      "status": "Terminated",
      "city": "Minneapolis",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78922",
      "recalling_firm": "Smiths Medical ASD Inc.",
      "address_1": "6000 Nathan Ln N",
      "address_2": "N/A",
      "postal_code": "55442-1690",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution- US nationwide Distribution and to the countries of : Australia, Barbados, Canada, Hong Kong, Indonesia, Jamaica, New Zealand, Philippines, Saudi Arabia, Singapore, South Africa, Trinidad and Tobago, United Arab Emirates, and United Kingdom.",
      "recall_number": "Z-1136-2018",
      "product_description": "Medfusion Syringe Pump 3010 Series    The Medfusion¿ Syringe Infusion Pumps are indicated for the following uses: 1) Administration of fluids requiring precisely controlled infusion rates including blood or blood products, lipids, drugs, antibiotics, enteral solutions and other therapeutic fluids; 2) By the following delivery routes: arterial, epidural, intravenous, intrathecal, subcutaneous, and enteral; 3)By the following delivery modes: continuous, volume/time, mass, body weight, intermittent, and bolus; 4) In critical care, anesthesia, neonatal, and pediatric applications or other healthcare settings where use of the syringe infusion pump can be monitored or supervised by a clinician; 5) Inside the MRI room mounted outside the 150 Gauss line and with shielded magnets of field strength of 1.5 Tesla.",
      "product_quantity": "16,600 pumps total",
      "reason_for_recall": "Certain Medfusion Syringe Pump Models, Series 3100, 3500, and 4000, may not recognize or may misidentify loaded medication syringes. The inability of a pump to recognize a syringe (i.e. the size of the syringe is unknown to the pump) results in an inability to complete pump programming. Misidentification of a syringe is where the pump misinterprets the syringe size.",
      "recall_initiation_date": "20171113",
      "center_classification_date": "20180319",
      "termination_date": "20250916",
      "report_date": "20180328",
      "code_info": "Product Codes 3010, 3010A with Model No. M03055, M02396, M02315, M02385, M02636, M02780, M02890, M03075, M03245, M03335, M03592, M03599, M03727, M03729, M03817, M03887, M04406, M04568, M04778, M04870, M05494, M05587, M05762, M05766, M05767, M05768, M05849, M06244, M07153, M07159, M07335, M07361, M07443, M07460, M08153, M08157, M08168, M08735, M08736, M08820, M08853, M08871, M08878, M09038, M09057, M09992, M10227, M10229, M10263, M10270, M10325, M10342, M10347, M10353, M10366, M10589, M10879, M10890, M10896, M10900, M11875, M11877, M11898, M11900, M11918, M11919"
    }
  ]
}