) 07332414124359 (21) 102259;  102260, (01) 07332414124359 (21) 102260;  102261, (01) 07332414124359 (21) 102261;  102262, (01) 07332414124359 (21) 102262;  102263, (01) 07332414124359 (21) 102263;  102264, (01) 07332414124359 (21) 102264;  102265, (01) 07332414124359 (21) 102265;  102266, (01) 07332414124359 (21) 102266;  102267, (01) 07332414124359 (21) 102267;  102268, (01) 07332414124359 (21) 102268;  102270, (01) 07332414124359 (21) 102270;  102271, (01) 07332414124359 (21) 102271;  102272, (01) 07332414124359 (21) 102272;  102273, (01) 07332414124359 (21) 102273;  102274, (01) 07332414124359 (21) 102274;  102275, (01) 07332414124359 (21) 102275;  102276, (01) 07332414124359 (21) 102276;  102277, (01) 07332414124359 (21) 102277;  102278, (01) 07332414124359 (21) 102278;  102279, (01) 07332414124359 (21) 102279;  102280, (01) 07332414124359 (21) 102280;  102281, (01) 07332414124359 (21) 102281;  102282, (01) 07332414124359 (21) 102282;  102283, (01) 07332414124359 (21) 102283;  102284, (01) 07332414124359 (21) 102284;  102285, (01) 07332414124359 (21) 102285;  102286, (01) 07332414124359 (21) 102286;  102287, (01) 07332414124359 (21) 102287;  102288, (01) 07332414124359 (21) 102288;  102289, (01) 07332414124359 (21) 102289;  102337, (01) 07332414124359 (21) 102337;  102338, (01) 07332414124359 (21) 102338;  102339, (01) 07332414124359 (21) 102339;  102340, (01) 07332414124359 (21) 102340;  102341, (01) 07332414124359 (21) 102341;  102342, (01) 07332414124359 (21) 102342;  102343, (01) 07332414124359 (21) 102343;  102344, (01) 07332414124359 (21) 102344;  102345, (01) 07332414124359 (21) 102345;  102346, (01) 07332414124359 (21) 102346;  102347, (01) 07332414124359 (21) 102347;  102348, (01) 07332414124359 (21) 102348;  102349, (01) 07332414124359 (21) 102349;  102350, (01) 07332414124359 (21) 102350;  102351, (01) 07332414124359 (21) 102351;  102532, (01) 07332414124359 (21) 102532;  102533, (01) 07332414124359 (21) 102533;  102534, (01) 07332414124359 (21) 102534;  102535, (01) 07332414124359 (21) 102535;  102536, (01) 07332414124359 (21) 102536;  102537, (01) 07332414124359 (21) 102537;  102538, (01) 07332414124359 (21) 102538;  102539, (01) 07332414124359 (21) 102539;  102540, (01) 07332414124359 (21) 102540;  102541, (01) 07332414124359 (21) 102541;  102542, (01) 07332414124359 (21) 102542;  102543, (01) 07332414124359 (21) 102543;  102544, (01) 07332414124359 (21) 102544;  102545, (01) 07332414124359 (21) 102545;  102546, (01) 07332414124359 (21) 102546;  102562, (01) 07332414124359 (21) 102562;  102563, (01) 07332414124359 (21) 102563;  102564, (01) 07332414124359 (21) 102564;  102565, (01) 07332414124359 (21) 102565;  102566, (01) 07332414124359 (21) 102566;  102567, (01) 07332414124359 (21) 102567;  102568, (01) 07332414124359 (21) 102568;  102569, (01) 07332414124359 (21) 102569;  102570, (01) 07332414124359 (21) 102570;  102571, (01) 07332414124359 (21) 102571;  102572, (01) 07332414124359 (21) 102572;  102573, (01) 07332414124359 (21) 102573;  102574, (01) 07332414124359 (21) 102574;  102575, (01) 07332414124359 (21) 102575;  102576, (01) 07332414124359 (21) 102576;  102592, (01) 07332414124359 (21) 102592;  102593, (01) 07332414124359 (21) 102593;  102594, (01) 07332414124359 (21) 102594;  102595, (01) 07332414124359 (21) 102595;  102596, (01) 07332414124359 (21) 102596;  102597, (01) 07332414124359 (21) 102597;  102598, (01) 07332414124359 (21) 102598;  102599, (01) 07332414124359 (21) 102599;  102600, (01) 07332414124359 (21) 102600;  102601, (01) 07332414124359 (21) 102601;  102602, (01) 07332414124359 (21) 102602;  102603, (01) 07332414124359 (21) 102603;  102604, (01) 07332414124359 (21) 102604;  102605, (01) 07332414124359 (21) 102605;  102606, (01) 07332414124359 (21) 102606;  102607, (01) 07332414124359 (21) 102607;  102608, (01) 07332414124359 (21) 102608;  102609, (01) 07332414124359 (21) 102609;  102610, (01) 07332414124359 (21) 102610;  102611, (01) 07332414124359 (21) 102611;  102612, (01) 07332414124359 (21) 102612;  102613, (01) 07332414124359 (21) 102613;  102614, (01) 07332414124359 (21) 102614;  102615, (01) 07332414124359 (21) 102615;  102616, (01) 07332414124359 (21) 102616;  102617, (01) 07332414124359 (21) 102617;  102618, (01) 07332414124359 (21) 102618;  102619, (01) 07332414124359 (21) 102619;  102620, (01) 07332414124359 (21) 102620;  102621, (01) 07332414124359 (21) 102621;  102637, (01) 07332414124359 (21) 102637;  102638, (01) 07332414124359 (21) 102638;  102639, (01) 07332414124359 (21) 102639;  102640, (01) 07332414124359 (21) 102640;  102641, (01) 07332414124359 (21) 102641;  102642, (01) 07332414124359 (21) 102642;  102643, (01) 07332414124359 (21) 102643;  102644, (01) 07332414124359 (21) 102644;  102645, (01) 07332414124359 (21) 102645;  102646, (01) 07332414124359 (21) 102646;  102647, (01) 07332414124359 (21) 102647;  102648, (01) 07332414124359 (21) 102648;  102649, (01) 07332414124359 (21) 102649;  102650, (01) 07332414124359 (21) 102650;  102651, (01) 07332414124359 (21) 102651;  102652, (01) 07332414124359 (21) 102652;  102653, (01) 07332414124359 (21) 102653;  102654, (01) 07332414124359 (21) 102654;  102655, (01) 07332414124359 (21) 102655;  102656, (01) 07332414124359 (21) 102656;  102657, (01) 07332414124359 (21) 102657;  102658, (01) 07332414124359 (21) 102658;  102659, (01) 07332414124359 (21) 102659;  102660, (01) 07332414124359 (21) 102660;  102661, (01) 07332414124359 (21) 102661;  102662, (01) 07332414124359 (21) 102662;  102663, (01) 07332414124359 (21) 102663;  102664, (01) 07332414124359 (21) 102664;  102665, (01) 07332414124359 (21) 102665;  102666, (01) 07332414124359 (21) 102666;  102706, (01) 07332414124359 (21) 102706;  102707, (01) 07332414124359 (21) 102707;  102708, (01) 07332414124359 (21) 102708;  102709, (01) 07332414124359 (21) 102709;  102710, (01) 07332414124359 (21) 102710;  102711, (01) 07332414124359 (21) 102711;  102712, (01) 07332414124359 (21) 102712;  102713, (01) 07332414124359 (21) 102713;  102714, (01) 07332414124359 (21) 102714;  102715, (01) 07332414124359 (21) 102715;  102716, (01) 07332414124359 (21) 102716;  102717, (01) 07332414124359 (21) 102717;  102718, (01) 07332414124359 (21) 102718;  102719, (01) 07332414124359 (21) 102719;  102720, (01) 07332414124359 (21) 102720;  102721, (01) 07332414124359 (21) 102721;  102722, (01) 07332414124359 (21) 102722;  102723, (01) 07332414124359 (21) 102723;  102724, (01) 07332414124359 (21) 102724;  102725, (01) 07332414124359 (21) 102725;  102726, (01) 07332414124359 (21) 102726;  102727, (01) 07332414124359 (21) 102727;  102728, (01) 07332414124359 (21) 102728;  102729, (01) 07332414124359 (21) 102729;  102730, (01) 07332414124359 (21) 102730;  102731, (01) 07332414124359 (21) 102731;  102732, (01) 07332414124359 (21) 102732;  102733, (01) 07332414124359 (21) 102733;  102734, (01) 07332414124359 (21) 102734;  102735, (01) 07332414124359 (21) 102735;  102736, (01) 07332414124359 (21) 102736;  102737, (01) 07332414124359 (21) 102737;  102738, (01) 07332414124359 (21) 102738;  102739, (01) 07332414124359 (21) 102739;  102740, (01) 07332414124359 (21) 102740;  102761, (01) 07332414124359 (21) 102761;  102762, (01) 07332414124359 (21) 102762;  102763, (01) 07332414124359 (21) 102763;  102764, (01) 07332414124359 (21) 102764;  102765, (01) 07332414124359 (21) 102765;  102766, (01) 07332414124359 (21) 102766;  102767, (01) 07332414124359 (21) 102767;  102768, (01) 07332414124359 (21) 102768;  102769, (01) 07332414124359 (21) 102769;  102770, (01) 07332414124359 (21) 102770;  102771, (01) 07332414124359 (21) 102771;  102772, (01) 07332414124359 (21) 102772;  102773, (01) 07332414124359 (21) 102773;  102774, (01) 07332414124359 (21) 102774;  102775, (01) 07332414124359 (21) 102775;  102776, (01) 07332414124359 (21) 102776;  102777, (01) 07332414124359 (21) 102777;  102778, (01) 07332414124359 (21) 102778;  102779, (01) 07332414124359 (21) 102779;  102780, (01) 07332414124359 (21) 102780;  102821, (01) 07332414124359 (21) 102821;  102822, (01) 07332414124359 (21) 102822;  102823, (01) 07332414124359 (21) 102823;  102824, (01) 07332414124359 (21) 102824;  102825, (01) 07332414124359 (21) 102825;  102826, (01) 07332414124359 (21) 102826;  102827, (01) 07332414124359 (21) 102827;  102828, (01) 07332414124359 (21) 102828;  102829, (01) 07332414124359 (21) 102829;  102830, (01) 07332414124359 (21) 102830;  102846, (01) 07332414124359 (21) 102846;  102847, (01) 07332414124359 (21) 102847;  102848, (01) 07332414124359 (21) 102848;  102849, (01) 07332414124359 (21) 102849;  102850, (01) 07332414124359 (21) 102850;  102851, (01) 07332414124359 (21) 102851;  102852, (01) 07332414124359 (21) 102852;  102853, (01) 07332414124359 (21) 102853;  102854, (01) 07332414124359 (21) 102854;  102855, (01) 07332414124359 (21) 102855;  102856, (01) 07332414124359 (21) 102856;  102857, (01) 07332414124359 (21) 102857;  102858, (01) 07332414124359 (21) 102858;  102859, (01) 07332414124359 (21) 102859;  102860, (01) 07332414124359 (21) 102860;  102861, (01) 07332414124359 (21) 102861;  102862, (01) 07332414124359 (21) 102862;  102863, (01) 07332414124359 (21) 102863;  102864, (01) 07332414124359 (21) 102864;  102865, (01) 07332414124359 (21) 102865;  102866, (01) 07332414124359 (21) 102866;  102867, (01) 07332414124359 (21) 102867;  102868, (01) 07332414124359 (21) 102868;  102869, (01) 07332414124359 (21) 102869;  102870, (01) 07332414124359 (21) 102870;  102871, (01) 07332414124359 (21) 102871;  102872, (01) 07332414124359 (21) 102872;  102873, (01) 07332414124359 (21) 102873;  102874, (01) 07332414124359 (21) 102874;  102875, (01) 07332414124359 (21) 102875;  102971, (01) 07332414124359 (21) 102971;  102972, (01) 07332414124359 (21) 102972;  102973, (01) 07332414124359 (21) 102973;  102974, (01) 07332414124359 (21) 102974;  102975, (01) 07332414124359 (21) 102975;  102976, (01) 07332414124359 (21) 102976;  102977, (01) 07332414124359 (21) 102977;  102978, (01) 07332414124359 (21) 102978;  102979, (01) 07332414124359 (21) 102979;  102980, (01) 07332414124359 (21) 102980;  102981, (01) 07332414124359 (21) 102981;  102982, (01) 07332414124359 (21) 102982;  102983, (01) 07332414124359 (21) 102983;  102984, (01) 07332414124359 (21) 102984;  102985, (01) 07332414124359 (21) 102985;  102986, (01) 07332414124359 (21) 102986;  102987, (01) 07332414124359 (21) 102987;  102988, (01) 07332414124359 (21) 102988;  102989, (01) 07332414124359 (21) 102989;  102990, (01) 07332414124359 (21) 102990;  102991, (01) 07332414124359 (21) 102991;  102992, (01) 07332414124359 (21) 102992;  102993, (01) 07332414124359 (21) 102993;  102994, (01) 07332414124359 (21) 102994;  102995, (01) 07332414124359 (21) 102995;  102996, (01) 07332414124359 (21) 102996;  102997, (01) 07332414124359 (21) 102997;  102998, (01) 07332414124359 (21) 102998;  102999, (01) 07332414124359 (21) 102999;  103000, (01) 07332414124359 (21) 103000;  103001, (01) 07332414124359 (21) 103001;  103002, (01) 07332414124359 (21) 103002;  103003, (01) 07332414124359 (21) 103003;  103004, (01) 07332414124359 (21) 103004;  103005, (01) 07332414124359 (21) 103005;  103006, (01) 07332414124359 (21) 103006;  103007, (01) 07332414124359 (21) 103007;  103008, (01) 07332414124359 (21) 103008;  103009, (01) 07332414124359 (21) 103009;  103010, (01) 07332414124359 (21) 103010;  103011, (01) 07332414124359 (21) 103011;  103012, (01) 07332414124359 (21) 103012;  103013, (01) 07332414124359 (21) 103013;  103014, (01) 07332414124359 (21) 103014;  103015, (01) 07332414124359 (21) 103015;  103036, (01) 07332414124359 (21) 103036;  103037, (01) 07332414124359 (21) 103037;  103038, (01) 07332414124359 (21) 103038;  103039, (01) 07332414124359 (21) 103039;  103040, (01) 07332414124359 (21) 103040;  103041, (01) 07332414124359 (21) 103041;  103042, (01) 07332414124359 (21) 103042;  103043, (01) 07332414124359 (21) 103043;  103044, (01) 07332414124359 (21) 103044;  103045, (01) 07332414124359 (21) 103045;  103046, (01) 07332414124359 (21) 103046;  103047, (01) 07332414124359 (21) 103047;  103048, (01) 07332414124359 (21) 103048;  103049, (01) 07332414124359 (21) 103049;  103050, (01) 07332414124359 (21) 103050;  103051, (01) 07332414124359 (21) 103051;  103052, (01) 07332414124359 (21) 103052;  103053, (01) 07332414124359 (21) 103053;  103054, (01) 07332414124359 (21) 103054;  103055, (01) 07332414124359 (21) 103055;  103056, (01) 07332414124359 (21) 103056;  103057, (01) 07332414124359 (21) 103057;  103058, (01) 07332414124359 (21) 103058;  103059, (01) 07332414124359 (21) 103059;  103060, (01) 07332414124359 (21) 103060;  103061, (01) 07332414124359 (21) 103061;  103062, (01) 07332414124359 (21) 103062;  103063, (01) 07332414124359 (21) 103063;  103064, (01) 07332414124359 (21) 103064;  103065, (01) 07332414124359 (21) 103065;  103066, (01) 07332414124359 (21) 103066;  103067, (01) 07332414124359 (21) 103067;  103068, (01) 07332414124359 (21) 103068;  103069, (01) 07332414124359 (21) 103069;  103070, (01) 07332414124359 (21) 103070;  103111, (01) 07332414124359 (21) 103111;  103112, (01) 07332414124359 (21) 103112;  103113, (01) 07332414124359 (21) 103113;  103114, (01) 07332414124359 (21) 103114;  103115, (01) 07332414124359 (21) 103115;  103116, (01) 07332414124359 (21) 103116;  103117, (01) 07332414124359 (21) 103117;  103118, (01) 07332414124359 (21) 103118;  103119, (01) 07332414124359 (21) 103119;  103120, (01) 07332414124359 (21) 103120;  103121, (01) 07332414124359 (21) 103121;  103122, (01) 07332414124359 (21) 103122;  103123, (01) 07332414124359 (21) 103123;  103124, (01) 07332414124359 (21) 103124;  103125, (01) 07332414124359 (21) 103125;  103146, (01) 07332414124359 (21) 103146;  103147, (01) 07332414124359 (21) 103147;  103148, (01) 07332414124359 (21) 103148;  103149, (01) 07332414124359 (21) 103149;  103150, (01) 07332414124359 (21) 103150;  103151, (01) 07332414124359 (21) 103151;  103152, (01) 07332414124359 (21) 103152;  103153, (01) 07332414124359 (21) 103153;  103154, (01) 07332414124359 (21) 103154;  103155, (01) 07332414124359 (21) 103155;  103156, (01) 07332414124359 (21) 103156;  103157, (01) 07332414124359 (21) 