  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Mansfield",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87179",
      "recalling_firm": "Covidien Llc",
      "address_1": "15 Hampshire St",
      "address_2": "N/A",
      "postal_code": "02048-1113",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the state of MA and the countries of France,  French Guiana, Germany, Luxembourg, Spain, United Kingdom.",
      "recall_number": "Z-1142-2021",
      "product_description": "Covidien Parietex hydrophilic anatomical mesh 15 cm x 10 cm polyester left - Product Usage: designed for the repair of inguinal hernias via laparoscopic or posterior open procedures Item Code: TECT1510AL",
      "product_quantity": "129 units WW",
      "reason_for_recall": "incorrect device is contained in the package. The packaging label indicates that the device is for left side use, the product inside the package is for right side use. Use of an opposite side mesh can result in implant failure and hernia recurrence",
      "recall_initiation_date": "20210113",
      "center_classification_date": "20210222",
      "termination_date": "20230214",
      "report_date": "20210303",
      "code_info": "Lot Number: SUH0709M"
    }
  ]
}