  "results": [
    {
      "status": "Terminated",
      "city": "Winterthur",
      "state": "N/A",
      "country": "Switzerland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70401",
      "recalling_firm": "Zimmer Gmbh",
      "address_1": "Sulzer Allee 8",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Domestic:  AK, AL, AR, AZ, CA, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MS, MA, MD, ME, MI, MN, MO, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY  International: Canada, Japan",
      "recall_number": "Z-1134-2015",
      "product_description": "Biolox Delta Femoral Head - The BIOLOX delta Ceramic Femoral Head is used in conjunction with compatible acetabular and femoral stem components in total hip arthroplasty. A variety of sizes and neck lengths are available for various patient anatomies and adjustment of the tension of the ligaments and reconstruction of the center of the natural head of the femur.",
      "product_quantity": "10340",
      "reason_for_recall": "Potential compromise of the inner sterile tray. If a compromised tray is detected, a delay in surgery may result. If undetected breach in both the inner and outer trays were to occur, and there is contamination present, a periprosthetic infection is possible, which could lead to revision.",
      "recall_initiation_date": "20141217",
      "center_classification_date": "20150220",
      "termination_date": "20150923",
      "report_date": "20150304",
      "code_info": "Item numbers:  00-8775-028-01; 00-8775-028-02; 00-8775-028-03;   00-8775-032-01; 00-8775-032-02; 00-8775-032-03;   00-8775-032-04; 00-8775-036-01; 00-8775-036-02;  00-8775-036-03; 00-8775-036-04; 00-8775-040-01;  00-8775-040-02; 00-8775-040-03; 00-8775-040-04.    Lot #'s:   2524205, 2527529, 2531817, 2543976, 2547030, 2522713, 2535758, 2541605, 2544263, 2538145, 2541613, 2524241, 2524242, 2535756, 2535757, 2538147, 2538184, 2543978, 2543979, 2549572, 2553888, 2557023, 2559372, 2524199, 2524200, 2527535, 2527536, 2527537, 2532038, 2535759, 2535765, 2535766, 2538148, 2538149, 2538152, 2538186, 2545261, 2546980, 2548230, 2551239, 2554875, 2557033, 2524204, 2527538, 2527539, 2538153, 2538189, 2544167, 2551250, 2556148, 2557156, 2535768, 2538190, 2541618, 2544169, 2552353, 2524244, 2524260, 2524261, 2527546, 2527547, 2531818, 2535769, 2535770, 2538165, 2538167, 2538168, 2541619, 2541620, 2544176, 2544177, 2547032, 2552354, 2552365, 2522849, 2524201, 2524202, 2524203, 2527548, 2527549, 2527590, 2531774, 2531815, 2531816, 2538169, 2538170, 2538171, 2538173, 2538175, 2538176, 2541852, 2541853, 2544189, 2544190, 2544191, 2548154, 2552367, 2552368, 2554887, 2554888, 2556180, 2557159, 2557170, 2524262, 2524263, 2527592, 2527593, 2531819, 2535772, 2538177, 2538178, 2538180, 2541866, 2544196, 2545262, 2548170, 2552369, 2556181, 2538191, 2541868, 2524271, 2527594, 2538194, 2538195, 2541869, 2541910, 2541911, 2544198, 2557173, 2568103, 2568982, 2582562, 2583925, 2590223, 2591046, 2591047, 2596627, 2597821, 2600144, 2602748, 2615192, 2617666, 2618635, 2621751, 2623054, 2632524, 2633562, 2635942, 2638898, 2641867, 2642795, 2645303, 2647759, 2651994, 2531222, 2538196, 2538197, 2541912, 2541913, 2544253, 2545245, 2558285, 2566263, 2566672, 2568104, 2570197, 2576875, 2576876, 2583926, 2591048, 2591049, 2596628, 2596954, 2596960, 2599455, 2602729, 2605035, 2606707, 2608757, 2614031, 2616153, 2618045, 2620408, 2622735, 2623070, 2628694, 2629838, 2633563, 2634788, 2637813, 2638899, 2639802, 2641868, 2642796, 2642797, 2645304, 2648615, 2651105, 2651106, 2524272, 2527606, 2538198, 2541914, 2541925, 2544254, 2567876, 2568983, 2573850, 2583927, 2591050, 2591051, 2597822, 2602735, 2606708, 2615193, 2618046, 2620409, 2627501, 2629839, 2632525, 2634789, 2638910, 2639803, 2642798, 2645305, 2647760, 2538199, 2538200, 2541926, 2541927, 2541928, 2544260, 2545246, 2573846, 2583928, 2591076, 2597823, 2601623, 2610484, 2617667, 2622736, 2627924, 2629840, 2634790, 2639808, 2639809, 2642799, 2646184, 2647761, 2648617, 2651108",
      "more_code_info": ""
    }
  ]
}