{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Marietta",
      "state": "GA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90022",
      "recalling_firm": "ProgenaCare",
      "address_1": "2275 Northwest Pkwy Se Ste 175",
      "address_2": "",
      "postal_code": "30067-9321",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of AL, GA, FL, TN, and TX.  There was no foreign/military/government distribution.",
      "recall_number": "Z-1137-2022",
      "product_description": "progenamatrix Human Keratin Matrix, sterile, Rx, with the following model numbers: (1) PM2012, 12 x 2 cm; (2) PM3002, 2 x 2 cm; (3) PM3004, 4 x 4 cm; PM3006, 6 x 6 cm; and PM3010, 10 x 10 cm.",
      "product_quantity": "547 packages",
      "reason_for_recall": "Inner pouch seal may be inadequate resulting in lack of sterility.",
      "recall_initiation_date": "20220418",
      "center_classification_date": "20220526",
      "termination_date": "20230414",
      "report_date": "20220601",
      "code_info": "(1) Model PM2012, lot #02SEP21-001, exp. 4/9/2024; lot #29SEP21-001, exp. 4/9/2024.  (2) Model PM3002, lot #28OCT21-001, exp. 3/7/2024.  (3) Model PM3004, lot #02DEC21-001, exp. 10/1/2023; lot #02SEP21-002, exp. 10/1/2023; lot #14OCT21-001, exp. 10/1/2023; and 23NOV21-001, exp. 10/1/2023.  (4) Model PM3006, lot #02SEP21-003, exp. 10/1/2023; lot #03DEC21-001, exp. 10/1/2023; lot #14OCT21-002, exp. 10/1/2023; and lot #23NOV21-002, exp. 10/1/2023.  (5) Model PM3010, lot #13JUL21-001, exp. 10/1/2023; and lot #15JUL21-002, exp. 3/6/2024.",
      "more_code_info": ""
    }
  ]
}