{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Murrysville",
      "address_1": "1001 Murry Ridge Ln",
      "reason_for_recall": "Internal testing revealed a potentially defective component on the power management board that could affect the function of certain Trilogy Ventilators resulting in loss of therapy to patients.",
      "address_2": "",
      "product_quantity": "572",
      "code_info": "Model (s), Catelogue, or Code Numbers were identified as: 1040000 Trilogy 202 International, Number 1040005 Trilogy200 USA,  1054096 Trilogy 100 International,   1054260 Trilogy 100 USA, 1040004 Trilogy 02, and 1058180 Garbin.  Multiple serial numbers also indicated: TV114011447; TV114011448; TV114011449  TV11401144A; TV11401144B; TV11401144C; TV11401144D; TV11401144E;  TV114011450; TV114011451; TV11401145B; TV11401145C; TV11401145D;  TV11401145F; TV114011460; TV114011461; TV114011462; TV114011464;   TV114012326; TV114012364;TV1140120A9; TV114012323; TV11401212C;  TV11401230F; TV11401235A; TV11401235C; TV114012370TV114012371  TV114012372; TV114012376; TV114012415; TV114012465; TV21401150C;  TV214011510; TV11401153D; TV11401213E; TV21401150F;TV114012165;  TV11401133A;TV114012318; TV114012433; TV11401223A; TV11401223E;  TV114012170; TV114011722; TV1140120B2;TV114012416; TV114012235;  TV114012148; TV114011550; TV1140120BB; TV114012469; TV114012459;  TV114012732;TV21401150A; TV11401130C; TV11401236D; TV114011455 ;  TV114011465; TV114012239;TV11401223B; TV11401224D;TV114011336  TV11401133B;TV21401150E;TV214011511;TV214011513;TV214011515;  TV214011516;TV214011519;TV21401151B;TV21401151F;TV21401210B  TV214012117;TV214012119;TV21401211A;TV21401211D;TV214012122;  TV114011712;TV114011458; TV114012133;TV114012361;  TV114012429  TV11401245B  TV114011724  TV114011727  TV114011740  TV214011512  TV21401210C  TV214012111  TV214012120  TV214012121  TV21401240D  TV114012333  TV11401233E  TV114012349  TV11401234C  TV114012422  TV11401243B  TV11401245F  TV11401231C  TV114012355  TV114012137  TV11401233C  TV1140120DB  TV114012453  TV114011635  TV114011752  TV11401134F  TV114011741  TV114011757  TV114012095  TV1140120CD  TV1140120E3  TV11401215B  TV114012130  TV114012134  TV114012343  TV114012347  TV114012462  TV114012473  TV114011554  TV114011711  TV114011719  TV114011723  TV114011725  TV114012259  TV014011405  TV014011503  TV014011504  TV21401210D  TV21401210F  TV214012114  TV214012116  TV11401207D  TV11401207F  TV114012085  TV11401208C  TV1140120A6  TV114011311  TV1140120C1  TV114012146  TV1140120E2  TV11401174E  TV114011754  TV1140120DA  TV11401233A  TV114012342  TV11401234B  TV11401241D  TV114012441  TV114012456  TV114012458  TV11401245A  TV11401245C  TV11401245D  TV014012302  TV014012404  TV014012204  TV014012206  TV014012207  TV014012208  TV014012402  TV014012405  TV014012406  TV014012407  TV014012408  TV014011601  TV014011607  TV21401151D  TV214011522  TV114011335  TV114012313  TV11401236C  TV114011744  TV114012090  TV11401209C  TV11401209F  TV114012098  TV11401230C  TV114012377  TV11401213B  TV11401214A  TV114012154  TV11401245E  TV11401230B  TV114012414  TV114012417  TV014011401  TV014011406  TV11401214D  TV11401246D  TV114012418  TV11401241C  TV11312282C  TV11401130F  TV11401146A  TV11401146C  TV11401146F  TV114011523  TV114011524  TV11401152A  TV11401152D  TV114011530  TV114011536  TV114011538  TV11401155C  TV11401170F  TV114011717  TV11401172C  TV1140120DC  TV114012160  TV114012163  TV11401216A  TV11401216B  TV114012240  TV114012446  TV11401246E  TV1140120D4  TV1140120D5  TV114012155  TV114012157  TV114012423  TV114012132  TV11401241E  TV114012432  TV11401212E  TV114012312  TV11401231D  TV114012351  TV11401235D  TV114012468  TV11401246F  TV114012470  TV114012472  TV11401241A  TV11401242A  TV1140120D0  TV11401224E  TV114012233   TV114012234  TV114012237  TV11401223F  TV11401224C  TV114012250  TV114012255  TV114012256  TV11401225B  TV114011746  TV114012322  TV11401154F  TV11401231A  TV114012330  TV114011317  TV11401152C  TV114011540  TV114011546  TV11401154A  TV11401154E  TV114011551  TV114011555  TV114011562  TV11401170D  TV114011718  TV11401172B  TV114011745  TV114011760  TV21401151A  TV21401151C  TV1140120D9  TV114012162  TV114012238  TV114012242  TV11401224B  TV114012358  TV11401161A  TV11401161F  TV114011623  TV11401162E  TV11401162F;",
      "center_classification_date": "20140306",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and the countries of  France, Great Britian, China, India, Italy, Korea, Kuwait, Nehterlands, and Singapore.",
      "state": "PA",
      "product_description": "Trilogy Ventilators  The Respironics Trilogy 100 system provides continuous or intermittent ventilator support for the care of individuals who require mechanical ventilation. Trilogy 100 is intended for pediatric through adult patients weighing at least 5 kg (11 lbs.). The device is intended to be used in home, institution/hospital, and portable applications such as wheelchairs and gurneys, and may be used for both invasive and non-invasive ventilation. It is not intended to be used as a transport ventilator. Trilogy 200 Intended Use: The Respironics Trilogy 200 system provides continuous or intermittent ventilator support for the care of individuals who require mechanical ventilation. Trilogy 200 is intended for pediatric through adult patients weighing at least 5 kg (11 lbs.). The device is intended to be used in home, institution/hospital, and portable applications such as wheelchairs and gurneys, and may be used for both invasive and non-invasive ventilation. Trilogy 202 Intended Use: The Respironics Trilogy 202 system provides continuous or intermittent ventilator support for the care of individuals who require mechanical ventilation with or without air/oxygen blending. Trilogy 202 is intended for pediatric through adult patients weighing at least 5 kg (11 lbs.). The device is intended to be used in hospitals and institutions, and for portable applications such as wheelchairs and gurneys only when in an institutional setting. It may be used for both invasive and non-invasive ventilation. It is not intended to be used as a transport ventilator.",
      "report_date": "20140312",
      "classification": "Class I",
      "openfda": {},
      "recalling_firm": "Phillips Respironics, Inc.",
      "recall_number": "Z-1143-2014",
      "initial_firm_notification": "Telephone",
      "product_type": "Devices",
      "event_id": "67550",
      "termination_date": "20150417",
      "more_code_info": "",
      "recall_initiation_date": "20140211",
      "postal_code": "15668-8517",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}