      "country": "United States",
      "city": "Irvine",
      "address_1": "34 Bunsen",
      "reason_for_recall": "HORIBA Medical is informing all ABX PENTRA MultiCal, N Control, and P Control customers running the  Creatinine Rate Blank Method assay on the ABX PENTRA 400 and PC200 that the Target Values for Creatinine Rate  Blank Method have been updated. This issue is only applicable to Creatinine Rate Blank Method values reported on the  ABX PENTRA 400 and PC200 Chemistry Analyzers only. No other analyzers",
      "address_2": "",
      "product_quantity": "2,023 units",
      "code_info": "Model Numbers:Multical - A11A01652,  N Control - A11A01653,  P Control - A11A01654.    Lot Numbers: MultiCal: 1204101 and 1304901,  N Control: 1300801,  P Control: 1300601.    Expiration Date (s): MultiCal: 1204101 Expires on 05/31/14,  MultiCal: 1304901 Expires on 07/15/15,  N Control: 1300801 Expires on 11/05/14,  P Control: 1300601 Expires on 11/05/14.",
      "center_classification_date": "20140305",
      "distribution_pattern": "United States nationwide distribution.",
      "state": "CA",
      "product_description": "ABX PENTRA Multical.  ABX PENTRA N Control.  ABX PENTRA P Control.    The ABX PENTRA 400 and PC200 are benchtop clinical chemistry analyzers using two  measuring principals: absorbance and ion selective electrodes.  The ABX PENTRA Multical is a lyophilized human serum calibrator with chemical  additives and materials of biological origin. The assigned values of the calibrator  components are given in the enclosed annex, ensuring optimal calibration of the  appropriate HORIBA ABX SAS methods on the ABX PENTRA Clinical Chemistry  Analyzer. This calibrator is provided in ten vials of 3 ml.",
      "report_date": "20140312",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Horiba Instruments, Inc dba Horiba Medical",
      "recall_number": "Z-1146-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "67332",
      "termination_date": "20141205",
      "more_code_info": "",
      "recall_initiation_date": "20140120",
      "postal_code": "92618-4210",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}