      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79540",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Distribution in the following U.S. states: AZ, CA, FL, IA, IL, LA, MI, NC, NJ, OH, OK, and TX.",
      "recall_number": "Z-1150-2018",
      "product_description": "AXIOM Artis angiographic x-ray system, Model Numbers: 7412807, 7413078, 7728392     AXIOM Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.",
      "product_quantity": "22",
      "reason_for_recall": "Certain systems may have leaking of the plastic coolant couplings, the liquid can enter into the flat panel detector leading to a failure of the detector. It may be necessary to cancel or restart the treatment or transfer the patient to an alternate or another system.",
      "recall_initiation_date": "20180105",
      "center_classification_date": "20180321",
      "termination_date": "20180823",
      "report_date": "20180328",
      "code_info": "Serial Numbers: 35061  35054  34999  35098  35059  35064  35073  35071  35057  35075  35083  35018  35097  35108  35046  35048  35050  46004  35045  35004  44011  35095"
    }
  ]
}