103157;  103158, (01) 07332414124359 (21) 103158;  103159, (01) 07332414124359 (21) 103159;  103160, (01) 07332414124359 (21) 103160;  103161, (01) 07332414124359 (21) 103161;  103162, (01) 07332414124359 (21) 103162;  103163, (01) 07332414124359 (21) 103163;  103164, (01) 07332414124359 (21) 103164;  103165, (01) 07332414124359 (21) 103165;  103166, (01) 07332414124359 (21) 103166;  103167, (01) 07332414124359 (21) 103167;  103168, (01) 07332414124359 (21) 103168;  103169, (01) 07332414124359 (21) 103169;  103170, (01) 07332414124359 (21) 103170;  103171, (01) 07332414124359 (21) 103171;  103172, (01) 07332414124359 (21) 103172;  103173, (01) 07332414124359 (21) 103173;  103174, (01) 07332414124359 (21) 103174;  103175, (01) 07332414124359 (21) 103175;  103236, (01) 07332414124359 (21) 103236;  103237, (01) 07332414124359 (21) 103237;  103238, (01) 07332414124359 (21) 103238;  103239, (01) 07332414124359 (21) 103239;  103240, (01) 07332414124359 (21) 103240;  103241, (01) 07332414124359 (21) 103241;  103242, (01) 07332414124359 (21) 103242;  103243, (01) 07332414124359 (21) 103243;  103244, (01) 07332414124359 (21) 103244;  103245, (01) 07332414124359 (21) 103245;  103246, (01) 07332414124359 (21) 103246;  103247, (01) 07332414124359 (21) 103247;  103248, (01) 07332414124359 (21) 103248;  103249, (01) 07332414124359 (21) 103249;  103250, (01) 07332414124359 (21) 103250;  103251, (01) 07332414124359 (21) 103251;  103252, (01) 07332414124359 (21) 103252;  103253, (01) 07332414124359 (21) 103253; 103254, (01) 07332414124359 (21) 103254;  103255, (01) 07332414124359 (21) 103255;  103256, (01) 07332414124359 (21) 103256;  103257, (01) 07332414124359 (21) 103257;  103258, (01) 07332414124359 (21) 103258;  103259, (01) 07332414124359 (21) 103259;  103260, (01) 07332414124359 (21) 103260;  103261, (01) 07332414124359 (21) 103261;  103262, (01) 07332414124359 (21) 103262;  103263, (01) 07332414124359 (21) 103263;  103264, (01) 07332414124359 (21) 103264;  103265, (01) 07332414124359 (21) 103265;  103266, (01) 07332414124359 (21) 103266;  103267, (01) 07332414124359 (21) 103267;  103268, (01) 07332414124359 (21) 103268;  103269, (01) 07332414124359 (21) 103269;  103270, (01) 07332414124359 (21) 103270;  103271, (01) 07332414124359 (21) 103271;  103272, (01) 07332414124359 (21) 103272;  103273, (01) 07332414124359 (21) 103273;  103274, (01) 07332414124359 (21) 103274;  103275, (01) 07332414124359 (21) 103275;  103276, (01) 07332414124359 (21) 103276;  103277, (01) 07332414124359 (21) 103277;  103278, (01) 07332414124359 (21) 103278;  103279, (01) 07332414124359 (21) 103279;  103280, (01) 07332414124359 (21) 103280;  103281, (01) 07332414124359 (21) 103281;  103282, (01) 07332414124359 (21) 103282;  103283, (01) 07332414124359 (21) 103283;  103284, (01) 07332414124359 (21) 103284;  103285, (01) 07332414124359 (21) 103285;  103296, (01) 07332414124359 (21) 103296;  103297, (01) 07332414124359 (21) 103297;  103298, (01) 07332414124359 (21) 103298;  103299, (01) 07332414124359 (21) 103299;  103300, (01) 07332414124359 (21) 103300;  103301, (01) 07332414124359 (21) 103301;  103302, (01) 07332414124359 (21) 103302;  103303, (01) 07332414124359 (21) 103303;  103304, (01) 07332414124359 (21) 103304;  103305, (01) 07332414124359 (21) 103305;  103306, (01) 07332414124359 (21) 103306;  103307, (01) 07332414124359 (21) 103307;  103308, (01) 07332414124359 (21) 103308;  103309, (01) 07332414124359 (21) 103309;  103310, (01) 07332414124359 (21) 103310;  103311, (01) 07332414124359 (21) 103311;  103312, (01) 07332414124359 (21) 103312;  103313, (01) 07332414124359 (21) 103313;  103314, (01) 07332414124359 (21) 103314;  103315, (01) 07332414124359 (21) 103315;  103316, (01) 07332414124359 (21) 103316;  103317, (01) 07332414124359 (21) 103317;  103318, (01) 07332414124359 (21) 103318;  103319, (01) 07332414124359 (21) 103319;  103320, (01) 07332414124359 (21) 103320;  103321, (01) 07332414124359 (21) 103321;  103322, (01) 07332414124359 (21) 103322;  103323, (01) 07332414124359 (21) 103323;  103324, (01) 07332414124359 (21) 103324;  103325, (01) 07332414124359 (21) 103325;  103376, (01) 07332414124359 (21) 103376;  103377, (01) 07332414124359 (21) 103377;  103378, (01) 07332414124359 (21) 103378;  103379, (01) 07332414124359 (21) 103379;  103380, (01) 07332414124359 (21) 103380;  103381, (01) 07332414124359 (21) 103381;  103382, (01) 07332414124359 (21) 103382;  103383, (01) 07332414124359 (21) 103383;  103384, (01) 07332414124359 (21) 103384;  103385, (01) 07332414124359 (21) 103385;  103386, (01) 07332414124359 (21) 103386;  103387, (01) 07332414124359 (21) 103387;  103388, (01) 07332414124359 (21) 103388;  103389, (01) 07332414124359 (21) 103389;  103390, (01) 07332414124359 (21) 103390;  103391, (01) 07332414124359 (21) 103391;  103392, (01) 07332414124359 (21) 103392;  103393, (01) 07332414124359 (21) 103393;  103394, (01) 07332414124359 (21) 103394;  103395, (01) 07332414124359 (21) 103395;  103436, (01) 07332414124359 (21) 103436;  103437, (01) 07332414124359 (21) 103437;  103438, (01) 07332414124359 (21) 103438;  103439, (01) 07332414124359 (21) 103439;  103440, (01) 07332414124359 (21) 103440;  103441, (01) 07332414124359 (21) 103441;  103442, (01) 07332414124359 (21) 103442;  103443, (01) 07332414124359 (21) 103443;  103444, (01) 07332414124359 (21) 103444;  103445, (01) 07332414124359 (21) 103445;  103486, (01) 07332414124359 (21) 103486;  103487, (01) 07332414124359 (21) 103487;  103488, (01) 07332414124359 (21) 103488;  103489, (01) 07332414124359 (21) 103489;  103490, (01) 07332414124359 (21) 103490;  103491, (01) 07332414124359 (21) 103491;  103492, (01) 07332414124359 (21) 103492;  103493, (01) 07332414124359 (21) 103493;  103494, (01) 07332414124359 (21) 103494;  103495, (01) 07332414124359 (21) 103495;  103496, (01) 07332414124359 (21) 103496;  103497, (01) 07332414124359 (21) 103497;  103498, (01) 07332414124359 (21) 103498;  103499, (01) 07332414124359 (21) 103499;  103500, (01) 07332414124359 (21) 103500;  103501, (01) 07332414124359 (21) 103501;  103502, (01) 07332414124359 (21) 103502;  103503, (01) 07332414124359 (21) 103503;  103504, (01) 07332414124359 (21) 103504;  103505, (01) 07332414124359 (21) 103505;  103506, (01) 07332414124359 (21) 103506;  103507, (01) 07332414124359 (21) 103507;  103508, (01) 07332414124359 (21) 103508;  103509, (01) 07332414124359 (21) 103509;  103510, (01) 07332414124359 (21) 103510;  103511, (01) 07332414124359 (21) 103511;  103512, (01) 07332414124359 (21) 103512;  103513, (01) 07332414124359 (21) 103513;  103514, (01) 07332414124359 (21) 103514;  103515, (01) 07332414124359 (21) 103515;  103516, (01) 07332414124359 (21) 103516;  103517, (01) 07332414124359 (21) 103517;  103518, (01) 07332414124359 (21) 103518;  103519, (01) 07332414124359 (21) 103519;  103520, (01) 07332414124359 (21) 103520;  103521, (01) 07332414124359 (21) 103521;  103522, (01) 07332414124359 (21) 103522;  103523, (01) 07332414124359 (21) 103523;  103524, (01) 07332414124359 (21) 103524;  103525, (01) 07332414124359 (21) 103525;  103526, (01) 