  "results": [
    {
      "status": "Terminated",
      "city": "Oakland",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82322",
      "recalling_firm": "Topcon Medical Systems, Inc.",
      "address_1": "111 Bauer Dr",
      "address_2": "N/A",
      "postal_code": "07436-3123",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide",
      "recall_number": "Z-1138-2019",
      "product_description": "IMAGEnet 6 v1.53-   IMAGEnet 6 Ophthalmic Data System is a software program that is intended for use in the collection, storage and management of digital images, patient data, diagnostic data and clinical information from Topcon devices without  controlling or altering the functions and parameters of any medical devices or through computerized networks.",
      "product_quantity": "450 units",
      "reason_for_recall": "When a patient is selected in the patient data and another patient is selected prior to the display of the primary patient's thumbnails is completed, then the primary patient's thumbnails are displayed with secondary patient's identification information such as name, patient's ID, etc. on the monitor screen of the PC used.",
      "recall_initiation_date": "20190306",
      "center_classification_date": "20190412",
      "termination_date": "20190717",
      "report_date": "20190424",
      "code_info": "System Version 1.53"
    }
  ]
}