{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Franklin Lakes",
      "state": "NJ",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79223",
      "recalling_firm": "Becton Dickinson & Company",
      "address_1": "1 Becton Dr",
      "address_2": "N/A",
      "postal_code": "07417-1815",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Distributed in 46 states: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NY, NV, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and Puerto Rico.",
      "recall_number": "Z-1139-2018",
      "product_description": "BD Vacutainer¿ EDTA Lavender Top Tubes: Catalog Numbers: 1. 366401 (UDI: 50382903664011); 2. 366450 (UDI: 50382903664509); 3. 366643 (UDI: 50382903666435); 4. 367835 (UDI: 50382903678353); 5. 367841 (UDI: 50382903678414); 6. 367844 (UDI: 50382903678445); 7. 367856 (UDI: 50382903678568); 8. 367861 (UDI: 50382903678612); 9. 367862 (UDI: 50382903678629); 10. 367863 (UDI: 50382903678636); 11. 368021 (UDI: 50382903680219); 12. 368047 (UDI: 50382903680479); 13. 368054 (UDI: 50382903680547); 14. 368661 (UDI: 50382903686617); 15. 364300 (UDI: pre-UDI)",
      "product_quantity": "1,116,012,050",
      "reason_for_recall": "BD Vacutainer EDTA Lavender, Tan, and Pink Top Tubes and BD Vacutainer Lithium Heparin Green Top Tubes can cause an underestimation of lead in blood samples when used with Magellan Diagnostics LeadCare assays, employing the Anodic Stripping Voltammetry (ASV) methodology, or any other assay employing ASV methodology.",
      "recall_initiation_date": "20180322",
      "center_classification_date": "20180322",
      "termination_date": "20220725",
      "report_date": "20180328",
      "code_info": "All lots. Catalog Number (Shelf Life): 1. 366401 (24 months); 2. 366450 (24 months); 3. 366643 (12 months); 4. 367835 (16 months); 5. 367841 (15 months); 6. 367844 (16 months); 7. 367856 (16 months); 8. 367861 (16 months); 9. 367862 (16 months); 10. 367863 (17 months); 11. 368021 (16 months); 12. 368047 (15 months); 13. 368054 (16 months); 14. 368661 (17 months); 15. 364300 (Requested)"
    }
  ]
}