07332414124359 (21) 103526;  103527, (01) 07332414124359 (21) 103527;  103528, (01) 07332414124359 (21) 103528;  103529, (01) 07332414124359 (21) 103529;  103530, (01) 07332414124359 (21) 103530;  103531, (01) 07332414124359 (21) 103531;  103532, (01) 07332414124359 (21) 103532;  103533, (01) 07332414124359 (21) 103533;  103534, (01) 07332414124359 (21) 103534;  103535, (01) 07332414124359 (21) 103535;  103586, (01) 07332414124359 (21) 103586;  103587, (01) 07332414124359 (21) 103587;  103588, (01) 07332414124359 (21) 103588;  103589, (01) 07332414124359 (21) 103589;  103590, (01) 07332414124359 (21) 103590;  103591, (01) 07332414124359 (21) 103591;  103592, (01) 07332414124359 (21) 103592;  103593, (01) 07332414124359 (21) 103593;  103594, (01) 07332414124359 (21) 103594;  103595, (01) 07332414124359 (21) 103595;  103596, (01) 07332414124359 (21) 103596;  103597, (01) 07332414124359 (21) 103597;  103598, (01) 07332414124359 (21) 103598;  103599, (01) 07332414124359 (21) 103599;  103600, (01) 07332414124359 (21) 103600;  103601, (01) 07332414124359 (21) 103601;  103602, (01) 07332414124359 (21) 103602;  103603, (01) 07332414124359 (21) 103603;  103604, (01) 07332414124359 (21) 103604;  103605, (01) 07332414124359 (21) 103605;  103616, (01) 07332414124359 (21) 103616;  103617, (01) 07332414124359 (21) 103617;  103618, (01) 07332414124359 (21) 103618;  103619, (01) 07332414124359 (21) 103619;  103620, (01) 07332414124359 (21) 103620;  103621, (01) 07332414124359 (21) 103621;  103622, (01) 07332414124359 (21) 103622;  103623, (01) 07332414124359 (21) 103623;  103624, (01) 07332414124359 (21) 103624;  103625, (01) 07332414124359 (21) 103625;  103636, (01) 07332414124359 (21) 103636;  103637, (01) 07332414124359 (21) 103637;  103638, (01) 07332414124359 (21) 103638;  103639, (01) 07332414124359 (21) 103639;  103640, (01) 07332414124359 (21) 103640;  103641, (01) 07332414124359 (21) 103641;  103642, (01) 07332414124359 (21) 103642;  103643, (01) 07332414124359 (21) 103643;  103644, (01) 07332414124359 (21) 103644;  103645, (01) 07332414124359 (21) 103645;  103656, (01) 07332414124359 (21) 103656;  103657, (01) 07332414124359 (21) 103657;  103658, (01) 07332414124359 (21) 103658;  103659, (01) 07332414124359 (21) 103659;  103660, (01) 07332414124359 (21) 103660;  103661, (01) 07332414124359 (21) 103661;  103662, (01) 07332414124359 (21) 103662;  103663, (01) 07332414124359 (21) 103663;  103664, (01) 07332414124359 (21) 103664;  103665, (01) 07332414124359 (21) 103665;  103676, (01) 07332414124359 (21) 103676;  103677, (01) 07332414124359 (21) 103677;  103678, (01) 07332414124359 (21) 103678;  103679, (01) 07332414124359 (21) 103679;  103680, (01) 07332414124359 (21) 103680;  103681, (01) 07332414124359 (21) 103681;  103682, (01) 07332414124359 (21) 103682;  103683, (01) 07332414124359 (21) 103683;  103684, (01) 07332414124359 (21) 103684;  103685, (01) 07332414124359 ( 21) 103685;  103756, (01) 07332414124359 (21) 103756;  103757, (01) 07332414124359 (21) 103757;  103758, (01) 07332414124359 (21) 103758;  103759, (01) 07332414124359 (21) 103759;  103760, (01) 07332414124359 (21) 103760;  103761, (01) 07332414124359 (21) 103761;  103762, (01) 07332414124359 (21) 103762;  103763, (01) 07332414124359 (21) 103763;  103764, (01) 07332414124359 (21) 103764;  103765, (01) 07332414124359 (21) 103765;  103776, (01) 07332414124359 (21) 103776;  103777, (01) 07332414124359 (21) 103777;  103778, (01) 07332414124359 (21) 103778;  103779, (01) 07332414124359 (21) 103779;  103780, (01) 07332414124359 (21) 103780;  103781, (01) 07332414124359 (21) 103781;  103782, (01) 07332414124359 (21) 103782;  103783, (01) 07332414124359 (21) 103783;  103784, (01) 07332414124359 (21) 103784;  103785, (01) 07332414124359 (21) 103785;  103796, (01) 07332414124359 (21) 103796;  103797, (01) 07332414124359 (21) 103797;  103798, (01) 07332414124359 (21) 103798;  103799, (01) 07332414124359 (21) 103799;  103800, (01) 07332414124359 (21) 103800;  103801, (01) 07332414124359 (21) 103801;  103802, (01) 07332414124359 (21) 103802;  103803, (01) 07332414124359 (21) 103803;  103804, (01) 07332414124359 (21) 103804;  103805, (01) 07332414124359 (21) 103805;  103816, (01) 07332414124359 (21) 103816;  103817, (01) 07332414124359 (21) 103817;  103818, (01) 07332414124359 (21) 103818;  103819, (01) 07332414124359 (21) 103819;  103820, (01) 07332414124359 (21) 103820;  103821, (01) 07332414124359 (21) 103821;  103822, (01) 07332414124359 (21) 103822;  103823, (01) 07332414124359 (21) 103823;  103824, (01) 07332414124359 (21) 103824;  103825, (01) 07332414124359 (21) 103825;  103836, (01) 07332414124359 (21) 103836;  103837, (01) 07332414124359 (21) 103837;  103838, (01) 07332414124359 (21) 103838;  103839, (01) 07332414124359 (21) 103839;  103840, (01) 07332414124359 (21) 103840;  103841, (01) 07332414124359 (21) 103841;  103842, (01) 07332414124359 (21) 103842;  103843, (01) 07332414124359 (21) 103843; 103844, (01) 07332414124359 (21) 103844;  103845, (01) 07332414124359 (21) 103845;  103996, (01) 07332414124359 (21) 103996;  103997, (01) 07332414124359 (21) 103997;  103998, (01) 07332414124359 (21) 103998;  103999, (01) 07332414124359 (21) 103999;  104000, (01) 07332414124359 (21) 104000;  104001, (01) 07332414124359 (21) 104001;  104002, (01) 07332414124359 (21) 104002;  104003, (01) 07332414124359 (21) 104003;  104004, (01) 07332414124359 (21) 104004;  104005, (01) 07332414124359 (21) 104005;  104016, (01) 07332414124359 (21) 104016;  104017, (01) 07332414124359 (21) 104017;  104018, (01) 07332414124359 (21) 104018;  104019, (01) 07332414124359 (21) 104019;  104020, (01) 07332414124359 (21) 104020;  104021, (01) 07332414124359 (21) 104021;  104022, (01) 07332414124359 (21) 104022;  104023, (01) 07332414124359 (21) 104023;  104024, (01) 07332414124359 (21) 104024;  104025, (01) 07332414124359 (21) 104025;  104036, (01) 07332414124359 (21) 104036;  104037, (01) 07332414124359 (21) 104037;  104038, (01) 07332414124359 (21) 104038;  104039, (01) 07332414124359 (21) 104039;  104040, (01) 07332414124359 (21) 104040;  104041, (01) 07332414124359 (21) 104041;  104042, (01) 07332414124359 (21) 104042;  104043, (01) 07332414124359 (21) 104043;  104044, (01) 07332414124359 (21) 104044;  104045, (01) 07332414124359 (21) 104045;  104056, (01) 07332414124359 (21) 104056;  104057, (01) 07332414124359 (21) 104057;  104058, (01) 07332414124359 (21) 104058;  104059, (01) 07332414124359 (21) 104059;  104060, (01) 07332414124359 (21) 104060;  104061, (01) 07332414124359 (21) 104061;  104062, (01) 07332414124359 (21) 104062;  104063, (01) 07332414124359 (21) 104063;  104064, (01) 07332414124359 (21) 104064;  104065, (01) 07332414124359 (21) 104065;  104266, (01) 07332414124359 (21) 104266;  104267, (01) 07332414124359 (21) 104267;  104268, (01) 07332414124359 (21) 104268;  104269, (01) 07332414124359 (21) 104269;  104270, (01) 07332414124359 (21) 104270;  104271, (01) 07332414124359 (21) 104271;  104272, (01) 07332414124359 (21) 104272;  104273, (01) 07332414124359 (21) 104273;  104274, (01) 07332414124359 (21) 104274;  104275, (01) 07332414124359 (21) 104275;  104286, (01) 07332414124359 (21) 104286;  104287, (01) 07332414124359 (21) 104287;  104288, (01) 07332414124359 (21) 104288;  104289, (01) 07332414124359 (21) 104289;  104290, (01) 07332414124359 (21) 104290;  104291, (01) 07332414124359 (21) 104291;  104292, (01) 07332414124359 (21) 104292;  104293, (01) 07332414124359 (21) 104293;  104294, (01) 07332414124359 (21) 104294;  104295, (01) 07332414124359 (21) 104295;  104306, (01) 07332414124359 (21) 104306;  104307, (01) 07332414124359 (21) 104307;  104308, (01) 07332414124359 (21) 104308;  104309, (01) 07332414124359 (21) 104309;  104310, (01) 07332414124359 (21) 104310;  104311, (01) 07332414124359 (21) 104311;  104312, (01) 07332414124359 (21) 104312;  104313, (01) 07332414124359 (21) 104313;  104314, (01) 07332414124359 (21) 104314;  104315, (01) 07332414124359 (21) 104315;  104396, (01) 07332414124359 (21) 104396;  104397, (01) 07332414124359 (21) 104397;  104398, (01) 07332414124359 (21) 104398;  104399, (01) 07332414124359 (21) 104399;  104400, (01) 07332414124359 (21) 104400;  104401, (01) 07332414124359 (21) 104401;  104402, (01) 07332414124359 (21) 104402;  104403, (01) 07332414124359 (21) 104403;  104404, (01) 07332414124359 (21) 104404;  104405, (01) 07332414124359 (21) 104405;  104436, (01) 07332414124359 (21) 104436;  104437, (01) 07332414124359 (21) 104437;  104438, (01) 07332414124359 (21) 104438;  104439, (01) 07332414124359 (21) 104439;  104440, (01) 07332414124359 (21) 104440;  104441, (01) 07332414124359 (21) 104441;  104442, (01) 07332414124359 (21) 104442;  104443, (01) 07332414124359 (21) 104443;  104444, (01) 07332414124359 (21) 104444;  104445, (01) 07332414124359 (21) 104445;  104496, (01) 07332414124359 (21) 104496;  104497, (01) 07332414124359 (21) 104497;  104498, (01) 07332414124359 (21) 104498;  104499, (01) 07332414124359 (21) 104499;  104500, (01) 07332414124359 (21) 104500;  104501, (01) 07332414124359 (21) 104501;  104502, (01) 07332414124359 (21) 104502;  104503, (01) 07332414124359 (21) 104503;  104504, (01) 07332414124359 (21) 104504;  104505, (01) 07332414124359 (21) 104505;  104546, (01) 07332414124359 (21) 104546;  104547, (01) 07332414124359 (21) 104547;  104548, (01) 07332414124359 (21) 104548;  104549, (01) 07332414124359 (21) 104549;  104550, (01) 07332414124359 (21) 104550;  104551, (01) 07332414124359 (21) 104551;  104552, (01) 07332414124359 (21) 104552;  104553, (01) 07332414124359 (21) 104553;  104554, (01) 07332414124359 (21) 104554;  104555, (01) 07332414124359 (21) 104555;  104576, (01) 07332414124359 (21) 104576;  104577, (01) 07332414124359 (21) 104577;  104578, (01) 07332414124359 (21) 104578;  104579, (01) 07332414124359 (21) 104579;  104580, (01) 07332414124359 (21) 104580;  104581, (01) 07332414124359 (21) 104581;  104582, (01) 07332414124359 (21) 104582;  104583, (01) 07332414124359 (21) 104583;  104584, (01) 07332414124359 (21) 104584;  104585, (01) 07332414124359 (21) 104585;  104596, (01) 07332414124359 (21) 104596;  104597, (01) 07332414124359 (21) 104597;  104598, (01) 07332414124359 (21) 104598;  104599, (01) 07332414124359 (21) 104599;  104600, (01) 07332414124359 (21) 104600;  104601, (01) 07332414124359 (21) 104601;  104602, (01) 07332414124359 (21) 104602;  104603, (01) 07332414124359 (21) 104603;  104604, (01) 07332414124359 (21) 104604;  104605, (01) 07332414124359 (21) 104605;  104626, (01) 07332414124359 (21) 104626;  104627, (01) 07332414124359 (21) 104627;  104628, (01) 07332414124359 (21) 104628;  104629, (01) 07332414124359 (21) 104629;  104630, (01) 07332414124359 (21) 104630;  104631, (01) 07332414124359 (21) 104631;  104632, (01) 07332414124359 (21) 104632;  104633, (01) 07332414124359 (21) 104633;  104634, (01) 07332414124359 (21) 104634;  104635, (01) 07332414124359 (21) 104635;  104766, (01) 07332414124359 (21) 104766;  104767, (01) 07332414124359 (21) 104767;  104768, (01) 07332414124359 (21) 104768;  104769, (01) 07332414124359 (21) 104769;  104770, (01) 07332414124359 (21) 104770;  104771, (01) 07332414124359 (21) 104771;  104772, (01) 07332414124359 (21) 104772;  104773, (01) 07332414124359 (21) 104773;  104774, (01) 07332414124359 (21) 104774;  104775, (01) 07332414124359 (21) 104775;  104786, (01) 07332414124359 (21) 104786;  104787, (01) 07332414124359 (21) 104787;  104788, (01) 07332414124359 (21) 104788;  104789, (01) 07332414124359 (21) 104789;  104790, (01) 07332414124359 (21) 104790;  104791, (01) 07332414124359 (21) 104791;  104792, (01) 07332414124359 (21) 104792;  104793, (01) 07332414124359 (21) 104793;  104794, (01) 07332414124359 (21) 104794;  104795, (01) 07332414124359 (21) 104795;  104796, (01) 07332414124359 (21) 104796;  104797, (01) 07332414124359 (21) 104797;  104798, (01) 07332414124359 (21) 104798;  104799, (01) 07332414124359 (21) 104799;  104800, (01) 07332414124359 (21) 104800;  104801, (01) 07332414124359 (21) 104801;  104802, (01) 07332414124359 (21) 104802;  104803, (01) 07332414124359 (21) 104803;  104804, (01) 07332414124359 (21) 104804;  104805, (01) 07332414124359 (21) 104805;  104806, (01) 07332414124359 (21) 104806;  104807, (01) 07332414124359 (21) 104807;  104808, (01) 07332414124359 (21) 104808;  104809, (01) 07332414124359 (21) 104809;  104810, (01) 07332414124359 (21) 104810;  104811, (01) 07332414124359 (21) 104811;  104812, (01) 07332414124359 (21) 104812;  104813, (01) 07332414124359 (21) 104813;  104814, (01) 07332414124359 (21) 104814;  104815, (01) 07332414124359 (21) 104815;  104826, (01) 07332414124359 (21) 104826;  104827, (01) 07332414124359 (21) 104827;  104828, (01) 07332414124359 (21) 104828;  104829, (01) 07332414124359 (21) 104829;  104830, (01) 07332414124359 (21) 104830;  104831, (01) 07332414124359 (21) 104831;  104832, (01) 07332414124359 (21) 104832;  104833, (01) 07332414124359 (21) 104833;  104834, (01) 07332414124359 (21) 104834;  104835, (01) 07332414124359 (21) 104835;  104906, (01) 07332414124359 (21) 104906;  104907, (01) 07332414124359 (21) 104907;  104908, (01) 07332414124359 (21) 104908;  104909, (01) 07332414124359 (21) 104909;  104910, (01) 07332414124359 (21) 104910;  104911, (01) 07332414124359 (21) 104911;  104912, (01) 07332414124359 (21) 104912;  104913, (01) 07332414124359 (21) 104913;  104914, (01) 07332414124359 (21) 104914;  104915, (01) 07332414124359 (21) 104915;  104936, (01) 07332414124359 (21) 104936;  104937, (01) 07332414124359 (21) 104937;  104938, (01) 07332414124359 (21) 104938;  104939, (01) 07332414124359 (21) 104939;  104940, (01) 07332414124359 (21) 104940;  104941, (01) 07332414124359 (21) 104941;  104942, (01) 07332414124359 (21) 104942;  104943, (01) 07332414124359 (21) 104943;  104944, (01) 07332414124359 (21) 104944;  104945, (01) 07332414124359 (21) 104945;  104966, (01) 07332414124359 (21) 104966;  104967, (01) 07332414124359 (21) 104967;  104968, (01) 07332414124359 (21) 104968;  104969, (01) 07332414124359 (21) 104969;  104970, (01) 07332414124359 (21) 104970;  104971, (01) 07332414124359 (21) 104971;  104972, (01) 07332414124359 (21) 