{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "San Jose",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84664",
      "recalling_firm": "Stryker Corporation",
      "address_1": "5900 Optical Ct",
      "address_2": "",
      "postal_code": "95138-1400",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US: GA, NJ, NY, KY, MD, TX, UT, AZ, IL, ID, CO, OH, FL, AK, CA, NV, AR, ME, WI",
      "recall_number": "Z-1147-2020",
      "product_description": "Stryker, Omega 4.75 mm,. PEEK Knotless Anchor System, Double-Double, REF 3910-500-472, UDI 07613327464832, Sterile R, RX Only CE 0197",
      "product_quantity": "73 devices",
      "reason_for_recall": "Complaints have been received regarding difficulty in removing the driver from the screw.  High driver removal forces can lead to the screw pulling out still attached to the driver. This could result  in an additional anchor insertion site in the bone being used.",
      "recall_initiation_date": "20190719",
      "center_classification_date": "20200211",
      "termination_date": "20240202",
      "report_date": "20200219",
      "code_info": "19113AG2, 19130AG2, 19134AG2, 19115AG2, 19131AG2, 19133AG2, 19154AG2,19155AG2, 19156AG2, 19157AG2, 19149AG2.",
      "more_code_info": ""
    }
  ]
}