{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Best",
      "state": "N/A",
      "country": "Netherlands",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82459",
      "recalling_firm": "Philips Medical Systems Nederlands",
      "address_1": "Veenpluis 4-6",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution in the states of  CA, FL, MI, OR, PA, TX and WA.",
      "recall_number": "Z-1140-2019",
      "product_description": "Foot Switches used with Philips MultiDiagnost Eleva with Flat Detector; Product Codes: 708037, 708038    Product The Allura Xper FD Series are intended for use on human patients to perform:   Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures.",
      "product_quantity": "7209 total",
      "reason_for_recall": "Additional units have been identified for previous recall Z-1280-2016 (RES 69339): If a Footswitch is frequently used on a non-flat surface or in the pedestal, the footswitch pedals may get bent. This might cause an intermittent or continuous inability of making live fluoroscopy images or exposures.",
      "recall_initiation_date": "20190219",
      "center_classification_date": "20190413",
      "termination_date": "20201014",
      "report_date": "20190424",
      "code_info": "Foot Switches:  452270000141 Footswitch CV 3p 4m; 452270000151 Footswitch MD 3p 6m;  452270000142 Footswitch CV 3p 4m; 452270000381 Footswitch CV 3p 8m; 452270000382 Footswitch CV 3p 8m; 459800076001 Footswitch CV 4p+2 4m; 459800076021 Footswitch CV 4p+2 8m; 459800415571 Wireless Footswitch 4p+2; 459800415581 Wireless footswitch set 4p+2;  459800415531 Wireless Footswitch 3P;  459800415561 Wireless footswitch set 3P."
    }
  ]
}