104972;  104973, (01) 07332414124359 (21) 104973;  104974, (01) 07332414124359 (21) 104974;  104975, (01) 07332414124359 (21) 104975;  104986, (01) 07332414124359 (21) 104986;  104987, (01) 07332414124359 (21) 104987;  104988, (01) 07332414124359 (21) 104988;  104989, (01) 07332414124359 (21) 104989;  104990, (01) 07332414124359 (21) 104990;  104991, (01) 07332414124359 (21) 104991;  104992, (01) 07332414124359 (21) 104992;  104993, (01) 07332414124359 (21) 104993;  104994, (01) 07332414124359 (21) 104994;  104995, (01) 07332414124359 (21) 104995;  105056, (01) 07332414124359 (21) 105056;  105057, (01) 07332414124359 (21) 105057;  105058, (01) 07332414124359 (21) 105058;  105059, (01) 07332414124359 (21) 105059;  105060, (01) 07332414124359 (21) 105060;  105061, (01) 07332414124359 (21) 105061;  105062, (01) 07332414124359 (21) 105062;  105063, (01) 07332414124359 (21) 105063;  105064, (01) 07332414124359 (21) 105064;  105065, (01) 07332414124359 (21) 105065;  105076, (01) 07332414124359 (21) 105076;  105077, (01) 07332414124359 (21) 105077;  105079, (01) 07332414124359 (21) 105079;  105080, (01) 07332414124359 (21) 105080;  105081, (01) 07332414124359 (21) 105081;  105082, (01) 07332414124359 (21) 105082;  105083, (01) 07332414124359 (21) 105083;  105084, (01) 07332414124359 (21) 105084;  105085, (01) 07332414124359 (21) 105085;  105086, (01) 07332414124359 (21) 105086;  105087, (01) 07332414124359 (21) 105087;  105088, (01) 07332414124359 (21) 105088;  105089, (01) 07332414124359 (21) 105089;  105090, (01) 07332414124359 (21) 105090;  105091, (01) 07332414124359 (21) 105091;  105092, (01) 07332414124359 (21) 105092;  105093, (01) 07332414124359 (21) 105093;  105094, (01) 07332414124359 (21) 105094;  105095, (01) 07332414124359 (21) 105095;  105106, (01) 07332414124359 (21) 105106;  105107, (01) 07332414124359 (21) 105107;  105108, (01) 07332414124359 (21) 105108;  105109, (01) 07332414124359 (21) 105109;  105110, (01) 07332414124359 (21) 105110;  105111, (01) 07332414124359 (21) 105111;  105112, (01) 07332414124359 (21) 105112;  105113, (01) 07332414124359 (21) 105113;  105114, (01) 07332414124359 (21) 105114;  105115, (01) 07332414124359 (21) 105115;  105126, (01) 07332414124359 (21) 105126;  105127, (01) 07332414124359 (21) 105127;  105128, (01) 07332414124359 (21) 105128;  105129, (01) 07332414124359 (21) 105129;  105130, (01) 07332414124359 (21) 105130;  105131, (01) 07332414124359 (21) 105131;  105132, (01) 07332414124359 (21) 105132;  105133, (01) 07332414124359 (21) 105133;  105134, (01) 07332414124359 (21) 105134;  105135, (01) 07332414124359 (21) 105135;  105146, (01) 07332414124359 (21) 105146;  105147, (01) 07332414124359 (21) 105147;  105148, (01) 07332414124359 (21) 105148;  105149, (01) 07332414124359 (21) 105149;  105150, (01) 07332414124359 (21) 105150;  105151, (01) 07332414124359 (21) 105151;  105152, (01) 07332414124359 (21) 105152;  105153, (01) 07332414124359 (21) 105153;  105154, (01) 07332414124359 (21) 105154;  105155, (01) 07332414124359 (21) 105155;  105156, (01) 07332414124359 (21) 105156;  105157, (01) 07332414124359 (21) 105157;  105158, (01) 07332414124359 (21) 105158;  105159, (01) 07332414124359 (21) 105159;  105160, (01) 07332414124359 (21) 105160;  105161, (01) 07332414124359 (21) 105161;  105162, (01) 07332414124359 (21) 105162;  105163, (01) 07332414124359 (21) 105163;  105164, (01) 07332414124359 (21) 105164;  105165, (01) 07332414124359 (21) 105165;  105176, (01) 07332414124359 (21) 105176;  105177, (01) 07332414124359 (21) 105177;  105178, (01) 07332414124359 (21) 105178;  105179, (01) 07332414124359 (21) 105179;  105180, (01) 07332414124359 (21) 105180;  105181, (01) 07332414124359 (21) 105181;  105182, (01) 07332414124359 (21) 105182;  105183, (01) 07332414124359 (21) 105183;  105184, (01) 07332414124359 (21) 105184;  105185, (01) 07332414124359 (21) 105185;  105206, (01) 07332414124359 (21) 105206;  105207, (01) 07332414124359 (21) 105207;  105208, (01) 07332414124359 (21) 105208;  105209, (01) 07332414124359 (21) 105209;  105210, (01) 07332414124359 (21) 105210;  105211, (01) 07332414124359 (21) 105211;  105212, (01) 07332414124359 (21) 105212;  105213, (01) 07332414124359 (21) 105213;  105214, (01) 07332414124359 (21) 105214;  105215, (01) 07332414124359 (21) 105215;  105226, (01) 07332414124359 (21) 105226;  105227, (01) 07332414124359 (21) 105227;  105228, (01) 07332414124359 (21) 105228;  105229, (01) 07332414124359 (21) 105229;  105230, (01) 07332414124359 (21) 105230;  105231, (01) 07332414124359 (21) 105231;  105232, (01) 07332414124359 (21) 105232;  105233, (01) 07332414124359 (21) 105233;  105234, (01) 07332414124359 (21) 105234;  105235, (01) 07332414124359 (21) 105235;  105236, (01) 07332414124359 (21) 105236;  105237, (01) 07332414124359 (21) 105237;  105238, (01) 07332414124359 (21) 105238;  105239, (01) 07332414124359 (21) 105239;  105240, (01) 07332414124359 (21) 105240;  105241, (01) 07332414124359 (21) 105241;  105242, (01) 07332414124359 (21) 105242;  105243, (01) 07332414124359 (21) 105243;  105244, (01) 07332414124359 (21) 105244;  105245, (01) 07332414124359 (21) 105245;  105286, (01) 07332414124359 (21) 105286;  105287, (01) 07332414124359 (21) 105287;  105288, (01) 07332414124359 (21) 105288;  105289, (01) 07332414124359 (21) 105289;  105290, (01) 07332414124359 (21) 105290;  105291, (01) 07332414124359 (21) 105291;  105292, (01) 07332414124359 (21) 105292;  105293, (01) 07332414124359 (21) 105293;  105294, (01) 07332414124359 (21) 105294;  105295, (01) 07332414124359 (21) 105295;  105306, (01) 07332414124359 (21) 105306;  105307, (01) 07332414124359 (21) 105307;  105308, (01) 07332414124359 (21) 105308;  105309, (01) 07332414124359 (21) 105309;  105310, (01) 07332414124359 (21) 105310;  105311, (01) 07332414124359 (21) 105311;  105312, (01) 07332414124359 (21) 105312;  105313, (01) 07332414124359 (21) 105313;  105314, (01) 07332414124359 (21) 105314;  105315, (01) 07332414124359 (21) 105315;  105316, (01) 07332414124359 (21) 105316;  105317, (01) 07332414124359 (21) 105317;  105318, (01) 07332414124359 (21) 105318;  105319, (01) 07332414124359 (21) 105319;  105320, (01) 07332414124359 (21) 105320;  105321, (01) 07332414124359 (21) 105321;  105322, (01) 07332414124359 (21) 105322;  105323, (01) 07332414124359 (21) 105323;  105324, (01) 07332414124359 (21) 105324;  105325, (01) 07332414124359 (21) 105325;  105336, (01) 07332414124359 (21) 105336;  105337, (01) 07332414124359 (21) 105337;  105338, (01) 07332414124359 (21) 105338;  105339, (01) 07332414124359 (21) 105339;  105340, (01) 07332414124359 (21) 105340;  105341, (01) 07332414124359 (21) 105341;  105342, (01) 07332414124359 (21) 105342;  105343, (01) 07332414124359 (21) 105343;  105344, (01) 07332414124359 (21) 105344;  105345, (01) 07332414124359 (21) 105345;  105356, (01) 07332414124359 (21) 105356;  105357, (01) 07332414124359 (21) 105357;  105358, (01) 07332414124359 (21) 105358;  105359, (01) 07332414124359 (21) 105359;  105360, (01) 07332414124359 (21) 105360;  105361, (01) 07332414124359 (21) 105361;  