      "status": "Terminated",
      "city": "Glens Falls",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70388",
      "recalling_firm": "Angiodynamics",
      "address_1": "10 Technical Park",
      "address_2": "N/A",
      "postal_code": "12804",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution -  US Distribution to the states of AZ, AR, FL, MI, MN, MO, NY, TX, VA, WA and WI., and to the countries of Hungary and Spain.",
      "recall_number": "Z-1143-2015",
      "product_description": "Starburst MRI Semi-Flex Electrode Device, 25cm/Attached Cable/14Ga/6.4F, Product No. H7877001039091, Catalog No. 700-103909; 25cm/14Ga/6.4F, Product No. H7877001026151, Catalog No. 700-102615; and 25cm/14Ga/6.4F, Product No. H7877001026151, Catalog No. 700-102615",
      "product_quantity": "43",
      "reason_for_recall": "Printed mark on the trocar intended to indicate a 5cm distance from the distal tip of the needle is actually located 4cm from the distal tip of the needle.",
      "recall_initiation_date": "20141219",
      "center_classification_date": "20150224",
      "termination_date": "20151127",
      "report_date": "20150304",
      "code_info": "Lot No. 4773676, 4776428 and 4780110",
      "more_code_info": ""
    }
  ]
}