105362, (01) 07332414124359 (21) 105362;  105363, (01) 07332414124359 (21) 105363;  105364, (01) 07332414124359 (21) 105364;  105365, (01) 07332414124359 (21) 105365;  105366, (01) 07332414124359 (21) 105366;  105367, (01) 07332414124359 (21) 105367;  105368, (01) 07332414124359 (21) 105368;  105369, (01) 07332414124359 (21) 105369;  105370, (01) 07332414124359 (21) 105370;  105371, (01) 07332414124359 (21) 105371;  105372, (01) 07332414124359 (21) 105372;  105373, (01) 07332414124359 (21) 105373;  105374, (01) 07332414124359 (21) 105374;  105375, (01) 07332414124359 (21) 105375;  105376, (01) 07332414124359 (21) 105376;  105378, (01) 07332414124359 (21) 105378;  105379, (01) 07332414124359 (21) 105379;  105380, (01) 07332414124359 (21) 105380;  105381, (01) 07332414124359 (21) 105381;  105382, (01) 07332414124359 (21) 105382;  105383, (01) 07332414124359 (21) 105383;  105384, (01) 07332414124359 (21) 105384;  105385, (01) 07332414124359 (21) 105385;  105386, (01) 07332414124359 (21) 105386;  105387, (01) 07332414124359 (21) 105387;  105388, (01) 07332414124359 (21) 105388;  105389, (01) 07332414124359 (21) 105389;  105390, (01) 07332414124359 (21) 105390;  105391, (01) 07332414124359 (21) 105391;  105392, (01) 07332414124359 (21) 105392;  105393, (01) 07332414124359 (21) 105393;  105394, (01) 07332414124359 (21) 105394;  105395, (01) 07332414124359 (21) 105395;  105396, (01) 07332414124359 (21) 105396;  105397, (01) 07332414124359 (21) 105397;  105398, (01) 07332414124359 (21) 105398;  105399, (01) 07332414124359 (21) 105399;  105400, (01) 07332414124359 (21) 105400;  105401, (01) 07332414124359 (21) 105401;  105402, (01) 07332414124359 (21) 105402;  105403, (01) 07332414124359 (21) 105403;  105404, (01) 07332414124359 (21) 105404;  105405, (01) 07332414124359 (21) 105405;  105406, (01) 07332414124359 (21) 105406;  105407, (01) 07332414124359 (21) 105407;  105408, (01) 07332414124359 (21) 105408;  105409, (01) 07332414124359 (21) 105409;  105410, (01) 07332414124359 (21) 105410;  105411, (01) 07332414124359 (21) 105411;  105412, (01) 07332414124359 (21) 105412;  105413, (01) 07332414124359 (21) 105413;  105414, (01) 07332414124359 (21) 105414;  105415, (01) 07332414124359 (21) 105415;  105426, (01) 07332414124359 (21) 105426;  105427, (01) 07332414124359 (21) 105427;  105428, (01) 07332414124359 (21) 105428;  105429, (01) 07332414124359 (21) 105429;  105430, (01) 07332414124359 (21) 105430;  105431, (01) 07332414124359 (21) 105431;  105432, (01) 07332414124359 (21) 105432;  105433, (01) 07332414124359 (21) 105433;  105434, (01) 07332414124359 (21) 105434;  105435, (01) 07332414124359 (21) 105435;  105446, (01) 07332414124359 (21) 105446;  105447, (01) 07332414124359 (21) 105447;  105448, (01) 07332414124359 (21) 105448;  105449, (01) 07332414124359 (21) 105449;  105450, (01) 07332414124359 (21) 105450;  105451, (01) 07332414124359 (21) 105451;  105452, (01) 07332414124359 (21) 105452;  105453, (01) 07332414124359 (21) 105453;  105454, (01) 07332414124359 (21) 105454;  105455, (01) 07332414124359 (21) 105455;  105546, (01) 07332414124359 (21) 105546;  105547, (01) 07332414124359 (21) 105547;  105548, (01) 07332414124359 (21) 105548;  105549, (01) 07332414124359 (21) 105549;  105550, (01) 07332414124359 (21) 105550;  105551, (01) 07332414124359 (21) 105551;  105552, (01) 07332414124359 (21) 105552;  105553, (01) 07332414124359 (21) 105553;  105554, (01) 07332414124359 (21) 105554;  105555, (01) 07332414124359 (21) 105555;  105566, (01) 07332414124359 (21) 105566;  105567, (01) 07332414124359 (21) 105567;  105568, (01) 07332414124359 (21) 105568;  105569, (01) 07332414124359 (21) 105569;  105570, (01) 07332414124359 (21) 105570;  105571, (01) 07332414124359 (21) 105571;  105572, (01) 07332414124359 (21) 105572;  105573, (01) 07332414124359 (21) 105573;  105574, (01) 07332414124359 (21) 105574;  105575, (01) 07332414124359 (21) 105575;  105586, (01) 07332414124359 (21) 105586;  105587, (01) 07332414124359 (21) 105587;  105588, (01) 07332414124359 (21) 105588;  105589, (01) 07332414124359 (21) 105589;  105590, (01) 07332414124359 (21) 105590;  105591, (01) 07332414124359 (21) 105591;  105592, (01) 07332414124359 (21) 105592;  105593, (01) 07332414124359 (21) 105593;  105594, (01) 07332414124359 (21) 105594;  105595, (01) 07332414124359 (21) 105595;  105606, (01) 07332414124359 (21) 105606;  105607, (01) 07332414124359 (21) 105607;  105608, (01) 07332414124359 (21) 105608;  105609, (01) 07332414124359 (21) 105609;  105610, (01) 07332414124359 (21) 105610;  105611, (01) 07332414124359 (21) 105611;  105612, (01) 07332414124359 (21) 105612;  105613, (01) 07332414124359 (21) 105613;  105614, (01) 07332414124359 (21) 105614;  105615, (01) 07332414124359 (21) 105615;  105712, (01) 07332414124359 (21) 105712;  105713, (01) 07332414124359 (21) 105713;  105714, (01) 07332414124359 (21) 105714;  105715, (01) 07332414124359 (21) 105715;  105716, (01) 07332414124359 (21) 105716;  105717, (01) 07332414124359 (21) 105717;  105718, (01) 07332414124359 (21) 105718;  105719, (01) 07332414124359 (21) 105719;  105720, (01) 07332414124359 (21) 105720;  105721, (01) 07332414124359 (21) 105721;  105722, (01) 07332414124359 (21) 105722;  105723, (01) 07332414124359 (21) 105723;  105724, (01) 07332414124359 (21) 105724;  105725, (01) 07332414124359 (21) 105725;  105726, (01) 07332414124359 (21) 105726;  105727, (01) 07332414124359 (21) 105727;  105728, (01) 07332414124359 (21) 105728;  105729, (01) 07332414124359 (21) 105729;  105730, (01) 07332414124359 (21) 105730;  105731, (01) 07332414124359 (21) 105731;  105742, (01) 07332414124359 (21) 105742;  105743, (01) 07332414124359 (21) 105743;  105744, (01) 07332414124359 (21) 105744;  105745, (01) 07332414124359 (21) 105745;  105746, (01) 07332414124359 (21) 105746;  105747, (01) 07332414124359 (21) 105747;  105748, (01) 07332414124359 (21) 105748;  105749, (01) 07332414124359 (21) 105749;  105750, (01) 07332414124359 (21) 105750;  105751, (01) 07332414124359 (21) 105751;  105752, (01) 07332414124359 (21) 105752;  105753, (01) 07332414124359 (21) 105753;  105754, (01) 07332414124359 (21) 105754;  105755, (01) 07332414124359 (21) 105755;  105756, (01) 07332414124359 (21) 105756;  105757, (01) 07332414124359 (21) 105757;  105758, (01) 07332414124359 (21) 105758;  105759, (01) 07332414124359 (21) 105759;  105760, (01) 07332414124359 (21) 105760;  105761, (01) 07332414124359 (21) 105761;  105792, (01) 07332414124359 (21) 105792;  105793, (01) 07332414124359 (21) 105793;  105794, (01) 07332414124359 (21) 105794;  105795, (01) 07332414124359 (21) 105795;  105796, (01) 07332414124359 (21) 105796;  105797, (01) 07332414124359 (21) 105797;  105798, (01) 07332414124359 (21) 105798;  105799, (01) 07332414124359 (21) 105799;  105800, (01) 07332414124359 (21) 105800;  105801, (01) 07332414124359 (21) 105801;  105832, (01) 07332414124359 (21) 105832;  105833, (01) 07332414124359 (21) 105833;  105834, (01) 07332414124359 (21) 105834;  105835, (01) 07332414124359 (21) 105835;  105836, (01) 07332414124359 (21) 105836;  105837, (01) 07332414124359 (21) 105837;  105838, (01) 07332414124359 (21) 105838;  105839, (01) 07332414124359 (21) 105839;  105840, (01) 07332414124359 (21) 105840;  105841, (01) 07332414124359 (21) 105841;  105842, (01) 07332414124359 (21) 105842;  105843, (01) 07332414124359 (21) 105843;  105844, (01) 07332414124359 (21) 105844;  105845, (01) 07332414124359 (21) 105845;  105846, (01) 07332414124359 (21) 105846;  105847, (01) 07332414124359 (21) 105847;  105848, (01) 07332414124359 (21) 105848;  105849, (01) 07332414124359 (21) 105849;  105850, (01) 07332414124359 (21) 105850;  105851, (01) 07332414124359 (21) 105851;  105882, (01) 07332414124359 (21) 105882;  105883, (01) 07332414124359 (21) 105883;  105884, (01) 07332414124359 (21) 105884;  105885, (01) 07332414124359 (21) 105885;  105886, (01) 07332414124359 (21) 105886;  105887, (01) 07332414124359 (21) 105887;  105889, (01) 07332414124359 (21) 105889;  105890, (01) 07332414124359 (21) 105890;  105891, (01) 07332414124359 (21) 105891;  105892, (01) 07332414124359 (21) 105892;  105893, (01) 07332414124359 (21) 105893;  105894, (01) 07332414124359 (21) 105894;  105895, (01) 07332414124359 (21) 105895;  105896, (01) 07332414124359 (21) 105896;  105897, (01) 07332414124359 (21) 105897;  105898, (01) 07332414124359 (21) 105898;  105899, (01) 07332414124359 (21) 105899;  105900, (01) 07332414124359 (21) 105900;  105901, (01) 07332414124359 (21) 105901;  105902, (01) 07332414124359 (21) 105902;  105903, (01) 07332414124359 (21) 105903;  105904, (01) 07332414124359 (21) 105904;  105905, (01) 07332414124359 (21) 105905;  105906, (01) 07332414124359 (21) 105906;  105907, (01) 07332414124359 (21) 105907;  105908, (01) 07332414124359 (21) 105908;  105909, (01) 07332414124359 (21) 105909;  105910, (01) 07332414124359 (21) 105910;  105911, (01) 07332414124359 (21) 105911;  105932, (01) 07332414124359 (21) 105932;  105933, (01) 07332414124359 (21) 105933;  105934, (01) 07332414124359 (21) 105934;  105935, (01) 07332414124359 (21) 105935;  105936, (01) 07332414124359 (21) 105936;  105938, (01) 07332414124359 (21) 105938;  105939, (01) 07332414124359 (21) 105939;  105940, (01) 07332414124359 (21) 105940;  105941, (01) 07332414124359 (21) 105941;  105942, (01) 07332414124359 (21) 105942;  105943, (01) 07332414124359 (21) 105943;  105944, (01) 07332414124359 (21) 105944;  105945, (01) 07332414124359 (21) 105945;  105946, (01) 07332414124359 (21) 105946;  105948, (01) 07332414124359 (21) 105948;  105949, (01) 07332414124359 (21) 105949;  105950, (01) 07332414124359 (21) 105950;  105951, (01) 07332414124359 (21) 105951;  105992, (01) 07332414124359 (21) 105992;  105993, (01) 07332414124359 (21) 105993;  105994, (01) 07332414124359 (21) 105994;  105995, (01) 07332414124359 (21) 105995;  105996, (01) 07332414124359 (21) 105996;  105997, (01) 07332414124359 (21) 105997;  105998, (01) 07332414124359 (21) 105998;  105999, (01) 07332414124359 (21) 105999;  106000, (01) 07332414124359 (21) 106000;  106001, (01) 07332414124359 (21) 106001;  106002, (01) 07332414124359 (21) 106002;  106003, (01) 07332414124359 (21) 106003;  106004, (01) 07332414124359 (21) 106004;  106005, (01) 07332414124359 (21) 106005;  106006, (01) 07332414124359 (21) 106006;  106007, (01) 07332414124359 (21) 106007;  106008, (01) 07332414124359 (21) 106008;  106009, (01) 07332414124359 (21) 106009;  106010, (01) 07332414124359 (21) 106010;  106011, (01) 07332414124359 (21) 106011;  106012, (01) 07332414124359 (21) 106012;  106013, (01) 07332414124359 (21) 106013;  106014, (01) 07332414124359 (21) 106014;  106015, (01) 07332414124359 (21) 106015;  106016, (01) 07332414124359 (21) 106016;  106017, (01) 07332414124359 (21) 106017;  106018, (01) 07332414124359 (21) 106018;  106020, (01) 07332414124359 (21) 106020;  106021, (01) 07332414124359 (21) 106021;  106052, (01) 07332414124359 (21) 106052;  106053, (01) 07332414124359 (21) 106053;  106054, (01) 07332414124359 (21) 106054;  106055, (01) 07332414124359 (21) 106055;  106056, (01) 07332414124359 (21) 106056;  106057, (01) 07332414124359 (21) 106057;  106058, (01) 07332414124359 (21) 106058;  106059, (01) 07332414124359 (21) 106059;  106060, (01) 07332414124359 (21) 106060;  106061, (01) 07332414124359 (21) 106061;  106122, (01) 07332414124359 (21) 106122;  106123, (01) 07332414124359 (21) 106123;  106125, (01) 07332414124359 (21) 106125;  106126, (01) 07332414124359 (21) 106126;  106127, (01) 07332414124359 (21) 106127;  106129, (01) 07332414124359 (21) 106129;  106130, (01) 07332414124359 (21) 106130;  106131, (01) 07332414124359 (21) 106131;  106132, (01) 07332414124359 (21) 106132;  106133, (01) 07332414124359 (21) 106133;  106134, (01) 07332414124359 (21) 106134;  106135, (01) 07332414124359 (21) 106135;  106137, (01) 07332414124359 (21) 106137;  106138, (01) 07332414124359 (21) 106138;  106140, (01) 07332414124359 (21) 106140;  106141, (01) 07332414124359 (21) 106141;  106143, (01) 07332414124359 (21) 106143;  106144, (01) 07332414124359 (21) 106144;  106146, (01) 07332414124359 (21) 106146;  106147, (01) 07332414124359 (21) 106147;  106148, (01) 07332414124359 (21) 106148;  106149, (01) 07332414124359 (21) 106149;  106150, (01) 07332414124359 (21) 106150;  106151, (01) 07332414124359 (21) 106151;  106202, (01) 07332414124359 (21) 106202;  106203, (01) 07332414124359 (21) 106203;  106204, (01) 07332414124359 (21) 106204;  106206, (01) 07332414124359 (21) 106206;  106208, (01) 07332414124359 (21) 106208;  106210, (01) 07332414124359 (21) 106210;  106220, (01) 07332414124359 (21) 106220 (1683 units)    c) 955701: 104446, (01) 07332414125844 (21) 104446;  104447, (01) 07332414125844 (21) 104447;  104448, (01) 07332414125844 (21) 104448;  104449, (01) 07332414125844 (21) 104449;  104450, (01) 07332414125844 (21) 104450;  104451, (01) 07332414125844 (21) 104451;  104452, (01) 07332414125844 (21) 104452;  104453, (01) 07332414125844 (21) 104453;  104454, (01) 07332414125844 (21) 104454;  104455, (01) 07332414125844 (21) 104455;  104466, (01) 07332414125844 (21) 104466;  104467, (01) 07332414125844 (21) 104467;  104468, (01) 07332414125844 (21) 104468;  104469, (01) 07332414125844 (21) 104469;  104470, (01) 07332414125844 (21) 104470;  104471, (01) 07332414125844 (21) 104471;  104472, (01) 07332414125844 (21) 104472;  104473, (01) 07332414125844 (21) 104473;  104474, (01) 07332414125844 (21) 104474;  104475, (01) 07332414125844 (21) 104475;  105646, (01) 07332414125844 (21) 105646;  105647, (01) 07332414125844 (21) 105647;  105648, (01) 07332414125844 (21) 105648;  105649, (01) 07332414125844 (21) 105649;  105650, (01) 07332414125844 (21) 105650;  105651, (01) 07332414125844 (21) 105651;  105652, (01) 07332414125844 (21) 105652;  105653, (01) 07332414125844 (21) 105653 (28 units)  Product Code SC6169:  Lot Numbers: 1286902, 1286903, 1286871, 1286870, 1286868, 1286869, 1286623, 1278425, 1281024, 1278909, 1278423, 1278424, 1277964, 1277966, 1277965, 1277743, 1276586, 1277559, 1275517  Product Code SC6035, Lot Numbers:  1287266, 1275298, 1275299  Product Code SC6053, Lot Numbers:  1277956, 1278671"
    }
  ]
}